Wisconsin Administrative Code
Public Service Commission
Chapter PSC 113 - Service Rules For Electrical Utilities
Subchapter IX - Electric Meter Testing Programs and Record Requirements
Section PSC 113.0921 - Statistical sample testing plan for in-service, electronic (solid state) meters and electromechanical, self-contained, single phase and 3-wire network meters
Universal Citation: WI Admin Code ยง PSC 113.0921
Current through August 26, 2024
(1) The statistical sample testing plan described in pars. (a) to (e) may be used for testing electronic (solid state) meters and self-contained, single phase and 3-wire network meters without mechanical demand registers or mechanical pulsing devices in place of the periodic testing requirements of s. PSC 113.0911 if the commission authorizes the adoption of the plan by a utility.
(a) All extended
range, surge proof designed meters shall be divided into homogeneous groups
based on meter design features and age. The groups shall be further divided
into lot sizes categorized by manufacturer, type, serial number, group size or
load duty cycle with lot sizes containing a minimum of 26 meters and a maximum
of 110,000 meters. The number of lots or lot composition and size may be
changed at the end of the sample testing year to allow for increasing or
decreasing analysis of accuracy testing requirements on any segment of meters
in any lot.
(b) Annually, from each
of the assembled lots, a coded sample size specified in Table A-2, Inspection
Level IV, page 4 of Military Standard 414, (MIL-STD-414) dated 11 June 1957 and
a corresponding actual sample size as shown on Table B-3 page 45,
(MIL-STD-414), shall be randomly selected for testing and analysis purposes.
Each meter in the lot sample shall be provided with a full load and light load
test for accuracy at unity power factor, as specified under s. . A separate
statistical analysis shall be performed on each lot sample at each of these 2
load ranges. All meters not registering at either full or light load test
points shall be removed from the lot sample and replaced with a different
randomly selected meter and not more than two lot sample meters operating
outside of the accuracy bandwidth of 95% to 105% at either test load point
shall be removed and replaced from the lot sample.
(c) The statistical analysis calculations for
both the full and light load accuracy results from the sample lot tests shall
be made following the example outlined on page 43 of MIL-STD-414 with the upper
and lower specification limits, U and L designated at 102% and 98%
respectively. The test criterion for acceptance or rejection of each lot shall
be by the Standard Deviation Method, Double Specification Limit with an
Acceptable Quality Level (AQL) of 1.00 for the full load analysis and 4.00 for
the light load analysis (both normal inspection) as shown on Table B-3, page 45
of MIL-STD-414.
(d) A lot shall be
deemed acceptable for continued use if the total estimated percent defective
(P) is less than the appropriate maximum allowable percent defective (M) as
determined from Table B-3, page 45 of MIL-STD-414, following the procedure of
par. (c) for both the full load and light load analysis test points at the
respective designated Acceptable Quality Levels. All of the meters in the
accepted lot may be retained in use without further accuracy adjustments and
will be concluded to have the accuracy characteristics specified in s.
PSC 113.0811(1)
(c). Meters in the sample lot may be adjusted
for acceptable accuracy as required or maintained as necessary and returned to
the lot.
(e) A lot shall be deemed
unacceptable and rejected for continued use if the total estimated percent
defective (P) is greater than or equal to the appropriate maximum allowable
percent defective (M) as determined from Table B-3, page 45 MIL-STD-414,
following the procedure of par. (c) for both the full load and light load
analysis test points at the respective designated Acceptable Quality Levels on
any 2 annual sample testing analysis years within a five-year period for the
lot or any meters in the lot. All meters in a rejected lot shall be provided
with an appropriate test within a period of 48 months from the date of
completion of the sample analysis and all the meters tested in the rejected lot
shall be adjusted to the accuracies specified in s.
PSC 113.0811(1)
(c). All meters in a rejected lot that will
not be returned to service shall be removed within 48 months from the date of
completion of the sample analysis. These meters may be retired without a test.
Annual statistical sample testing shall be terminated during the period when
all of the meters in a rejected lot are being provided with a test and accuracy
adjustment.
(f) All meters in any
lot may be tested and adjusted for proper accuracy over a 48-month period at
the discretion of the utility without a sample analysis determination
specifying the lot test is necessary.
(g) Any meter that is included in this plan,
which is removed from service for retirement, may be retired without a
test.
Disclaimer: These regulations may not be the most recent version. Wisconsin may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google
Privacy Policy and
Terms of Service apply.