Current through August 26, 2024
(1)
(a) Except as provided in subs. (4) to (6),
the remanufacturer or other person shall monitor each pump in light liquid
service monthly to detect leaks by the methods specified in s. NR 661.1063(2).
(b) The remanufacturer
or other person shall check each pump in light liquid service by visual
inspection each calendar week for indications of liquids dripping from the pump
seal.
(2)
(a) If an instrument reading of 10,000 ppm or
greater is measured, a leak is detected.
(b) If there are indications of liquids
dripping from the pump seal, a leak is detected.
(3)
(a)
When a leak is detected, it shall be repaired as soon as practicable, but not
later than 15 calendar days after it is detected, except as provided in s. NR 661.1059.
(b) A first attempt at
repair, such as tightening the packing gland, shall be made no later than 5
calendar days after each leak is detected.
(4) Each pump equipped with a dual mechanical
seal system that includes a barrier fluid system is exempt from the
requirements under sub. (1), provided all of the following requirements are
met:
(a) Each dual mechanical seal system
shall be one of the following:
1. Operated
with the barrier fluid at a pressure that is at all times greater than the pump
stuffing box pressure.
2. Equipped
with a barrier fluid degassing reservoir that is connected by a closed-vent
system to a control device that complies with the requirements specified in s.
NR 661.1060.
3. Equipped with a
system that purges the barrier fluid into a hazardous secondary material stream
with no detectable emissions to the atmosphere.
(b) The barrier fluid system may not be a
hazardous secondary material with organic concentrations 10 percent or greater
by weight.
(c) Each barrier fluid
system shall be equipped with a sensor that will detect failure of the seal
system, the barrier fluid system, or both.
(d) Each pump shall be checked the
remanufacturer or other person by visual inspection, each calendar week, for
indications of liquids dripping from the pump seals.
(e)
1. Each
sensor as described in par. (c) shall be checked by the remanufacturer or other
person daily or be equipped with an audible alarm that shall be checked monthly
to ensure that it is functioning properly.
2. A remanufacturer or other person that
stores or treats the hazardous secondary material shall determine, based on
design considerations and operating experience, a criterion that indicates
failure of the seal system, the barrier fluid system, or both.
(f)
1. If there are indications of liquids
dripping from the pump seal or the sensor indicates failure of the seal system,
the barrier fluid system, or both based on the criterion determined in par. (e)
2., a leak is detected.
2. When a
leak is detected, it shall be repaired as soon as practicable, but not later
than 15 calendar days after it is detected, except as provided in s. NR 661.1059.
3. A first attempt at
repair, such as relapping the seal, shall be made no later than 5 calendar days
after each leak is detected.
(5) Any pump that is designated, as described
in s. NR 661.1064(7) (b), for no detectable emissions, as indicated by an
instrument reading of less than 500 ppm above background, is exempt from the
requirements under subs. (1), (3), and (4) if the pump meets all of the
following requirements:
(a) Has no externally
actuated shaft penetrating the pump housing.
(b) Operates with no detectable emissions as
indicated by an instrument reading of less than 500 ppm above background as
measured by the methods specified in s. NR 661.1063(3).
(c) Is tested for compliance with par. (b)
initially upon designation, annually, and at other times as requested by the
department.
(6) If any
pump is equipped with a closed-vent system capable of capturing and
transporting any leakage from the seal or seals to a control device that
complies with the requirements specified in s. NR 661.1060, it is exempt from
the requirements specified in subs. (1) to (5).