Current through August 26, 2024
(1) GENERAL
REQUIREMENTS.
(a) A laboratory shall
establish a quality control program that includes the analysis of appropriate
quality control samples and quality control procedures that define their
practices. These quality control procedures shall be used to assess all the
following:
1. The level of background
contamination associated with the preparation and analysis of all
samples.
2. The sensitivity of all
tests performed.
3. The level of
control of an entire analytical system.
4. The bias contributed to sample results by
all preparation and analysis steps.
5. The reproducibility of test
results.
6. The selectivity of test
methods.
(b) A
laboratory may not adjust or correct the sample results by the recoveries of
associated quality control samples or surrogates unless otherwise allowed by
method, regulation, or covered program. A laboratory may not subtract analyte
concentrations found in method blanks from sample results unless otherwise
allowed by method, regulation, or covered program.
(c) A laboratory shall document deviations
from the laboratory's quality system or exceedances of quality control samples.
To the extent the department's data systems allow, the deviations shall be
communicated with the results.
(d)
A laboratory shall establish procedures for identifying and documenting
preparation batches that facilitate determining compliance with the frequencies
of quality control samples required under this chapter.
(2) LOD.
(a) A laboratory shall determine the LOD for
all tests performed except for any of the following:
1. Biochemical oxygen demand and carbonaceous
biochemical oxygen demand.
2. Tests
for which analyzing a fortified sample is impossible or impractical.
3. Titrimetric tests.
4. Gravimetric tests, other than oil and
grease as HEM.
(b) A
laboratory shall determine the LOD of an analyte annually by 40 CFR Part 136, Appendix B. All sample-processing steps of a method shall be included in the
determination of a LOD.
Note: Links to the 40 CFR Part 136, Appendix B
can be found on the Wisconsin department of natural resources laboratory
accreditation program website.
(c) The LOD shall meet the regulatory limits
required by the covered programs.
Note: Exemptions to LOD requirements for
specific compounds are provided on the Wisconsin department of natural
resources laboratory accreditation program website.
(d) The LOD shall be adjusted when the sample
amounts used are different than those used for the LOD determination.
(e) For tests exempted from performing an LOD
under par. (a), the laboratory shall establish a reporting limit, or an
estimate of a test's sensitivity based on the intended use of the data for a
given application.
(f) The LOD
shall be determined each time there is a change in a method or instrumentation
that affects the sensitivity of an analysis.
(g) For HRGC/MS technology, the estimated
detection limit is defined in SW-846 8290A and is equivalent to the
LOD.
(3) LOQ.
(a) A laboratory shall establish the LOQ for
all tests performed except for those exempted from an LOD under sub. (2) (a).
(b) The LOQ shall meet the
regulatory limits required by the covered programs.
Note: Exemptions to LOQ requirements for
specific compounds are provided on the Wisconsin department of natural
resources laboratory accreditation program website.
(c) Except for ICP and ICP/MS single point
initial calibrations, the LOQ shall be established as 10/3 the LOD or at the
concentration of the lowest standard in the initial calibration. For ICP and
ICP/MS, when single point initial calibrations are utilized, the LOQ shall be
established as 10/3 the LOD or at the "lower limit of quantitation."
Note: The "lower limit of quantitation" is
referenced in SW-846 6010C, 6010D, 6020A, and 6020B.
(d) The LOQ shall be greater than the
LOD.
(4) REPORTING
LIMITS.
(a) Reporting limits are reserved for
those analytes exempted under sub. (2) (a) and shall be established based on a
test's sensitivity and the intended use of the data.
(b) For biochemical oxygen demand and
carbonaceous biochemical oxygen demand, the minimum reporting limit is 2 mg/L
which is based on a 300 mL sample volume. When no dilution is equal to 300 mL,
the reporting limit shall be adjusted based on the lowest dilution
reported.
(c) For total suspended
solids, the reporting limit shall be determined using the following formula:
Reporting Limit (mg/L) = 1000 / (sample volume filtered in
mL).
(5) METHOD BLANK.
(a) The laboratory shall process method
blanks along with and under the same conditions, including all sample
preparation steps, as the associated samples in a preparation batch.
Note: Method blanks are not required for
analysis of pH, alkalinity, acidity, conductivity, and solids
determinations.
(b) The
laboratory shall process method blanks at a frequency of at least one per
preparation batch up to 20 environmental samples. When samples are analyzed by
methods that do not require a preparation step before analysis, a method blank
shall be analyzed at the frequency of one per analytical batch up to 20
environmental samples.
(c) Whenever
the concentration of the method blank contains analytes of interest greater
than the LOD, the laboratory shall evaluate the nature of the interference and
its effect on each sample in a preparation batch.
(d) The acceptance criteria for method blanks
are analyte and sample specific and are established based on the highest of any
of the following:
1. The LOD.
2. Five percent of the regulatory limit for
that analyte.
3. Ten percent of the
measured concentration in the sample.
(6) LCS.
(a) Unless otherwise exempted by this
subsection, the laboratory shall process an LCS at a frequency of at least one
sample per preparation batch up to 20 environmental samples, along with and
under the same conditions as the associated samples in a preparation batch.
These conditions shall include all sample preparation steps, except for
leaching procedure extractions.
Note: TCLP leachates for metals analysis are
fortified after the leaching step is completed and before acid
preservation.
(b) The
laboratory shall fortify the LCS for the biochemical oxygen demand and
carbonaceous biochemical oxygen demand tests with a mixture of glucose-glutamic
acid as specified in approved methods of analysis. The LCS shall be processed
at a frequency of at least one sample per analytical batch for a laboratory
that analyzes more than 20 samples per week. A laboratory that analyzes fewer
than 20 samples per week shall analyze one LCS per week.
(c) The laboratory is not required to process
an LCS for tests for which analyzing a fortified sample is impossible or
impractical.
Note: An LCS need not be analyzed for the
following tests: pH, solids determinations, chlorophyll a, and color.
(d) The LCS shall be fortified with
the analytes specified by method, regulation, or covered program or all
reported analytes, except as allowed in par. (e).
(e) For analyses of polychlorinated
biphenyls, the laboratory shall fortify an LCS with at least one aroclor per
preparation batch. For other tests that determine analytes with responses that
encompass more than one chromatographic peak, as in the case of toxa-phene and
chlordane, the laboratory may fortify an LCS with a single multi-peak analyte
per preparation batch. The laboratory shall ensure that all multi-peak analytes
detectable by a method are fortified in an LCS at least once every year that
any of those analytes are reported at a detectable concentration.
(f) When the method, regulation, or covered
program do not specify control limits, the laboratory shall evaluate LCS
recoveries and generate in-house control limits, following exclusion of
outliers with a statistical technique and using the mean plus or minus 3 times
the sample standard deviation. Annually, the laboratory shall review its
generated in-house control limits and update those limits whenever the
performance characteristics change.
(g) In lieu of using generated in-house
control limits for the LCS, the laboratory may opt to use the CCV standard
limits.
(7) SELECTIVITY.
The laboratory shall establish procedures to confirm the detections of organic
analytes determined by technologies that, unlike mass spectrometry or diode
array liquid chromatography, do not provide a positive unique identification
when a covered program requires it or when the history of a sample source does
not suggest the likely presence of the detected ana-lyte.
(a) The laboratory shall develop and document
acceptance criteria, which consider retention time shifts, for chromatographic
retention time windows.
(b) The
laboratory shall document acceptance criteria for mass spectral
tuning.