Wisconsin Administrative Code
Hearing and Speech Examining Board
Chapter HAS 4 - Hearing Instrument Specialists Measurement Of Human Hearing
Section HAS 4.03 - Equipment used to measure hearing
Current through August 26, 2024
(1) Pure tone audiometry must be conducted with a pure tone audiometer which conforms to the American National Standards Institute, Standard ANSI S3.6 2018 approved September 20, 2018. Such audiometer shall be capable of generating a minimum of 9 discrete frequencies, ranging from 250 Hz through 8 KHz (250, 500, 750, 1000, 1500, 2000, 3000, 4000, 6000, 8000 Hz). Output levels over the frequency range shall conform to standard ANSI S3.6 specified above.
Note: A copy of Standard ANSI S3.6 may be obtained from the American National Standards Institute website at www.ANSI.org. A copy of the Standard is on file at the Legislative Reference Bureau.
(2) A masking source shall be either available within, or capable of being attached to, the audiometer. The masking source shall have one of the following:
(3) Audiometric equipment used in the evaluation of hearing sensitivity for the fitting and sale of hearing instruments shall be calibrated not less than once every 12 months.