(1) ADEQUATE FACILITIES. A certified laboratory shall have all of the following:
(a) Adequate facilities, equipment and
supplies to perform the tests which the operator is certified to perform at
that laboratory.
(b) At least 5
linear feet of level bench top working space per analyst.
(c) At least 50 foot candles of light on each
bench top.
(d) Effective climate
controls to maintain proper temperature and humidity for testing.
(2) QUALITY CONTROL AND MAINTENANCE. The operator of a certified laboratory shall do all of the following:
(a) Maintain laboratory
facilities, equipment and supplies in proper working condition.
(b) Maintain temperature-controlled
equipment, including incubators, water baths, refrigerators and freezers, so
that the equipment functions effectively. The operator shall measure and record
equipment temperatures at least daily when the equipment is in use and the
laboratory is staffed with trained personnel.
(c) Maintain autoclaves to ensure proper
temperature and pressure throughout the entire sterilization cycle, and keep a
record of autoclave maintenance.
(d) Calibrate each electronic test instrument
as recommended by the instrument manufacturer, and record the
results.
(e) Ensure that each
reagent and solution is clearly labeled to show its identity, its titer or
concentration, its expiration or preparation date, instructions for proper
storage, and other pertinent information. An operator shall store reagents and
solutions according to the manufacturer's specifications.
(f) Ensure that glassware is free of chips,
excessive scratches or cloudiness. Glassware used for measurement, including
pipettes, graduated cylinders and volumetric flasks, shall have clear
graduations.
(3) NOTICE OF MATERIAL CHANGES. An operator shall notify the department whenever the operator remodels laboratory facilities or installs new equipment if the remodeling or installation materially affects the performance of tests under s. ATCP 77.02 which the operator is certified to perform at that laboratory.
Disclaimer: These regulations may not be the most recent version. Wisconsin may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.