Washington Administrative Code
Title 246 - Health, Department of
RADIATION
Chapter 246-244 - Radiation protection - Wireline services
Section 246-244-080 - Leak testing of sealed sources
Current through Register Vol. 24-18, September 15, 2024
(1) Testing and recordkeeping requirements. Each licensee who uses a sealed source shall have the source tested for leakage periodically. The licensee shall keep a record of leak test results in units of becquerels (or microcuries) and retain the record for inspection by the department for three years after the leak test is performed.
(2) Method of testing. The wipe of a sealed source must be performed using a leak test kit or method approved by the department, an agreement state, a licensing state, or the United States Nuclear Regulatory Commission. The wipe sample must be taken from the nearest accessible point to the sealed source where contamination might accumulate. The wipe sample must be analyzed for radioactive contamination. The analysis must be capable of detecting the presence of 185 Bq (0.005 microcurie) of radioactive material on the test sample and must be performed by a person approved by the department, an agreement state, a licensing state, or the United States Nuclear Regulatory Commission to perform the analysis.
(3) Test frequency.
(4) Removal of leaking source from service.
(5) Exemptions from testing requirements. The following sealed sources are exempt from the periodic leak test requirements set out in subsections (1) through (4) of this section:
Statutory Authority: RCW 70.98.050. 03-12-062, § 246-244-080, filed 6/2/03, effective 7/3/03. Statutory Authority: RCW 70.98.050 and 70.98.080. 91-15-112 (Order 184), § 246-244-080, filed 7/24/91, effective 8/24/91. Statutory Authority: RCW 43.70.040. 91-02-049 (Order 121), recodified as § 246-244-080, filed 12/27/90, effective 1/31/91. Statutory Authority: RCW 70.98.080. 87-01-031 (Order 2450), § 402-38-140, filed 12/11/86.