Chapter 246-244 - Radiation protection - Wireline services
Section 246-244-070 - Radiation survey instruments
Current through Register Vol. 24-18, September 15, 2024
(1) The licensee or registrant shall maintain and use sufficient calibrated and operable radiation survey instruments at each field station and temporary job site to make physical radiation surveys as required. Instrumentation shall be capable of measuring 0.001 mSv (0.1 millirem) per hour through at least 0.5 mSv (50 millirem) per hour.
(2) Each radiation survey instrument shall be calibrated:
(3) Each licensee shall have available additional calibrated and operable radiation detection instruments capable of detecting radiation and contamination levels that could be encountered during well-logging operations or during the event of an accident, e.g., an alpha meter in case of Am-241 source rupture, a contamination meter and probe, and a high level meter capable of detecting radiation levels up to at least one roentgen per hour. The licensee may own such instruments or may make prior arrangements to obtain them expeditiously from a second party as necessary.
(4) Calibration records shall be maintained for a period of at least three years for inspection by the department.
Statutory Authority: RCW 70.98.050. 01-05-110, § 246-244-070, filed 2/21/01, effective 3/24/01. Statutory Authority: RCW 43.70.040. 91-02-049 (Order 121), recodified as § 246-244-070, filed 12/27/90, effective 1/31/91. Statutory Authority: RCW 70.98.080. 87-01-031 (Order 2450), § 402-38-120, filed 12/11/86.