Washington Administrative Code
Title 246 - Health, Department of
RADIATION
Chapter 246-240 - Radiation protection Medical use of radioactive material
PHOTON EMITTING REMOTE AFTERLOADER UNITS, TELETHERAPY UNITS, AND GAMMA STEREOTACTIC RADIOSURGERY UNITS
Section 246-240-387 - Additional technical requirements for mobile remote afterloader units
Current through Register Vol. 24-18, September 15, 2024
(1) A licensee providing mobile remote afterloader service shall:
(2) In addition to the periodic spot-checks required by WAC 246-240-381, a licensee authorized to use mobile afterloaders for medical use shall perform checks on each remote afterloader unit before use at each address of use. At a minimum, checks must be made to verify the operation of:
(3) In addition to the requirements for checks in subsection (2) of this section, a licensee shall ensure overall proper operation of the remote afterloader unit by conducting a simulated cycle of treatment before use at each address of use.
(4) If the results of the checks required in subsection (2) of this section indicate the malfunction of any system, a licensee shall lock the control console in the off position and not use the unit except as may be necessary to repair, replace, or check the malfunctioning system.
(5) A licensee shall retain a record of each check required by subsection (2) of this section in accordance with WAC 246-240-626.
Statutory Authority: RCW 70.98.050. 06-05-019, § 246-240-387, filed 2/6/06, effective 3/9/06.