PHOTON EMITTING REMOTE AFTERLOADER UNITS, TELETHERAPY UNITS, AND GAMMA STEREOTACTIC RADIOSURGERY UNITS
Section 246-240-366 - Dosimetry equipment
Current through Register Vol. 24-24, December 15, 2024
(1) Except for low dose-rate remote afterloader sources where the source output or activity is determined by the manufacturer, a licensee shall have a calibrated dosimetry system available for use. To satisfy this requirement, one of the following two conditions must be met:
(2) The licensee shall have a dosimetry system available for use for spot-check output measurements, if applicable. To satisfy this requirement, the system may be compared with a system that has been calibrated in accordance with subsection (1) of this section. This comparison must have been performed within the previous year and after each servicing that may have affected system calibration. The spot-check system may be the same system used to meet the requirement in subsection (1) of this section.
(3) The licensee shall retain a record of each calibration, intercomparison, and comparison in accordance with WAC 246-240-611.
Statutory Authority: RCW 70.98.050. 06-05-019, § 246-240-366, filed 2/6/06, effective 3/9/06.