Chapter 246-240 - Radiation protection Medical use of radioactive material
MANUAL BRACHYTHERAPY
Section 246-240-275 - Therapy-related computer systems

Universal Citation: WA Admin Code 246-240-275

Current through Register Vol. 24-18, September 15, 2024

The licensee shall perform acceptance testing on the treatment planning system of therapy-related computer systems in accordance with published protocols accepted by nationally recognized bodies. At a minimum, the acceptance testing must include, as applicable, verification of:

(1) The source-specific input parameters required by the dose calculation algorithm;

(2) The accuracy of dose, dwell time, and treatment time calculations at representative points;

(3) The accuracy of isodose plots and graphic displays; and

(4) The accuracy of the software used to determine sealed source positions from radiographic images.

Statutory Authority: RCW 70.98.050. 06-05-019, § 246-240-275, filed 2/6/06, effective 3/9/06.

Disclaimer: These regulations may not be the most recent version. Washington may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.