Washington Administrative Code
Title 246 - Health, Department of
RADIATION
Chapter 246-240 - Radiation protection Medical use of radioactive material
GENERAL TECHNICAL REQUIREMENTS
Section 246-240-113 - Requirements for possession of sealed sources and brachytherapy sources
Current through Register Vol. 24-24, December 15, 2024
(1) A licensee in possession of any sealed source or brachytherapy source shall follow the radiation safety and handling instructions supplied by the manufacturer.
(2) A licensee in possession of a sealed source shall:
(3) To satisfy the leak test requirements of this section, the licensee shall ensure the sample is analyzed by such method that the leak test can detect the presence of 185 becquerels (0.005 microcuries) of radioactive material in the sample.
(4) A licensee shall retain leak test records in accordance with WAC 246-240-572(1).
(5) If the leak test reveals the presence of 185 becquerels (0.005 microcurie) or more of removable contamination, the licensee shall:
(6) A licensee need not perform a leak test on the following sources:
(7) A licensee in possession of sealed sources or brachytherapy sources, except for gamma stereotactic radiosurgery sources, shall conduct a physical inventory of all the sources in its possession at intervals not to exceed six months. The licensee shall retain each inventory record in accordance with WAC 246-240-572.
Statutory Authority: RCW 70.98.050 and 70.98.080. 09-06-003, § 246-240-113, filed 2/18/09, effective 3/21/09. Statutory Authority: RCW 70.98.050. 06-05-019, § 246-240-113, filed 2/6/06, effective 3/9/06.