Chapter 246-240 - Radiation protection Medical use of radioactive material
GENERAL TECHNICAL REQUIREMENTS
Section 246-240-104 - Calibration of survey instruments
Universal Citation: WA Admin Code 246-240-104
Current through Register Vol. 24-24, December 15, 2024
(1) A licensee shall calibrate the survey instruments used to show compliance with this section and WAC 246-240-587 before first use, annually, and following a repair that affects the calibration. A licensee shall:
(a) Calibrate
all scales with readings up to 10 mSv (1000 mrem) per hour with a radiation
source;
(b) Calibrate two separated
readings on each scale or decade that will be used to show compliance;
and
(c) Conspicuously note on the
instrument the date of calibration.
(2) A licensee may not use survey instruments if the difference between the indicated exposure rate and the calculated exposure rate is more than 20 percent.
(3) A licensee shall retain a record of each survey instrument calibration in accordance with WAC 246-240-566.
Statutory Authority: RCW 70.98.050. 06-05-019, § 246-240-104, filed 2/6/06, effective 3/9/06.
Disclaimer: These regulations may not be the most recent version. Washington may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.