Section 246-226-050 - Dose limits

Universal Citation: WA Admin Code 246-226-050
Current through Register Vol. 24-18, September 15, 2024

The CTDIvol for the following CT procedure on phantoms may not exceed the dose limits established in Table 1.

Table 1: Dose limits

CT Procedure Phantom Size Dose Limit: CTDIvol (mGy)
Adult head 16 cm 80
Adult abdomen 32 cm 30
Pediatric head (one year old) 16 cm 40
Pediatric abdomen (40 pounds) 16 cm 20

Disclaimer: These regulations may not be the most recent version. Washington may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.