Washington Administrative Code
Title 246 - Health, Department of
RADIATION
Chapter 246-226 - Radiation protection - Computed tomography
Section 246-226-030 - Design requirements
Current through Register Vol. 24-18, September 15, 2024
(1) The location of a CT X-ray system must be designed and constructed:
(2) After January 1, 2017, and within thirty days of first use of a CT X-ray system, the registrant shall complete and keep on file a radiation protection survey of the room and surrounding areas consistent with National Council on Radiation Protection and Measurements Report #147 (2004). For CT X-ray systems in use before January 1, 2017, the registrant shall complete the radiation protection survey by January 1, 2019, and keep on file.
(3) For fixed or mobile CT X-ray systems:
(4) The registrant shall submit a revised radiation shielding plan for department review in accordance with WAC 246-225-030 after replacement of the CT X-ray system, or any change in the CT X-ray system room's construction or surrounding rooms' construction.
(5) Rooms in which a portable CT X-ray system is used are exempt from the requirements of subsections (2), (3), and (4) of this section. However, the operator must be protected to ensure exposure does not exceed dose limits established in chapter 246-221 WAC.