Standards in this section are designed to assure that optimal X-ray image quality and diagnostic information are produced so that fewer retakes are needed, and associated patient and operator exposure is minimized.
(1) When performing manual film processing (also known as hand tank processing) registrants or an operator working under the registrant's direction shall:
Table 1
THERMOMETER READINGS (DEGREES) |
MINIMUM DEVELOPING TIMES (MINUTES) |
|
C |
F |
|
27 |
80 |
2 |
79 |
2 |
|
78 |
2 1/2 |
|
77 |
2 1/2 |
|
24 |
76 |
3 |
75 |
3 |
|
74 |
3 1/2 |
|
73 |
3 1/2 |
|
22 |
72 |
4 |
71 |
4 |
|
70 |
4 1/2 |
|
69 |
4 1/2 |
|
20 |
68 |
5 |
67 |
5 1/2 |
|
66 |
5 1/2 |
|
65 |
6 |
|
18 |
64 |
6 1/2 |
63 |
7 |
|
62 |
8 |
|
61 |
8 1/2 |
|
16 |
60 |
9 1/2 |
(2) When performing automatic film processing, registrants or an operator working under the registrant's direction shall:
Table 2
DEVELOPER TEMPERATURE |
PROCESSOR DEVELOPER IMMERSION TIME* |
||
°C |
°F |
Seconds |
|
35 |
95 |
20 |
|
34.5 |
94 |
21 |
|
34 |
93 |
22 |
|
33.5 |
92 |
23 |
|
33 |
91 |
24 |
|
32 |
90 |
25 |
|
31.5 |
89 |
26 |
|
31 |
88 |
27 |
|
30.5 |
87 |
28 |
|
30 |
86 |
29 |
|
29.5 |
85 |
30 |
* |
Immersion time only, no cross-over time included. |
(3) When developing film, registrants or an operator working under the registrant's direction shall:
(4) When processing digital images, registrants or an operator working under the registrant's direction shall:
(5) The department may make X-ray film development and darkroom tests as necessary to determine compliance with this section.
Statutory Authority: RCW 70.98.050 and 70.98.080. 11-19-013, § 246-225A-090, filed 9/7/11, effective 10/8/11; 08-14-074, § 246-225A-090, filed 6/26/08, effective 7/27/08.