Chapter 246-225A - Radiation safety and diagnostic image quality standards for dental facilities
Section 246-225A-080 - Special requirements for dental intra-oral radiography
Universal Citation: WA Admin Code 246-225A-080
Current through Register Vol. 24-18, September 15, 2024
(1) Registrants using an X-ray system designed for use with an intra-oral image receptor shall use equipment that:
(a) Limits the SSD to not
less than 18 centimeters;
(b)
Limits the X-ray beam so that the beam diameter at the minimum SSD is no
greater than 7 centimeters in diameter;
(c) Has an open-ended position-indicating
device; and
(d) Has shielding
included in the beam-limiting device or position-indicating device equivalent
to that required for the diagnostic source assembly under WAC
246-225A-060(1).
(2) Registrants shall not use diagnostic dental X-ray systems with a fixed, nominal kVp of less than 55.
Statutory Authority: RCW 70.98.050 and 70.98.080. 11-19-013, § 246-225A-080, filed 9/7/11, effective 10/8/11; 08-14-074, § 246-225A-080, filed 6/26/08, effective 7/27/08.
Disclaimer: These regulations may not be the most recent version. Washington may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.