(1)
Equipment requirements.
(a)
Leakage radiation. When
the tube is operated at its leakage technique factors, the leakage radiation
shall not exceed the value specified at the distance specified for the
classification of that X-ray system:
(i)
Contact therapy systems. Leakage radiation shall not exceed 100 milliroentgens
per hour at five centimeters from the surface of the tube housing
assembly;
(ii) Zero to one hundred
fifty kVp systems. Systems shall have a leakage radiation which does not exceed
one roentgen in one hour at one meter from the source;
(iii) One hundred fifty-one to nine hundred
ninety-nine kVp systems. The leakage radiation shall not exceed one roentgen in
one hour at one meter from the source except systems that operate in excess of
500 kVp may have a leakage radiation at one meter from the source equivalent to
the exposure within one hour of the useful beam at one meter from the source
multiplied by a factor of 0.001.
(b)
Permanent beam limiting devices.
Permanent fixed diaphragms or cones used for limiting the useful beam
shall provide the same or higher degree of protection as that required by the
tube housing assembly.
(c)
Removable and adjustable beam limiting devices.
(i) Removable beam limiting devices shall,
for the portion of the useful beam to be blocked by these devices, transmit not
more than one percent of the original X-ray beam at the maximum kilovoltage and
maximum treatment filter;
(ii)
Adjustable beam limiting devices installed after the effective date of this
section shall meet the requirements of (c)(i) of this subsection;
(iii) Adjustable beam limiting devices
installed before the effective date of this section shall, for the portion of
the X-ray beam to be blocked by these devices, transmit not more than five
percent of the original X-ray beam at the maximum kilovoltage and maximum
treatment filter.
(d)
Filter and wedge systems. Filter systems shall meet the
following requirements:
(i) Filters cannot be
accidently displaced from the useful beam at any possible tube
orientation;
(ii) Each filter is
marked as to its material of construction and its thickness or wedge angle for
wedge filters;
(iii) It shall be
possible for the operator to determine the presence or absence of each filter
in the useful beam when the operator is at the control panel, either by display
at the control panel or by direct observation; and
(iv) The filter insertion slot opening shall
be covered with an attenuator equivalent to four-pound lead under operating
conditions.
(e)
Tube immobilization. The tube housing assembly shall be
capable of being immobilized during stationary treatments.
(f)
Focal spot marking. The
tube housing assembly shall be so marked that it is possible to determine the
location of the focal spot to within five millimeters, and such marking shall
be readily accessible for use during calibration procedures.
(g)
Timer.
(i) A timer shall be provided which has a
display at the treatment control panel. The timer shall be graduated in minutes
and fractions of minutes. The timer shall have a preset time selector and a
means of determining elapsed time;
(ii) The timer shall be a cumulative timer
which activates with radiation and retains its reading after irradiation is
interrupted or terminated;
(iii)
The timer shall terminate irradiation when a preselected time has
elapsed;
(iv) The timer shall
permit accurate presetting and determination of exposure times as short as 1
second;
(v) The timer shall
terminate irradiation when set to zero;
(vi) The timer shall not activate until the
shutter is opened, when patient irradiation is controlled by a shutter
mechanism.
(h)
Control panel functions. The control panel, in addition to the
displays required in other provisions of this chapter, shall have:
(i) An indication of whether X rays are being
produced;
(ii) Means for indicating
kV and X-ray tube current;
(iii)
The means for terminating an exposure at any time;
(iv) A locking device which will prevent
unauthorized use of the X-ray system; and
(v) For X-ray equipment manufactured after
the effective date of this section, a positive display of specific filter(s) in
the beam.
(i)
Source-to-patient distance. There shall be means of
determining the source-to-patient distance to within five
millimeters.
(j)
Shutters.
Unless it is possible to bring the X-ray output to the prescribed
exposure parameters within five seconds, the entire useful beam shall be
automatically attenuated by a shutter having a lead equivalency not less than
that of the tube housing assembly. In addition:
(i) After the unit is at operating
parameters, the shutter shall be controlled electrically by the operator from
the control panel;
(ii) An
indication of shutter position shall appear at the control panel.
(k)
Low filtration X-ray
tubes. Each X-ray system equipped with a beryllium or other
low-filtration window shall be clearly labeled as such upon the tube housing
assembly and at the control panel;
(l)
Alignment. When the
therapy X-ray system is equipped with a light field indicating the X-ray field,
the misalignment of one field edge to the other shall not exceed one percent of
any source-to-treatment distance.
(2)
Facility design requirements for
systems capable of operating above 50 kVp.
In addition to shielding adequate to meet requirements of
chapters 246-235 and 246-221 WAC and the shielding plan review provisions of
WAC 246-225-030, the treatment room
shall meet the following design requirements:
(a)
Warning lights.
Treatment rooms to which access is possible through more than one
entrance shall be provided with warning lights, in a readily observable
position near the outside of all access doors, which will indicate when the
useful beam is "on." Or, as an alternative, entrances other than the main one
shall be equipped with interior locks, activated for the period of exposure,
and the main entrance shall be under control of the operator.
(b)
Voice communication.
Provision shall be made for two-way aural communication between the
patient and the operator at the control panel; however, where excessive noise
levels make aural communication impractical, other methods of communication
shall be used.
(c)
Viewing
systems. Windows, mirrors, closed-circuit television, or an equivalent
system shall be provided to permit continuous observation of the patient during
irradiation and shall be so located that the operator can observe the patient
from the control panel. When the primary viewing system is by electronic means
(e.g., television), an alternate viewing system shall be available for use in
the event of electronic failure or treatment must be discontinued until repair
is made. If treatment is to be discontinued, this policy shall be included in
the written safety procedures. A copy of the safety procedures shall be
provided to the operator.
(d)
Additional requirements. Treatment rooms which contain an
X-ray system capable of operating above 150 kVp shall meet the following
additional requirements:
(i) All necessary
shielding, except for any beam interceptor, shall be provided by fixed
barriers;
(ii) The control panel
shall be outside the treatment room;
(iii) All doors of the treatment room shall
be electronically connected to the control panel such that X-ray production
cannot occur unless all doors are closed;
(iv) When the doors referred to in (d)(iii)
of this subsection are opened while the X-ray tube is activated:
(A) X-ray production shall terminate within
one second; or
(B) The radiation at
a distance of one meter from the source shall be reduced to less than 100
milliroentgens per hour within one second.
(v) After the radiation output of the X-ray
tube has been affected by the opening of any door referred to in (d)(iii) of
this subsection, it shall be possible to restore the X-ray system to full
operation only upon:
(A) Closing the door;
and subsequently
(B) Reinitiating
the exposure at the control panel.
(e)
Calibrations.
(i) The calibration of an X-ray system shall
be performed at intervals not to exceed one year and after any change or
replacement of components which could cause a change in the radiation
output.
(ii) The calibration of the
radiation output of the X-ray system shall be performed by a qualified expert
who is physically present at the facility during such calibration.
(iii) Calibration of the radiation output of
an X-ray system shall be performed with a calibrated instrument. The
calibration of such instrument shall be traceable to a national standard. The
instrument shall have been calibrated within the preceding two years.
(iv) The calibrations made pursuant to (e)(i)
of this subsection shall be such that the dose at a reference point in soft
tissue can be calculated to within ± five percent.
(v) The calibration of the X-ray system shall
include, but not be limited to, the following determinations:
(A) The exposure rates for each combination
of field size, technique factors, filter, and treatment distance
used;
(B) The degree of alignment
between the radiation field and the field indicated by the localizing device if
such device is present; and
(C) An
evaluation of the uniformity of the radiation field symmetry for the field
sizes used and any dependence upon tube housing assembly orientation.
(vi) Records of calibration
performed pursuant to (e) of this subsection shall be maintained by the
registrant for two years after completion of the calibration.
(vii) A copy of the most recent X-ray system
calibration shall be available for use by the operator at the control
panel.
(f)
Operating procedures.(i)
When a patient must be held in position for radiation therapy, mechanical
supporting or restraining devices shall be used;
(ii) The tube housing assembly shall not be
held by an individual during exposures;
(iii) No individual other than the patient
shall be in the treatment room unless such individual is protected by a barrier
sufficient to meet the requirements of chapter 246-221 WAC. No individual other
than the patient shall be in the treatment room during exposures when the kVp
exceeds 150;
(iv) The X-ray system
shall not be used in the administration of radiation therapy unless the
requirements of (e) of this subsection have been met.
Statutory Authority:
RCW
70.98.050 and
70.98.080. 91-15-083 (Order 183),
§ 246-225-120, filed 7/23/91, effective 8/23/91. Statutory Authority:
RCW
43.70.040. 91-02-049 (Order 121), recodified
as § 246-225-120, filed 12/27/90, effective 1/31/91. Statutory Authority:
RCW
70.98.080. 87-01-031 (Order 2450), §
402-28-091, filed 12/11/86; 83-19-050 (Order 2026), § 402-28-091, filed
9/16/83. Statutory Authority:
RCW
70.98.050. 81-01-011 (Order 1570), §
402-28-091, filed 12/8/80.