Current through Register Vol. 24-18, September 15, 2024
Fluoroscopic X-ray systems shall meet the following
requirements:
(1) Limitation of useful
beam.
(a) The fluoroscopic tube shall not
produce X rays unless the primary barrier is in position to intercept the
entire useful beam at all times.
(b) The entire cross section of the useful
beam shall be intercepted by the primary protective barrier of the fluoroscopic
image assembly at any source-to-image-distance (SID).
(c) Nonimage-intensified fluoroscopic
equipment shall not be used.
(d)
For fluoroscopic equipment without a spot film device, neither the length nor
the width of the fluoroscopic X-ray field in the plane of the image receptor
shall exceed that of the visible area of the image receptor by more than three
percent of the SID. The sum of the excess length and the excess width shall be
no greater than four percent of the SID. Measurements shall be made at the
minimum SID available but at no less than 20 centimeters (8 inches) table top
to image receptor distance.
(e) For
uncertified fluoroscopic equipment with a spot film device, the fluoroscopic
X-ray beam with the shutters wide open (during either fluoroscopy itself or
spot films) shall be no larger than the dimensions of the largest spot film
size for which the device is designed. Measurements shall be made at the
minimum SID available, but at no less than 20 centimeters (8 inches) table top
to the film plane distance.
(f) For
certified ( 21 C.F.R. 1020) fluoroscopic equipment with a spot film device,
neither the length nor the width of the fluoroscopic X-ray field in the plane
of the image receptor shall exceed that of the visible area of the image
receptor by more than three percent of the SID. The sum of the excess length
and width shall be no greater than four percent of the SID. Measurements shall
be made at the minimum SID available, but at no less than 20 centimeters (8
inches) table top to film plane distance.
(g) Fluoroscopic equipment beam limitation:
(i) Means shall be provided to reduce the
beam size at the plane of the image receptor to 125 square centimeters or less;
and
(ii) The minimum field size at
the greatest SID shall be equal to or less than 5 centimeters by 5
centimeters.
(2)
Activation of the fluoroscopic
tube. X-ray production in the fluoroscopic mode shall be controlled by
a deadman switch.
(3)
Entrance exposure rate allowable
limits.
(a) For equipment with or
without automatic brightness control, the exposure
rate measured at the point where the center of the useful beam enters the
patient shall not exceed 2.58 x 10-3 coulombs per
kilogram per minute (ten roentgens per minute), except during film recording of
fluoroscopic images or when an optional high level control (HLC) is
activated.
(b) For equipment
provided with HLC, the equipment shall not be operable at a combination of tube
potential and current which will result in an
exposure
rate in excess of 1.29 x 10-3 coulombs per kilogram
per minute (5 roentgens per minute) at the point where the center of the useful
beam enters the patient, unless the HLC is activated.
(i) Special means of activation of high level
controls, such as additional pressure applied continuously by the operator,
shall be required to avoid accidental use.
(ii) A continuous signal audible to the
fluoroscopist shall indicate the high level control is employed.
(c) Measuring compliance of
entrance
exposure rate limits. Compliance with
subsection (3) of this section shall be determined as follows:
(i) Movable grids and compression devices
shall be removed from the useful beam during the measurement;
(ii) If the source is below the table,
exposure rate shall be measured 1 centimeter above the
table top or cradle;
(iii) If the
source is above the table, the exposure rate shall be
measured at 30 centimeters above the table top with the end of the
beam-limiting device or spacer positioned as closely as possible to the point
of measurement;
(iv) In a C-arm
type of fluoroscope, the exposure rate shall be
measured 30 centimeters from the input surface of the fluoroscopic imaging
assembly, with the source positioned at any available SID, provided the end of
the beam-limiting device or spacer is no closer than 30 centimeters from the
input surface of fluoroscopic imaging assembly; and
(v) In a lateral-type fluoroscope, the
exposure rate shall be measured at a point 15 centimeters from the center line
of the X-ray table with the end of the beam-limiting device or spacer
positioned as closely as possible to the point of measurement. If the table top
is movable, the table top shall be positioned as closely as possible to the
lateral X-ray source, with the end of the beam-limiting device or spacer no
closer than 15 centimeters to the center line of the X-ray table.
(d) Periodic measurement of
entrance
exposure rate limits.
(i) Periodic measurements of the
exposure rate shall be made. An adequate period for
such measurements shall be annually or after maintenance of the system
affecting the exposure rate.
(ii) Results of
exposure rate measurements shall be available where
the fluoroscopist has ready access to the measurements while using that
fluoroscope. Results of the measurements shall include:
(A) The maximum possible coulombs per
kilogram per minute (R/minute), as well as the technique factors associated
with it;
(B) The name of the person
performing the measurements;
(C)
The last date the measurements were performed; and
(D) The type of device used in making the
measurements.
(iii)
Conditions of measurement:
(A) The kVp shall
be adjusted to that which will produce the maximum entrance exposure
rate;
(B) The high level control,
if present, shall not be activated;
(C) The X-ray systems that incorporate
automatic exposure rate control (automatic brightness
control) shall have sufficient material, for example, lead or lead equivalence,
placed in the useful beam to produce the maximum output of the X-ray system;
and
(D) X-ray systems not
incorporating automatic exposure rate control shall
utilize whatever combination of kVp, mA, and other selectable parameters that
will generate the highest exposure rate of the X-ray system. Materials, for
example, an attenuation block, may be placed in the useful beam to protect the
imaging system, as long as the material does not affect the measurement of the
exposure rate.
(4)
Barrier transmitted radiation
rate limits.
(a) The
exposure rate due to transmission through the primary
protective barrier with the attenuation block in the useful beam, combined with
radiation from the image intensifier, if provided, shall not exceed 5.16 x
10-7 coulombs per kilogram per hour (2
milliroentgens per hour) for each 2.58 x 10-4
coulombs per kilogram per minute (roentgen per minute) of entrance
exposure rate. The barrier transmission measurement
shall be made at 10 centimeters from an accessible surface of the fluoroscopic
imaging assembly beyond the plane of the image receptor.
(b) Measuring compliance of barrier
transmission.
(i) The
exposure rate due to transmission through the primary
protective barrier combined with radiation from the image intensifier shall be
determined by measurements averaged over an area of 100 square centimeters with
no linear dimension greater than 20 centimeters.
(ii) If the source is below the table top,
the measurement shall be made with the input surface of the fluoroscopic
imaging assembly positioned 30 centimeters above the table top.
(iii) If the source is above the table top
and the SID is variable, the measurement shall be made with the end of the
beam-limiting device or spacer as close to the table top as it can be placed,
provided the beam-limiting device or spacer shall not be closer than 30
centimeters.
(iv) Movable grids and
compression devices shall be removed from the useful beam during the
measurement.
(5)
Indication of potential and
current. During fluoroscopy and cinefluorography, X-ray tube potential
and current shall be continuously indicated.
(6)
Source-skin distance (SSD).
The source to skin distance shall not be less than:
(a) 38 centimeters on stationary
fluoroscopes;
(b) 30 centimeters on
mobile fluoroscopes; and
(c) 20
centimeters for image intensified fluoroscopes used for specific surgical
application. The user must provide precautionary measures for the use of the
fluoroscope due to its short SSD.
(7)
Fluoroscopic timer.
(a) Means shall be provided to preset the
cumulative on-time of the fluoroscopic tube. The maximum cumulative time of the
timing device shall not exceed five minutes without resetting.
(b) A signal audible to the fluoroscopist
shall indicate the completion of a preset cumulative on-time. Such signal shall
continue to sound while X rays are produced until the timing device is reset.
Alternatively, the timing device may terminate exposures at the end of the
preset time.
(c) Total fluoroscopic
on-time for each patient shall be recorded, either in patient's chart or in a
separate log.
(8)
Control of scattered radiation.
(a) Fluoroscopic table designs when combined
with normal operating procedures shall be such that no unprotected part of
staff or ancillary person's body shall be exposed to unattenuated scattered
radiation which originates from under the table. The attenuation required shall
be not less than 0.25 mm lead equivalent.
(b) Equipment configuration when combined
with procedures shall be such that no portion of staff or ancillary person's
body, except the extremities, shall be exposed to the unattenuated scattered
radiation emanating from above the table top unless:
(i) The radiation has passed through not less
than 0.25 mm lead equivalent material, for example, drapes, Bucky-slot
cover-sliding or folding panel, or self-supporting curtains, in addition to
lead equivalency provided by the protective apron referred to under WAC
246-225-020(2)(e);
and
(ii) Exceptions to subdivision
(b) of this subsection may be made in some special procedures where a sterile
field will not permit the use of the normal protective barriers. Where the use
of prefitted sterilized covers for the barriers is practical, the department
shall not permit such exception.
(9)
Radiation therapy simulation
systems. Radiation therapy simulation systems shall be exempt from the
requirements of subsection (3) of this section. In addition, these systems
shall be exempt from:
(a) Subsections (1) and
(4) of this section provided such systems are designed and used in such a
manner that no individual other than the patient is in the X-ray room when the
system is producing X rays; and
(b)
Subsection (7) of this section if such systems are provided with a means of
indicating the cumulative time that an individual patient has been exposed to X
rays.
Statutory Authority:
RCW
70.98.050. 94-01-073, § 246-225-050,
filed 12/9/93, effective 1/9/94. Statutory Authority:
RCW
70.98.050 and
70.98.080. 91-15-083 (Order 183),
§ 246-225-050, filed 7/23/91, effective 8/23/91. Statutory Authority:
RCW
43.70.040. 91-02-049 (Order 121), recodified
as § 246-225-050, filed 12/27/90, effective 1/31/91. Statutory Authority:
RCW
70.98.080. 87-01-031 (Order 2450), §
402-28-040, filed 12/11/86; 83-19-050 (Order 2026), § 402-28-040, filed
9/16/83. Statutory Authority:
RCW
70.98.050. 81-01-011 (Order 1570), §
402-28-040, filed 12/8/80; Order 1084, § 402-28-040, filed 1/14/76; Order
1, § 402-28-040, filed 1/8/69; Rules (part), filed
10/26/66.