Chapter 246-760 - Auditory and visual standards - School districts
AUDITORY ACUITY STANDARDS
Section 246-760-030 - What are the auditory acuity screening standards for screening equipment and procedures?

Universal Citation: WA Admin Code 246-760-030

Current through Register Vol. 24-18, September 15, 2024

(1) Schools shall use auditory screening equipment providing tonal stimuli at frequencies at one thousand, two thousand, and four thousand herz (Hz) at hearing levels of twenty decibels (dB), as measured at the earphones, in reference to American National Standards Institute (ANSI) 1996 standards.

(2) Qualified persons will check the calibration of frequencies and intensity at least every twelve months, at the earphones, using equipment designed for audiometer calibration.

Statutory Authority: RCW 28A.210.200. 02-20-079, § 246-760-030, filed 9/30/02, effective 10/31/02. Statutory Authority: RCW 43.20.050. 91-02-051 (Order 124B), recodified as § 246-760-030, filed 12/27/90, effective 1/31/91. Statutory Authority: RCW 28A.31.030. 87-22-010 (Order 306), § 248-148-031, filed 10/26/87.

Disclaimer: These regulations may not be the most recent version. Washington may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.