Current through Register Vol. 24-18, September 15, 2024
The medical test site must use quality control procedures,
providing and assuring accurate and reliable test results and reports, meeting
the requirements of this chapter.
(1)
The medical test site must have and follow written procedures and policies
available in the work area for:
(a) Analytical
methods used by the technical personnel including:
(i) Principle;
(ii) Specimen collection and processing
procedures;
(iii)
Equipment/reagent/supplies required;
(iv) Preparation of solutions, reagents, and
stains;
(v) Test
methodology;
(vi) Quality control
procedures;
(vii) Procedures for
reporting results (normal, abnormal, and critical values);
(viii) Reference range;
(ix) Troubleshooting guidelines - limitations
of methodology;
(x) Calibration
procedures; and
(xi) Pertinent
literature references; and
(b) Alternative or backup methods for
performing tests including the use of a reference facility if
applicable.
(2) The
medical test site must establish written criteria for and maintain appropriate
documentation of:
(a) Temperature-controlled
spaces and equipment;
(b)
Preventive maintenance activities;
(c) Equipment function checks;
(d) Procedure calibrations; and
(e) Method/instrument validation
procedures.
(3) The
medical test site must maintain documentation of:
(a) Expiration date, lot numbers, and other
pertinent information for:
(i)
Reagents;
(ii) Solutions;
(iii) Culture media;
(iv) Controls;
(v) Calibrators;
(vi) Standards;
(vii) Reference materials; and
(viii) Other testing materials; and
(b) Testing of quality control
samples.
(4) For
quantitative tests, the medical test site must perform quality
control as follows:
(a) Include two reference
materials of different concentrations each day of testing unknown samples, if
these reference materials are available; or
(b) Follow an equivalent quality testing
procedure that meets federal CLIA regulations.
(5) For qualitative tests, the
medical test site must perform quality control as follows:
(a) Use positive and negative reference
material each day of testing unknown samples; or
(b) Follow an equivalent quality testing
procedure that meets federal CLIA regulations.
(6) The medical test site must:
(a) Use materials within their documented
expiration date;
(b) Not
interchange components of kits with different lot numbers, unless specified by
the manufacturer;
(c) Determine the
statistical limits for each lot number of unassayed reference materials through
repeated testing;
(d) Use the
manufacturer's reference material limits for assayed material, provided they
are:
(i) Verified by the medical test site;
and
(ii) Appropriate for the
methods and instrument used by the medical test site;
(e) Make reference material limits readily
available;
(f) Report patient
results only when reference materials are within acceptable limits;
(g) Rotate control material testing among all
persons who perform the test;
(h)
Use calibration material from a different lot number than that used to
establish a cut-off value or to calibrate the test system, if using calibration
material as a control material;
(i) For each test system that has an
extraction phase, include two control materials, including one that is capable
of detecting errors in the extraction process;
(j) For each molecular amplification
procedure, include two control materials and, if reaction inhibition is a
significant source of false negative results, a control material capable of
detecting the inhibition is required; and
(k) Comply with general quality control
requirements as described in Table 090-1, unless otherwise specified in
subsection (9)(a) through (l) of this section.
(7) The medical test site must perform, when
applicable:
(a) Calibration and calibration
verification for moderate and high complexity testing as described
in Table 090-2;
(b) Validation for
moderate complexity testing by verifying the following performance
characteristics when the medical test site introduces a new procedure
classified as moderate complexity:
(i)
Accuracy;
(ii) Precision;
(iii) Reportable range of patient test
results; and
(iv) If using the
reference range provided by the manufacturer, that it is appropriate for the
patient population;
(c)
Validation for
high complexity testing:
(i) When the medical test site introduces a
new procedure classified as high complexity;
(ii) For each method that is developed
in-house, is a modification of the manufacturer's test procedure, or is an
instrument, kit or test system that has not been cleared by FDA; and
(iii) By verifying the following performance
characteristics:
(A) Accuracy;
(B) Precision;
(C) Analytical sensitivity;
(D) Analytical specificity to include
interfering substances;
(E)
Reference ranges (normal values);
(F) Reportable range of patient test results;
and
(G) Any other performance
characteristic required for test performance.
(8) When patient values are above
the maximum or below the minimum calibration point or the reportable range, the
medical test site must:
(a) Report the
patient results as greater than the upper limit or less than the lower limit or
an equivalent designation; or
(b)
Use an appropriate procedure to rerun the sample allowing results to fall
within the established linear range.
Table 090-1 General Quality Control
Requirements
|
Control Material
|
Frequency
|
(a) Each batch or shipment of reagents,
discs, antisera, and identification systems |
* Appropriate control materials for
positive and negative reactivity |
* When prepared or opened, unless otherwise
specified |
(b) Each batch or shipment of
stains |
* Appropriate control materials
for positive and negative reactivity |
* When prepared or opened; and |
* Each day of use, unless otherwise
specified |
(c) Fluorescent and immunohistochemical
stains |
* Appropriate control materials for
positive and negative reactivity |
* Each time of use, unless otherwise
specified |
(d) Quality control for each
specialty and subspecialty |
* Appropriate control materials; or |
* At least as frequently as specified in
this section; |
* Equivalent mechanism to
assure the quality, accuracy, and precision of the test if reference materials
are not available |
* More frequently if recommended by the
manufacturer of the instrument or test procedure; or |
* More frequently if specified by the
medical test site |
(e) Direct antigen detection
systems without procedural controls |
* Positive and negative
controls that evaluate both the extraction and reaction phase |
* Each batch, shipment, and new lot number;
and |
* Each day of use |
Table 090-2 Calibration and Calibration
Verification-Moderate and High Complexity Testing
|
Calibration Material
|
Frequency
|
CALIBRATION
|
* Calibration materials
appropriate for methodology |
* Initial on-site
installation/implementation of instrument/method; |
* At the frequency recommended by the
manufacturer; and |
* Whenever calibration verification fails
to meet the medical test site's acceptable limits for calibration verification.
|
CALIBRATION
VERIFICATION
|
* Use assayed material, if available, at
the lower, mid-point, and upper limits of procedure's report-able range; or
|
* At least every six months; |
* Demonstrate alternate method
of assuring accuracy at the lower, mid-point, and upper limits of procedure's
reportable range |
* When there is a complete change of
reagents (i.e., new lot number or different manufacturer) is introduced; |
* When major preventive maintenance is
performed or there is a replacement of critical parts of equipment; or |
* When controls are outside of the medical
test site's acceptable limits or exhibit trends. |
(9) The medical test site must perform
quality control procedures as described for each specialty and subspecialty in
(a) through (l) of this subsection.
(a)
Chemistry.
Perform quality control procedures for chemistry as described
in Table 090-3 or follow an equivalent quality testing procedure that meets
federal CLIA regulations.
Table 090-3 Quality Control
Procedures-Chemistry
Subspecialty/Test
|
Qualitative
|
Quantitative
|
|
Control Material |
Frequency |
Control Material |
Frequency |
Routine Chemistry |
* Positive and negative reference material
|
* Each day of use |
* Two levels of reference material in
different concentrations |
* Each day of use |
Toxicology |
|
|
|
|
* GC/MS for drug screening |
* Analyte-specific control |
* With each run of patient specimens |
* Analyte-specific control |
* With each analytical run |
* Urine drug screen |
* Positive control containing at least one
drug representative of each drug class to be reported; must go through each
phase of use including extraction |
* With each run of patient specimens |
|
|
Urinalysis |
|
|
|
|
* Nonwaived instrument |
|
|
* Two levels of control material |
* Each day of use |
* Refractometer for specific
gravity |
|
|
* Calibrate to zero with distilled water
|
* Each day of use |
|
|
* One level of control material |
|
Blood Gas Analysis |
|
|
* Calibration |
* Follow manufacturer's specifications and
frequency |
|
|
* One level of control material |
* Each eight hours of testing, using both
low and high values on each day of testing |
|
|
* One-point calibration or one control
material |
* Each time patient specimen is tested,
unless automated instrument internally verifies calibration every thirty
minutes |
Electrophoresis |
* One control containing fractions
representative of those routinely reported in patient specimens |
* In each electrophoretic cell |
* One control containing fractions
representative of those routinely reported in patient specimens |
* In each electrophoretic cell |
(b)
Hematology.
(i) Run patient and
quality control samples in duplicate for manual cell counts;
(ii) If reference material is unavailable,
document the mechanism used to assure the quality, accuracy, and precision of
the test; and
(iii) Perform quality
control procedures for hematology as described in Table 090-4 or follow an
equivalent quality testing procedure that meets federal CLIA regulations.
Table 090-4 Quality Control
Procedures-Hematology
|
Control Material
|
Frequency
|
Automated |
* Two levels of reference material in
different concentrations |
* Each day that patient samples are tested
|
Manual Blood Counts |
* One level of reference material |
* Every eight hours that patient samples
are tested |
Qualitative Tests |
* Positive and negative reference material
|
* Each day of testing |
(c)
Coagulation.
(i) Run patient and quality control samples
in duplicate for manual coagulation test (tilt tube);
(ii) If reference material is unavailable,
document the mechanism used to assure the quality, accuracy, and precision of
the test; and
(iii) Perform quality
control procedures for coagulation as described in Table 090-5 or follow an
equivalent quality testing procedure that meets federal CLIA regulations.
Table 090-5 Quality Control
Procedures-Coagulation
|
Control Material
|
Frequency
|
Automated |
* Two levels of reference
material in different concentrations |
* Every eight hours that patient samples
are tested; and |
* Each time reagents are changed |
Manual Tilt Tube Method |
* Two levels of reference
material in different concentrations |
* Every eight hours that patient samples
are tested; and |
* Each time reagents are changed |
(d)
General immunology.
(i) Employ reference materials for all test
components to ensure reactivity;
(ii) Report test results only when the
predetermined reactivity pattern of the reference material is observed;
and
(iii) Perform quality control
procedures for general immunology as described in Table 090-6 or follow an
equivalent quality testing procedure that meets federal CLIA regulations.
Table 090-6 Quality Control Procedures-General
Immunology
|
Control Material
|
Frequency
|
Serologic tests on unknown specimens |
* Positive and negative reference material
|
* Each day of testing |
Kits with procedural (internal)
controls |
* Positive and negative reference material
(external controls) |
* When kit is opened; and |
* Procedural (internal)
controls |
* Each day of testing, or follow an
equivalent quality testing procedure that meets federal CLIA regulations |
* Each time patient sample is tested |
(e)
Syphilis serology.
(i) Use equipment, glassware, reagents,
controls, and techniques that conform to manufacturer's
specifications;
(ii) Employ
reference materials for all test components to ensure reactivity; and
(iii) Perform serologic tests on unknown
specimens each day of testing with a positive serum reference material with
known titer or graded reactivity and a negative reference material.
(f)
Microbiology.
(i) Have available and use:
(A) Appropriate stock organisms for quality
control purposes; and
(B) A
collection of slides, photographs, gross specimens, or text books for reference
sources to aid in identification of microorganisms;
(ii) Document all steps (reactions) used in
the identification of microorganisms on patient specimens;
(iii) For antimicrobial susceptibility
testing:
(A) Record zone sizes or minimum
inhibitory concentration for reference organisms; and
(B) Zone sizes or minimum inhibitory
concentration for reference organisms must be within established limits before
reporting patient results; and
(C)
Perform quality control on antimicrobial susceptibility testing media as
described in Table 090-8;
(iv) For noncommercial media, check each
batch or shipment for sterility, ability to support growth and, if appropriate,
selectivity, inhibition, or biochemical response;
(v) For commercial media:
(A) Verify that the product insert specifies
that the quality control checks meet the requirements for media quality control
as outlined by the Clinical Laboratory Standards Institute (CLSI).
M22-A3 Quality Control for Commercially Prepared Microbiological
Culture Media; Approved Standard-Third Edition. June 2004. (Volume 24,
Number 19);
(B) Keep records of the
manufacturer's quality control results;
(C) Document visual inspection of the media
for proper filling of the plate, temperature or shipment damage, and
contamination before use; and
(D)
Follow the manufacturer's specifications for using the media; and
(vi) For microbiology
subspecialties:
(A)
Bacteriology: Perform quality control procedures for bacteriology
as described in Tables 090-7 and 090-8.
Table 090-7 Quality Control
Procedures-Bacteriology
|
Control Material
|
Frequency
|
Reagents, disks, and identification systems
|
* Positive and negative reference
organisms, unless otherwise specified |
* Each batch, shipment, and new lot number
unless otherwise specified |
Catalase, coagulase, oxidase, and
Beta-lactamase CefinaseTM reagents |
|
|
Bacitracin, optochin, ONPG, X and V disks
or strips |
|
|
Stains, unless otherwise
specified; DNA probes; and all beta-lactamase methods other than
CefinaseTM
|
* Positive and negative
reference organisms |
* Each batch, shipment, and new lot number;
and |
* Each day of use |
Fluorescent stains |
* Positive and negative
reference organisms |
* Each batch, shipment, and new lot number;
and |
* Each time of use |
Gram stains |
* Positive and negative
reference organisms |
* Each batch, shipment, and new lot number;
and |
* Each week of use |
Direct antigen detection
systems without procedural controls |
* Positive and negative
controls that evaluate both the extraction and reaction phase |
* Each batch, shipment, and new lot number;
and |
* Each day of use |
Test kits with procedural
(internal) controls |
* Positive and negative reference material
(external) controls |
* Each batch, shipment, and new lot number;
and |
* Procedural (internal)
controls |
* Each day of testing, or follow an
equivalent quality testing procedure that meets federal CLIA regulations |
* Each time patient sample is tested |
Antisera |
* Positive and negative
reference material |
* Each batch, shipment, and new lot number;
and |
* Every six months |
Table 090-8 Quality Control Procedures-Bacteriology -
Media for Antimicrobial Susceptibility Testing
|
Control Material
|
Frequency
|
Check each new batch of media and each new
lot of antimicrobial disks or other testing systems (MIC) |
* Approved reference organisms (ATCC
organisms) |
* Before initial use and each day of testing;
or
* May be done weekly if the medical test site can
meet the quality control requirements for antimicrobial disk susceptibility
testing as outlined by CLSI M100S Performance Standards for
Antimicrobial Susceptibility Testing; Twenty-Sixth Edition.
|
(B)
Mycobacteriology: Perform quality control procedures for
mycobacteriology as described in Table 090-9.
Table 090-9 Quality Control
Procedures-Mycobacteriology
|
Control Material
|
Frequency
|
All reagents or test procedures used for
mycobacteria identification unless otherwise specified |
* Acid-fast organism that produces a
positive reaction and an acid-fast organism that produces a negative reaction
|
* Each day of use |
Acid-fast stains |
* Acid-fast organism that produces a
positive reaction and an organism that produces a negative reaction |
* Each day of use |
Fluorochrome acid-fast stains |
* Acid-fast organism that produces a
positive reaction and an acid-fast organism that produces a negative reaction
|
* Each time of use |
Susceptibility tests performed
on Mycobacterium tuberculosis isolates |
* Appropriate control
organism(s) |
* Each batch of media, and each lot number
and shipment of antimycobacterial agent(s) before, or concurrent with, initial
use |
* Each week of use |
(C)
Mycology: Perform quality control procedures for mycology as
described in Table 090-10.
Table 090-10 Quality Control
Procedures-Mycology
|
Control Material
|
Frequency
|
Susceptibility tests: Each drug NOTE:
Establish control limits and criteria for acceptable control results prior to
reporting patient results |
* One control strain that is susceptible to
the drug |
* Each day of use |
Lactophenol cotton blue stain |
* Appropriate control organism(s) |
* Each batch or shipment and each lot
number |
Acid-fast stains |
* Organisms that produce positive and
negative reactions |
* Each day of use |
Reagents for biochemical and other
identification test procedures |
* Appropriate control organism(s) |
* Each batch or shipment and each lot
number |
Commercial identification systems utilizing
two or more substrates |
* Organisms that verify positive and
negative reactivity of each media type |
* Each batch or shipment and each lot
number |
(D)
Parasitology:
(I) Have available
and use:
* Reference collection of slides or photographs and, if
available, gross specimens for parasite identification; and
* Calibrated ocular micrometer for determining the size of
ova and parasites, if size is a critical parameter.
(II) Check permanent stains each month of use
with reference materials.
(E)
Virology:
(I) Have available:
* Host systems for isolation of viruses; and
* Test methods for identification of viruses that cover the
entire range of viruses that are etiologically related to the clinical diseases
for which services are offered; and
(II) Simultaneously culture uninoculated
cells or cell substrate as a negative control when performing virus
identification.
(g)
Histopathology: Fluorescent
and immunohistochemical stains must be checked for positive and negative
reactivity each time of use. For all other differential or special stains,
include a control slide of known reactivity with each slide or group of slides
and document reactions.
(h)
Cytology.
(i) Processing
specimens:
(A) Stain all gynecological smears
using a Papanicolaou or a modified Papanicolaou staining method;
(B) Have methods to prevent
cross-contamination between gynecologic and nongynecologic specimens during the
staining process; and
(C) Stain
nongynecological specimens that have a high potential for cross-contamination
separately from other non-gynecological specimens, and filter or change the
stains following staining.
(ii) Performing specimen examinations:
(A) All cytology preparations must be
evaluated on the premises of the medical test site;
(B) Technical personnel must examine, unless
federal law and regulation specify otherwise, no more than one hundred
cytological slides (one patient specimen per slide; gynecologic,
nongynecologic, or both) in a twenty-four-hour period and in no less than an
eight-hour work period;
(C)
Previously examined negative, reactive, reparative, atypical, premalignant or
malignant gynecological cases and previously examined nongynecologic cytology
preparations and tissue pathology slides examined by a technical supervisor are
not included in the one hundred slide limit;
(D) Each nongynecologic slide preparation
made using liquid-based slide preparatory techniques that result in cell
dispersion over one-half or less of the total available slide may be counted as
one-half slide; and
(E) Records of
the total number of slides examined by each individual at all sites during each
twenty-four-hour period must be maintained.
(iii) Establish and implement a quality
assurance program that ensures:
(A) There is
criteria for submission of material;
(B) All providers submitting specimens are
informed of these criteria;
(C) All
samples submitted are assessed for adequacy;
(D) Records of initial examinations and
rescreening results are available and documented;
(E) Rescreening of benign gynecological
slides is:
(I) Performed by an individual who
meets the personnel requirements for technical or general supervisor in
cytology as defined under 42 C.F.R. Part 493 Subpart M;
(II) Completed before reporting patient
results on those selected cases;
(III) Performed and documented on:
* No less than ten percent of the benign gynecological
slides; and
* Includes cases selected at random from the total caseload
and from patients or groups of patients that are identified as having a high
probability of developing cervical cancer, based on available patient
information;
(F)
The technical supervisor:
(I) Confirms all
gynecological smears interpreted to be showing reactive or reparative changes,
atypical squamous or glandular cells of undetermined significance, or to be in
the premalignant (dysplasia, cervical intraepithelial neoplasia or all squamous
intraepithelial neoplasia lesions including human papillomavirus-associated
changes) or malignant category;
(II) Reviews all nongynecological cytological
preparations; and
(III)
Establishes, documents, and reassesses, at least every six months, the workload
limits for each cytotechnologist;
(G) All cytology reports with a diagnosis of
high-grade squamous intraepithelial lesion (HSIL), adenocarcinoma, or other
malignant neoplasms are correlated with prior cytology reports and with
histopathology reports if available, and the causes of any discrepancies are
determined;
(H) Review of all
normal or negative gynecological specimens received within the previous five
years, if available in the laboratory system, or records of previous reviews,
for each patient with a current high grade intraepithelial lesion or moderate
dysplasia of CIN-2 or above;
(I)
Notification of the patient's physician if significant discrepancies are found
that would affect patient care and issuance of an amended report;
(J) An annual statistical evaluation of the
number of cytology cases examined, number of specimens processed by specimen
type, volume of patient cases reported by diagnosis, number of cases where
cytology and histology are discrepant, number of cases where histology results
were unavailable for comparison, and number of cases where rescreen of negative
slides resulted in reclassification as abnormal; and
(K) Evaluation and documentation of the
performance of each individual examining slides against the medical test site's
overall statistical values, with documentation of any discrepancies, including
reasons for the deviation and corrective action, if appropriate.
(i)
Immunohematology/transfusion services.
(i) Perform ABO grouping, Rh (D) typing,
antibody detection and identification, and compatibility testing as described
by the Food and Drug Administration (FDA) under 21 C.F.R. Parts 606 and 640.
(A) Perform ABO grouping:
(I) By concurrently testing unknown red cells
with FDA approved anti-A and anti-B grouping sera;
(II) Confirm ABO grouping of unknown serum
with known A1 and B red cells;
(B) Perform Rh (D) typing by testing unknown
red cells with anti-D (anti-Rh) blood grouping serum; and
(C) Perform quality control procedures for
immunohematology as described in Table 090-11.
(ii) Blood and blood products:
(A) Collecting, processing, and distributing:
(I) Must comply with FDA requirements listed
under 21 C.F.R. Parts 606, 610.40, 610.53, and 640; and
(II) Must establish, document, and follow
policies to ensure positive identification of a blood or blood product
recipient.
(B) Labeling
and dating must comply with FDA requirements listed under 21 C.F.R. 606 Subpart
G, and 610.53.
(C) Storing:
(I) There must be an adequate temperature
alarm system that is regularly inspected.
(II) The system must have an audible alarm
system that monitors proper blood and blood product storage temperature over a
twenty-four-hour period.
(III) High
and low temperature checks of the alarm system must be documented.
(D) Collection of heterologous or
autologous blood products on-site:
(I) Must
register with the FDA; and
(II)
Have a current copy of the form FDA 2830 "Blood Establishment Registration and
Product Listing."
(iii) Must have an agreement approved by the
director for procurement, transfer, and availability to receive products from
outside entities.
(iv) Promptly
investigate transfusion reactions according to established procedures, and take
any necessary remedial action.
Table 090-11 Quality Control
Procedures-Immunohematology
Reagent
|
Control Material
|
Frequency
|
ABO antisera |
* Positive control |
* Each day of use |
Rh antisera |
* Positive and negative controls |
* Each day of use |
* Patient control to detect false positive
Rh test results |
* When required by the manufacturer |
Other antisera |
* Positive and negative controls |
* Each day of use |
ABO reagent red cells |
* Positive control |
* Each day of use |
Antibody screening cells |
* Positive control using at least one known
antibody |
* Each day of use |
(j)
Histocompatibility.(i) Use
applicable quality control standards for immunohematology, transfusion
services, and diagnostic immunology as described in this chapter; and
(ii) Meet the standards for
histocompatibility as listed in 42 C.F.R. Part 493.1278, Standard:
Histocompatibility, available from the department upon request.
(k)
Cytogenetics.
(i) Document:
(A) Number of metaphase chromosome spreads
and cells counted and karyotyped;
(B) Number of chromosomes counted for each
metaphase spread;
(C) Media
used;
(D) Reactions
observed;
(E) Quality of banding;
and
(F) Sufficient resolution
appropriate for the type of tissue or specimen and the type of study required
based on the clinical information provided;
(ii) Assure an adequate number of karyotypes
are prepared for each patient according to the indication given for performing
cytogenetics study;
(iii) Use an
adequate patient identification system for:
(A) Patient specimens;
(B) Photographs, photographic negatives, or
computer stored images of metaphase spreads and karyotypes;
(C) Slides; and
(D) Records; and
(iv) Perform full chromosome analysis for
determination of sex.
(l)
Radiobioassay and
radioimmunoassay.
(i) Check the
counting equipment for stability each day of use with radioactive standards or
reference sources; and
(ii) Meet
Washington state radiation standards described under chapter 70.98 RCW and
chapters 246-220, 246-221, 246-222, 246-232, 246-233, 246-235, 246-239,
246-247, 246-249, and 246-254 WAC.
Statutory Authority:
RCW
70.42.005 and 42 C.F.R. Part 493. 05-04-040,
§ 246-338-090, filed 1/27/05, effective 3/19/05. Statutory Authority:
RCW
70.42.005,
70.42.060. 01-02-069, §
246-338-090, filed 12/29/00, effective 1/29/01. Statutory Authority:
RCW
70.42.005,
70.42.060 and chapter 70.42 RCW.
00-06-079, § 246-338-090, filed 3/1/00, effective 4/1/00. Statutory
Authority:
RCW
70.42.005. 97-14-113, § 246-338-090,
filed 7/2/97, effective 8/2/97. Statutory Authority: Chapter 70.42 RCW.
93-18-091 (Order 390), § 246-338-090, filed 9/1/93, effective 10/2/93;
91-21-062 (Order 205), § 246-338-090, filed 10/16/91, effective 10/16/91.
Statutory Authority:
RCW
43.70.040. 91-02-049 (Order 121), recodified
as § 246-338-090, filed 12/27/90, effective 1/31/91. Statutory Authority:
Chapter 70.42 RCW. 90-20-017 (Order 090), § 248-38-090, filed 9/21/90,
effective 10/22/90.