Section 9VAC5-91-700 - Calibration of exhaust gas analyzers

Universal Citation: 9 VA Admin Code 5-91-700

Current through Register Vol. 41, No. 3, September 23, 2024

The department shall use, and require for use, gases and containers meeting the following parameters, blends, and specifications in the calibration of exhaust gas analyzers:

1. The calibration gases for standardizing instruments shall conform to the provisions outlined in EPA420-B-04-011.

2. Gas analyzers shall pass a two-point gas calibration for HC, CO, CO2 and NO, within 72 hours before each test.

3. Gas analyzers shall pass a five-point gas audit for HC, CO, CO2, and NO at least two times within 12 months, and each time an analyzer system emissions measurement system, sensor, or other electronic components are repaired or replaced in response to an audit failure.

4. The director may reduce the frequency of calibrations in subdivision 2 of this section and audits in subdivision 3 of this section or both if evidence indicates accuracy would not be adversely affected.

Statutory Authority

§ 46.2-1180 of the Code of Virginia; § 182 of the Clean Air Act; 40 CFR Part 51, Subpart S.

Disclaimer: These regulations may not be the most recent version. Virginia may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.