Current through Register Vol. 41, No. 3, September 23, 2024
A. The provisions governing test methods and
procedures shall be as follows:
1. With
regard to the emissions standards in 9VAC5-40-8080 and 9VAC5-40-8090, the
provisions of 9VAC5-40-30 (Emission testing) apply.
2. With regard to the emission limits in
9VAC5-40-7970 through 9VAC5-40-8070, the following provisions apply:
a.9VAC5-40-30 D and G.
b.
40 CFR
60.8, with the exception of paragraph
(a).
c.
40 CFR
60.11 and
40 CFR
60.13.
d. Subsections B through J of this
section.
B.
The procedures and test methods specified in this subsection shall be used to
determine compliance with the emission limits for particulate matter and
opacity under 9VAC5-40-7970 and 9VAC5-40-8060.
1. Reference Method 1 shall be used to select
sampling site and number of traverse points.
2. Reference Method 3, 3A, or 3B, as
applicable, shall be used for gas analysis.
3. Reference Method 5 shall be used for
determining compliance with the particulate matter emission limit. The minimum
sample volume shall be 1.7 cubic meters. The probe and filter holder heating
systems in the sample train shall be set to provide a gas temperature no
greater than 160 ± 14 degrees Centigrade. An oxygen or carbon dioxide
measurement shall be obtained simultaneously with each Reference Method 5
run.
4. The owner of an affected
facility may request that compliance with the particulate matter emission limit
be determined using carbon dioxide measurements corrected to an equivalent of
7.0% oxygen. The relationship between oxygen and carbon dioxide levels for the
affected facility shall be established as specified in 9VAC5-40-8150 B
6.
5. As specified in
40 CFR
60.8, all emission tests shall consist of
three test runs. The average of the particulate matter emission concentrations
from the three test runs is used to determine compliance.
6. In accordance with subdivisions 7 and 10
of this subsection, Reference Method 9 shall be used for determining compliance
with the opacity limit except as provided in
40 CFR
60.11(e).
7. The owner of an affected facility shall
conduct an initial emission test for particulate matter emissions and opacity
as required in 9VAC5-40-8100.
8.
The owner of an affected facility shall install, calibrate, maintain, and
operate a continuous opacity monitoring system for measuring opacity and shall
follow the methods and procedures specified in subdivisions 8 a through 8 d of
this subsection.
a. The output of the
continuous opacity monitoring system shall be recorded on a 6-minute average
basis.
b. The continuous opacity
monitoring system shall be installed, evaluated, and operated in accordance
with 40 CFR
60.13.
c. The continuous opacity monitoring system
shall conform to Performance Specification 1 in Appendix B of 40 CFR Part
60.
d. The initial performance
evaluation shall be completed as specified in 9VAC5-40-8100.
9. Following the date that the
initial emission test for particulate matter is completed or is required to be
completed in 9VAC5-40-8100 for an affected facility, the owner shall conduct an
emission test for particulate matter on an annual basis (no more than 12
calendar months following the previous emission test).
10. Following the date that the initial
emission test for opacity is completed or is required to be completed in
9VAC5-40-8100 for an affected facility, the owner shall conduct an emission
test for opacity on an annual basis (no more than 12 calendar months following
the previous emission test) using the test method specified in subdivision 6 of
this subsection.
C. The
procedures and test methods specified in this subsection shall be used to
determine compliance with the emission limits for cadmium, lead, and mercury
under 9VAC5-40-7990, 9VAC5-40-8000, and 9VAC5-40-8010.
1. The procedures and test methods specified
in subdivisions 1 a through 1 g of this subsection shall be used to determine
compliance with the emission limits for cadmium and lead under 9VAC5-40-7990
and 9VAC5-40-8000.
a. Reference Method 1
shall be used for determining the location and number of sampling
points.
b. Reference Method 3, 3A,
or 3B, as applicable, shall be used for flue gas analysis.
c. Reference Method 29 shall be used for
determining compliance with the cadmium and lead emission limits.
d. An oxygen or carbon dioxide measurement
shall be obtained simultaneously with each Reference Method 29 test run for
cadmium and lead required under subdivision 1 c of this subsection.
e. The owner of an affected facility may
request that compliance with the cadmium or lead emission limit be determined
using carbon dioxide measurements corrected to an equivalent of 7.0% oxygen.
The relationship between oxygen and carbon dioxide levels for the affected
facility shall be established as specified in 9VAC5-40-8150 B 6.
f. All emission tests shall consist of a
minimum of three test runs conducted under representative full load operating
conditions. The average of the cadmium or lead emission concentrations from
three test runs or more shall be used to determine compliance.
g. Following the date of the initial emission
test or the date on which the initial emission test is required to be completed
in 9VAC5-40-8100, the owner of an affected facility shall conduct an emission
test for compliance with the emission limits for cadmium and lead on an annual
basis (no more than 12 calendar months following the previous emission
test).
2. The procedures
and test methods specified in subdivisions 2 a through 2 j of this subsection
shall be used to determine compliance with the mercury emission limit under
9VAC5-40-8010.
a. Reference Method 1 shall be
used for determining the location and number of sampling points.
b. Reference Method 3, 3A, or 3B, as
applicable, shall be used for flue gas analysis.
c. Reference Method 29 shall be used to
determine the mercury emission concentration. The minimum sample volume when
using Reference Method 29 for mercury shall be 1.7 cubic meters.
d. An oxygen (or carbon dioxide) measurement
shall be obtained simultaneously with each Reference Method 29 test run for
mercury required under subdivision 2 c of this subsection.
e. The percent reduction in the potential
mercury emissions (% pHg) is computed using the
following equation:
View Image
where:
%PHg = percent reduction of the
potential mercury emissions achieved.
Ei = potential mercury emission
concentration measured at the control device inlet, corrected to 7.0% oxygen
(dry basis).
EO = controlled mercury emission
concentration measured at the mercury control device outlet, corrected to 7.0%
oxygen (dry basis).
f. All
emission tests shall consist of a minimum of three test runs conducted under
representative full load operating conditions. The average of the mercury
emission concentrations or percent reductions from three test runs or more is
used to determine compliance.
g.
The owner of an affected facility may request that compliance with the mercury
emission limit be determined using carbon dioxide measurements corrected to an
equivalent of 7.0% oxygen. The relationship between oxygen and carbon dioxide
levels for the affected facility shall be established as specified in
9VAC5-40-8150 B 6.
h. The owner of
an affected facility shall conduct an initial emission test for mercury
emissions as required in 9VAC5-40-8100.
i. Following the date that the initial
emission test for mercury is completed or is required to be completed in
9VAC5-40-8100, the owner of an affected facility shall conduct an emission test
for mercury emissions on an annual basis (no more than 12 calendar months from
the previous emission test).
j. The
owner of an affected facility where activated carbon injection is used to
comply with the mercury emission limit shall follow the procedures specified in
subsection J of this section for measuring and calculating carbon
usage.
D. The
procedures and test methods specified in this subsection shall be used for
determining compliance with the sulfur dioxide emission limit under
9VAC5-40-8020.
1. Reference Method 19,
section 4.3, shall be used to calculate the daily geometric average sulfur
dioxide emission concentration.
2.
Reference Method 19, section 5.4, shall be used to determine the daily
geometric average percent reduction in the potential sulfur dioxide emission
concentration.
3. The owner of an
affected facility may request that compliance with the sulfur dioxide emission
limit be determined using carbon dioxide measurements corrected to an
equivalent of 7.0% oxygen. The relationship between oxygen and carbon dioxide
levels for the affected facility shall be established as specified in
9VAC5-40-8150 B 6.
4. The owner of
an affected facility shall conduct an initial emission test for sulfur dioxide
emissions as required in 9VAC5-40-8100. Compliance with the sulfur dioxide
emission limit (concentration or percent reduction) shall be determined by
using the continuous emission monitoring system specified in subdivision 5 of
this subsection to measure sulfur dioxide and calculating a 24-hour daily
geometric average emission concentration or a 24-hour daily geometric average
percent reduction using Reference Method 19, sections 4.3 and 5.4, as
applicable.
5. The owner of an
affected facility shall install, calibrate, maintain, and operate a continuous
emission monitoring system for measuring sulfur dioxide emissions discharged to
the atmosphere and record the output of the system.
6. Following the date that the initial
emission test for sulfur dioxide is completed or is required to be completed in
9VAC5-40-8100, compliance with the sulfur dioxide emission limit shall be
determined based on the 24-hour daily geometric average of the hourly
arithmetic average emission concentrations using continuous emission monitoring
system outlet data if compliance is based on an emission concentration, or
continuous emission monitoring system inlet and outlet data if compliance is
based on a percent reduction.
7. At
a minimum, valid continuous monitoring system hourly averages shall be obtained
as specified in subdivisions 7 a and 7 b of this subsection for 75% of the
operating hours per day for 90% of the operating days per calendar quarter that
the affected facility is combusting municipal solid waste.
a. At least two data points per hour shall be
used to calculate each one-hour arithmetic average.
b. Each sulfur dioxide one-hour arithmetic
average shall be corrected to 7.0% oxygen on an hourly basis using the one-hour
arithmetic average of the oxygen (or carbon dioxide) continuous emission
monitoring system data.
8. The one-hour arithmetic averages required
under subdivision 6 of this subsection shall be expressed in parts per million
corrected to 7.0% oxygen (dry basis) and used to calculate the 24-hour daily
geometric average emission concentrations and daily geometric average emission
percent reductions. The one-hour arithmetic averages shall be calculated using
the data points required in 40 CFR
60.13(e)(2).
9. All valid continuous emission monitoring
system data shall be used in calculating average emission concentrations and
percent reductions even if the minimum continuous emission monitoring system
data requirements of subdivision 7 of this subsection are not met.
10. The procedures in
40 CFR
60.13 shall be followed for installation,
evaluation, and operation of the continuous emission monitoring
system.
11. The initial performance
evaluation shall be completed as specified in 9VAC5-40-8100.
12. The continuous emission monitoring system
shall be operated according to Performance Specification 2 in Appendix B of 40
CFR Part 60.
a. During each relative accuracy
test run of the continuous emission monitoring system required by Performance
Specification 2 in Appendix B of 40 CFR Part 60, sulfur dioxide and oxygen (or
carbon dioxide) data shall be collected concurrently (or within a 30- to
60-minute period) by both the continuous emission monitors and the test methods
specified in subdivisions 12 a (1) and 12 a (2) of this subsection.
(1) For sulfur dioxide, Reference Method 6,
6A, or 6C shall be used.
(2) For
oxygen (or carbon dioxide), Reference Method 3, 3A, or 3B, as applicable, shall
be used.
b. The span
value of the continuous emissions monitoring system at the inlet to the sulfur
dioxide control device shall be 125% of the maximum estimated hourly potential
sulfur dioxide emissions of the municipal waste combustor unit. The span value
of the continuous emission monitoring system at the outlet of the sulfur
dioxide control device shall be 50% of the maximum estimated hourly potential
sulfur dioxide emissions of the municipal waste combustor unit.
13. Quarterly accuracy
determinations and daily calibration drift tests shall be performed in
accordance with procedure 1 in Appendix F of 40 CFR Part 60.
14. When sulfur dioxide emissions data are
not obtained because of continuous emission monitoring system breakdowns,
repairs, calibration checks, and zero and span adjustments, emissions data
shall be obtained by using other monitoring systems as approved by the board or
Reference Method 19 to provide, as necessary, valid emissions data for a
minimum of 75% of the hours per day that the affected facility is operated and
combusting municipal solid waste for 90% of the days per calendar quarter that
the affected facility is operated and combusting municipal solid
waste.
E. The procedures
and test methods specified in this subsection shall be used for determining
compliance with the hydrogen chloride emission limit under 9VAC5-40-8030.
1. Reference Method 26 or 26A, as applicable,
shall be used to determine the hydrogen chloride emission concentration. The
minimum sampling time for Reference Method 26 shall be one hour.
2. An oxygen (or carbon dioxide) measurement
shall be obtained simultaneously with each Reference Method 26 test run for
hydrogen chloride required by subdivision 1 of this subsection.
3. The percent reduction in potential
hydrogen chloride emissions (%PHCl) is computed using
the following equation:
View Image
where:
%PHCl = percent reduction of the
potential hydrogen chloride emissions achieved.
Ei = potential hydrogen chloride
emission concentration measured at the control device inlet, corrected to 7.0%
oxygen (dry basis).
EO = controlled hydrogen chloride
emission concentration measured at the control device outlet, corrected to 7.0%
oxygen (dry basis).
4. The
owner of an affected facility may request that compliance with the hydrogen
chloride emission limit be determined using carbon dioxide measurements
corrected to an equivalent of 7.0% oxygen. The relationship between oxygen and
carbon dioxide levels for the affected facility shall be established as
specified in 9VAC5-40-8150 B 6.
5.
As specified in 40 CFR
60.8, all emission tests shall consist of
three test runs. The average of the hydrogen chloride emission concentrations
or percent reductions from the three test runs is used to determine
compliance.
6. The owner of an
affected facility shall conduct an initial emission test for hydrogen chloride
as required in 9VAC5-40-8100.
7.
Following the date that the initial emission test for hydrogen chloride is
completed or is required to be completed in 9VAC5-40-8100, the owner of an
affected facility shall conduct an emission test for hydrogen chloride
emissions on an annual basis (no more than 12 calendar months following the
previous emission test).
F. The procedures and test methods specified
in this subsection shall be used to determine compliance with the limits for
dioxin/furan emissions under 9VAC5-40-8040.
1. Reference Method 1 shall be used for
determining the location and number of sampling points.
2. Reference Method 3, 3A, or 3B, as
applicable, shall be used for flue gas analysis.
3. Reference Method 23 shall be used for
determining the dioxin/furan emission concentration.
a. The minimum sample time shall be four
hours per test run.
b. An oxygen
(or carbon dioxide) measurement shall be obtained simultaneously with each
Reference Method 23 test run for dioxins/furans.
4. The owner of an affected facility shall
conduct an initial emission test for dioxin/furan emissions in accordance with
subdivision 3 of this subsection, as required in 9VAC5-40-8100.
5. Following the date that the initial
emission test for dioxins/furans is completed or is required to be completed in
9VAC5-40-8100, the owner of an affected facility shall conduct emission tests
for dioxin/furan emissions in accordance with subdivision 3 of this subsection,
according to one of the schedules specified in subdivision 5 a or 5 b of this
subsection.
a. For affected facilities,
emission tests shall be conducted on an annual basis (no more than 12 calendar
months following the previous emission test).
b. Where all emission tests over a two-year
period indicate that dioxin/furan emissions are less than or equal to 15
nanograms per dry standard cubic meter (total mass) for all affected facilities
located within a municipal waste combustor plant, the owner of the municipal
waste combustor plant may elect to conduct annual emission tests for one
affected facility (i.e., unit) per year at the municipal waste combustor plant.
At a minimum, an emission test for dioxin/furan emissions shall be conducted
annually (no more than 12 months following the previous emission test) for one
affected facility at the municipal waste combustor plant. Each year a different
affected facility at the municipal waste combustor plant shall be tested, and
the affected facilities at the plant shall be tested in sequence (e.g., unit 1,
unit 2, unit 3, as applicable). If each annual emission test continues to
indicate a dioxin/furan emission level less than or equal to 15 nanograms per
dry standard cubic meter (total mass), the owner may continue conducting an
emission test on only one affected facility per year. If any annual emission
test indicates a dioxin/furan emission level greater than 15 nanograms per dry
standard cubic meter (total mass), emission tests thereafter shall be conducted
annually on all affected facilities at the plant until and unless all annual
emission tests for all affected facilities at the plant over a two-year period
indicate a dioxin/furan emission level less than or equal to 15 nanograms per
dry standard cubic meter (total mass).
6. The owner of an affected facility that
selects to follow the emission testing schedule specified in subdivision 5 a or
5 b of this subsection shall follow the procedures specified in 9VAC5-40-8160 D
4 for reporting the selection of this schedule.
7. The owner of an affected facility where
activated carbon is used to comply with the dioxin/furan emission limits
specified in 9VAC5-40-8040 or the dioxin/furan emission level specified in
subdivision 5 a or 5 b of this subsection shall follow the procedures specified
in subsection J of this section for measuring and calculating the carbon usage
rate.
8. The owner may request that
compliance with the dioxin/furan emission limit be determined using carbon
dioxide measurements corrected to an equivalent of 7.0% oxygen. The
relationship between oxygen and carbon dioxide levels for the affected facility
shall be established as specified in 9VAC5-40-8150 B 6.
9. As specified in
40 CFR
60.8, all emission tests shall consist of
three test runs. The average of the dioxin/furan emission concentrations from
the three test runs is used to determine compliance.
G. The procedures and test methods specified
in this subsection shall be used to determine compliance with the nitrogen
oxides emission limit for affected facilities under 9VAC5-40-8050.
1. Reference Method 19, section 4.1, shall be
used for determining the daily arithmetic average nitrogen oxides emission
concentration.
2. The owner of an
affected facility may request that compliance with the nitrogen oxides emission
limit be determined using carbon dioxide measurements corrected to an
equivalent of 7.0% oxygen. The relationship between oxygen and carbon dioxide
levels for the affected facility shall be established as specified in
9VAC5-40-8150 B 6.
3. The owner of
an affected facility subject to the nitrogen oxides limit under 9VAC5-40-8050
shall conduct an initial emission test for nitrogen oxides as required in
9VAC5-40-8100. Compliance with the nitrogen oxides emission limit shall be
determined by using the continuous emission monitoring system specified in
subdivision 4 of this subsection for measuring nitrogen oxides and calculating
a 24-hour daily arithmetic average emission concentration using Reference
Method 19, section 4.1.
4. The
owner of an affected facility subject to the nitrogen oxides emission limit
under 9VAC5-40-8050 shall install, calibrate, maintain, and operate a
continuous emission monitoring system for measuring nitrogen oxides discharged
to the atmosphere, and record the output of the system.
5. Following the date that the initial
emission test for nitrogen oxides is completed or is required to be completed
in 9VAC5-40-8100, compliance with the emission limit for nitrogen oxides
required under 9VAC5-40-8050 shall be determined based on the 24-hour daily
arithmetic average of the hourly emission concentrations using continuous
emission monitoring system outlet data.
6. At a minimum, valid continuous emission
monitoring system hourly averages shall be obtained as specified in
subdivisions 6 a and 6 b of this subsection for 75% of the operating hours per
day for 90% of the operating days per calendar quarter that the affected
facility is combusting municipal solid waste.
a. At least two data points per hour shall be
used to calculate each one-hour arithmetic average.
b. Each nitrogen oxides one-hour arithmetic
average shall be corrected to 7.0% oxygen on an hourly basis using the one-hour
arithmetic average of the oxygen (or carbon dioxide) continuous emission
monitoring system data.
7. The one-hour arithmetic averages required
by subdivision 5 of this subsection shall be expressed in parts per million by
volume (dry basis) and used to calculate the 24-hour daily arithmetic average
concentrations. The one-hour arithmetic averages shall be calculated using the
data points required in 40
CFR 60.13(e)(2).
8. All valid continuous emission monitoring
system data must be used in calculating emission averages even if the minimum
continuous emission monitoring system data requirements of subdivision 6 of
this subsection are not met.
9. The
procedures in 40 CFR
60.13 shall be followed for installation,
evaluation, and operation of the continuous emission monitoring system. The
initial performance evaluation shall be completed as specified in
9VAC5-40-8100.
10. The owner of an
affected facility shall operate the continuous emission monitoring system
according to Performance Specification 2 in Appendix B of 40 CFR Part 60 and
shall follow the procedures and methods specified in subdivisions 10 a and 10 b
of this subsection.
a. During each relative
accuracy test run of the continuous emission monitoring system required by
Performance Specification 2 of Appendix B of 40 CFR Part 60, nitrogen oxides
and oxygen (or carbon dioxide) data shall be collected concurrently (or within
a 30- to 60-minute period) by both the continuous emission monitors and the
test methods specified in subdivisions 10 a (1) and 10 a (2) of this
subsection.
(1) For nitrogen oxides,
Reference Method 7, 7A, 7C, 7D, or 7E shall be used.
(2) For oxygen (or carbon dioxide), Reference
Method 3, 3A, or 3B, as applicable, shall be used.
b. The span value of the continuous emission
monitoring system shall be 125% of the maximum estimated hourly potential
nitrogen oxide emissions of the municipal waste combustor unit.
11. Quarterly accuracy
determinations and daily calibration drift tests shall be performed in
accordance with procedure 1 in Appendix F of 40 CFR Part 60.
12. When nitrogen oxides continuous emissions
data are not obtained because of continuous emission monitoring system
breakdowns, repairs, calibration checks, and zero and span adjustments,
emissions data shall be obtained using other monitoring systems as approved by
the board or Reference Method 19 to provide, as necessary, valid emissions data
for a minimum of 75% of the hours per day for 90% of the days per calendar
quarter the unit is operated and combusting municipal solid waste.
H. The procedures specified in
this subsection shall be used for determining compliance with the fugitive ash
emission limit under 9VAC5-40-8070.
1.
Reference Method 22 shall be used for determining compliance with the fugitive
ash emission limit under 9VAC5-40-8070. The minimum observation time shall be a
series of three one-hour observations. The observation period shall include
times when the facility is transferring ash from the municipal waste combustor
unit to the area where ash is stored or loaded into containers or
trucks.
2. The average duration of
visible emissions per hour shall be calculated from the three one-hour
observations. The average shall be used to determine compliance with
9VAC5-40-8070.
3. The owner of an
affected facility shall conduct an initial emission test for fugitive ash
emissions as required in 9VAC5-40-8100.
4. Following the date that the initial
emission test for fugitive ash emissions is completed or is required to be
completed in 9VAC5-40-8100 for an affected facility, the owner shall conduct an
emission test for fugitive ash emissions on an annual basis (no more than 12
calendar months following the previous emission test).
I. The procedures specified in this
subsection shall be used to determine compliance with the opacity limit for air
curtain incinerators under 9VAC5-40-8060 C.
1. Reference Method 9 shall be used for
determining compliance with the opacity limit.
2. The owner of the air curtain incinerator
shall conduct an initial emission test for opacity as required in
9VAC5-40-8100.
3. Following the
date that the initial emission test is completed or is required to be completed
in 9VAC5-40-8100, the owner of the air curtain incinerator shall conduct an
emission test for opacity on an annual basis (no more than 12 calendar months
following the previous emission test).
J. The owner of an affected facility where
activated carbon injection is used to comply with the mercury emission limit
under 9VAC5-40-8010, or the dioxin/furan emission limits under 9VAC5-40-8040,
or the dioxin/furan emission level specified in subdivision F 5 b of this
section shall follow the procedures specified in this subsection.
1. During the emission tests for
dioxins/furans and mercury, as applicable, the owner shall estimate an average
carbon mass feed rate based on carbon injection system operating parameters
such as the screw feeder speed, hopper volume, hopper refill frequency, or
other parameters appropriate to the feed system being employed, as specified in
subdivisions 1 a and 1 b of this subsection.
a. An average carbon mass feed rate in
kilograms per hour or pounds per hour shall be estimated during the initial
emission test for mercury emissions and each subsequent emission test for
mercury emissions.
b. An average
carbon mass feed rate in kilograms per hour or pounds per hour shall be
estimated during the initial emission test for dioxin/furan emissions and each
subsequent emission test for dioxin/furan emissions.
2. During operation of the affected facility,
the carbon injection system operating parameter(s) that are the primary
indicator(s) of the carbon mass feed rate (e.g., screw feeder setting) must
equal or exceed the level(s) documented during the emission tests specified
under subdivisions 1 a and 1 b of this subsection.
3. The owner of an affected facility shall
estimate the total carbon usage of the plant (kilograms or pounds) for each
calendar quarter by two independent methods, according to the procedures in
subdivisions 3 a and 3 b of this subsection.
a. The weight of carbon delivered to the
plant.
b. Estimate the average
carbon mass feed rate in kilograms per hour or pounds per hour for each hour of
operation for each affected facility based on the parameters specified under
subdivision 1 of this subsection, and sum the results for all affected
facilities at the plant for the total number of hours of operation during the
calendar quarter.
Statutory Authority
§ 10.1-1308 of the Code of Virginia.