Utah Administrative Code
Topic - Environmental Quality
Title R313 - Waste Management and Radiation Control, Radiation
Rule R313-28 - Use of X-Rays in the Healing Arts
Section R313-28-57 - Radiographic Systems Other Than Fluoroscopic, or Dental Intraoral Systems - mA/mAs Linearity
Current through Bulletin 2024-06, March 15, 2024
The following requirements apply when the equipment is operated on a power supply as specified by the manufacturer for fixed x-ray tube potentials within the range of 40 percent to 100 percent of the maximum rated potentials.
(1) Equipment having independent selection of x-ray tube current, mA. Where the tube current is continuous, the average ratios of exposure to the indicated milliampere-seconds product, C/kg/mAs or mR/mAs, obtained at two consecutive tube current settings or at two settings differing by no more than a factor of two shall not differ by more than 0.10 times their sum.
(2) Equipment having a combined x-ray tube current-exposure time product, mAs, selector, but not a separate tube current, mA, selector. Where the tube current is continuous, the average ratios of exposure to the indicated milliampere-seconds product, C/kg/mAs or mR/mAs, obtained at two consecutive milliampere-seconds settings or at two settings differing by no more than a factor of two shall not differ by more than 0.10 times their sum.