Rule R313-28 - Use of X-Rays in the Healing Arts
Section R313-28-55 - Radiographic Systems - Standby Radiation From Capacitor Discharge Equipment

Universal Citation: UT Admin Code R 313-28-55
Current through Bulletin 2024-06, March 15, 2024

Radiation emitted from the x-ray tube when the system is fully charged and the exposure switch or timer is not activated shall not exceed a rate of 0.516 uC/kg (two milliroentgens) per hour at five centimeters from accessible surfaces of the diagnostic source assembly, with the beam-limiting device fully open.

Disclaimer: These regulations may not be the most recent version. Utah may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.