Rule R313-28 - Use of X-Rays in the Healing Arts
Section R313-28-54 - Radiographic Systems Other Than Fluoroscopic, or Dental Intraoral Systems - Exposure Reproducibility

Universal Citation: UT Admin Code R 313-28-54
Current through Bulletin 2024-06, March 15, 2024

When technique factors, including control panel selections associated with automatic exposure control systems, are held constant the coefficient of variation of exposure for both manual and automatic exposure control systems shall not exceed 0.05. This requirement applies to clinically used techniques.

Disclaimer: These regulations may not be the most recent version. Utah may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.