Utah Administrative Code
Topic - Environmental Quality
Title R313 - Waste Management and Radiation Control, Radiation
Rule R313-28 - Use of X-Rays in the Healing Arts
Section R313-28-52 - Radiographic Systems Other Than Fluoroscopic, Dental Intraoral, or Computed Tomography - Radiation Exposure Control Devices

Universal Citation: UT Admin Code R 313-28-52

Current through Bulletin 2024-06, March 15, 2024

(1) Exposure Initiation. Means shall be provided to initiate the radiation exposure by a deliberate action on the part of the operator, for example, the depression of a switch. Radiation exposure shall not be initiated without a deliberate action. In addition, it shall not be possible to initiate an exposure when the timer is set to a "zero" or "off" position if either position is provided.

(2) Exposure termination. Means shall be provided to terminate the exposure at a preset time interval, preset product of current and time, a preset number of pulses, or a preset radiation exposure to the image receptor. Except for dental panoramic systems, termination of an exposure shall cause automatic resetting of the timer to its initial setting or to "zero."

(3) Manual Exposure Control: An x-ray control shall be incorporated into x-ray systems so that an exposure can be terminated at times except for:

(a) exposure of one-half second or less; or

(b) during serial radiography when means shall be provided to permit completion of a single exposure of the series in process.

(4) Automatic EXPOSURE controls, phototimers. When automatic EXPOSURE control is provided:

(a) indication shall be made on the control panel when this mode of operation is selected;

(b) when the x-ray tube potential is equal to or greater than 51 kVp, the minimum exposure time for field emission equipment rated for pulsed operation shall be equal to or less than the interval equivalent to two pulses; and

(c) the minimum exposure time for all equipment other than that specified in R313-28-52(4)(b) shall be equal to or less than 1/60 second or a time interval required to deliver five mAs, whichever is greater.

(5) Exposure Indication. Means shall be provided for visual indication observable at or from the operator's protected position whenever x-rays are produced. In addition, a signal audible to the operator shall indicate that the exposure has terminated.

(6) Exposure Duration, Timer, Linearity. For systems having independent selection of exposure time settings, the average ratio of exposure to the indicated milliampere-seconds product obtained at two consecutive timer settings or at two settings not differing by more than a factor of two shall not differ by more than 0.10 times their sum.

(7) Exposure Control Location. The x-ray exposure control shall be placed so that the operator can view the patient while making the exposure.

(8) Operator Protection.

(a) Stationary x-ray systems shall be required to have the x-ray exposure switch permanently mounted in a protected area.

(b) Mobile and portable x-ray systems which are:
(i) used continuously for greater than one week at the same location, one room or suite, shall meet the requirements of R313-28-52(8)(a); or

(ii) used for less than one week at one location, one room, or suite shall be provided with either a protective barrier at least two meters (6.5 feet) high for operator protection during exposures, or means shall be provided to allow the operator to be at least 2.7 meters (nine feet) from the tube housing assembly during the exposure.

Disclaimer: These regulations may not be the most recent version. Utah may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.