Utah Administrative Code
Topic - Environmental Quality
Title R313 - Waste Management and Radiation Control, Radiation
Rule R313-28 - Use of X-Rays in the Healing Arts
Section R313-28-51 - Radiographic Systems Other than Fluoroscopic, Dental Intraoral, or Computed Tomography - Beam Limitation

Universal Citation: UT Admin Code R 313-28-51

Current through Bulletin 2024-06, March 15, 2024

The useful beam shall be limited to the area of clinical interest and show evidence of collimation. This shall be deemed to have been met if a positive beam limiting device meeting the manufacturer's specifications or the requirements of R313-28-300 has been properly used or if evidence of collimation is shown on at least three sides or three corners of the film, for example, projections of the shutters of the collimator, cone cutting at the corners or a border at the film's edge.

(1) General purpose stationary and mobile x-ray systems.

(a) Only x-ray systems provided with a means for independent stepless adjustment of at least two dimensions of the x-ray field shall be used.

(b) A method shall be provided for visually defining the perimeter of the x-ray field. The total misalignment of the edges of the visually defined field with the respective edges of the x-ray field along either the length or width of the visually defined field shall not exceed two percent of the distance from the source to the center of the visually defined field when the surface upon which it appears is perpendicular to the axis of the x-ray beam.

(c) The Board may grant an exemption on non-certified x-ray systems to R313-28-51(1)(a) and (b) provided the registrant makes a written application for the exemption and in that application:
(i) demonstrates it is impractical to comply with R313-28-51(1)(a) and (b); and

(ii) demonstrates the purpose of R313-28-51(1)(a) and (b) will be met by other methods.

(2) In addition to the requirements of R313-28-51(1) above, stationary general purpose x-ray systems, both certified and non-certified shall meet the following requirements:

(a) a method shall be provided to indicate when the axis of the x-ray beam is perpendicular to the plane of the image receptor, to align the center of the x-ray field with respect to the center of the image receptor to within two percent of the SID, and to indicate the SID to within two percent;

(b) the beam-limiting device shall numerically indicate the field size in the plane of the image receptor to which it is adjusted; and

(c) indication of field size dimensions and SID's shall be specified in inches or centimeters and shall be such that aperture adjustments result in x-ray field dimensions in the plane of the image receptor which correspond to those of the image receptor to within two percent of the SID when the beam axis is perpendicular to the plane of the image receptor.

(3) Radiographic equipment designed for only one image receptor size at a fixed SID shall be provided with means to limit the field at the plane of the image receptor to dimensions no greater than those of the image receptor, and to align the center of the x-ray field with the center of the image receptor to within two percent of the SID, or shall be provided with means to both size and align the x-ray field so that the x-ray field at the plane of the image receptor does not extend beyond the edges of the image receptor.

(4) Special purpose x-ray systems.

(a) Means shall be provided to limit the x-ray field in the plane of the image receptor so that the x-ray field does not exceed each dimension of the image receptor by more than two percent of the SID when the axis of the x-ray beam is perpendicular to the plane of the image receptor.

(b) Means shall be provided to align the center of the x-ray field with the center of the image receptor to within two percent of the SID, or means shall be provided to both size and align the x-ray field so that the x-ray field at the plane of the image receptor does not extend beyond the edges of the image receptor. Compliance shall be determined with the axis of the x-ray beam perpendicular to the plane of the image receptor.

(c)R313-28-51(4)(a) and R313-28-51(4)(b) may be met with a system that meets the requirements for a general purpose x-ray system as specified in R313-28-51(1) or, when alignment means are also provided, may be met with either;
(i) an assortment of removable, fixed-aperture, beam-limiting devices sufficient to meet the requirements for the combination of image receptor sizes and SID's for which the unit is designed with the beam limiting device having clear and permanent markings to indicate the image receptor size and SID for which it is designed; or

(ii) a beam-limiting device having multiple fixed apertures sufficient to meet the requirement for the combinations of image receptor sizes and SID's for which the unit is designed. Permanent, clearly legible markings shall indicate the image receptor size and SID for which the aperture is designed and shall indicate which aperture is in position for use.

Disclaimer: These regulations may not be the most recent version. Utah may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.