Utah Administrative Code
Topic - Environmental Quality
Title R313 - Waste Management and Radiation Control, Radiation
Rule R313-28 - Use of X-Rays in the Healing Arts
Section R313-28-40 - Fluoroscopic X-Ray Systems

Universal Citation: UT Admin Code R 313-28-40

Current through Bulletin 2024-06, March 15, 2024

All fluoroscopic x-ray systems used shall be image intensified and meet the following requirements:

(1) Primary barrier.

(a) The fluoroscopic imaging assembly shall be provided with a primary protective barrier which intercepts the entire cross section of the useful beam at SIDs for which the unit was designed.

(b) The x-ray tube used for fluoroscopy shall not produce x-rays unless the barrier is in position to intercept the entire useful beam.

(2) Fluoroscopic beam limitation.

(a) For certified fluoroscopic systems with or without a spot film device neither the length nor the width of the x-ray field in the plane of the image receptor shall exceed that of the visible area of the image receptor by more than three percent of the SID. The sum of the excess length and the excess width shall be no greater than four percent of the SID.

(b) For uncertified fluoroscopic systems with a spot film device, the x-ray beam with the shutters fully open, during fluoroscopy or spot filming, shall be no larger than the largest image receptor size for which the device is designed. Measurements shall be made at the minimum SID available but at no less than 20 centimeters table top to the film plane distance.

(c) For uncertified fluoroscopic systems without a spot film device, the requirements of R313-28-40(1) apply.

(d) Other requirements for fluoroscopic beam limitation:
(i) means shall be provided to permit further limitation of the field. Beam-limiting devices manufactured after May 22, 1979, and incorporated in equipment with a variable SID or visible area of greater than 300 square centimeters shall be provided with means for stepless adjustment of the x-ray field;

(ii) equipment with a fixed SID and a visible area of 300 square centimeters or less shall be provided with either stepless adjustment of the x-ray field or with means to further limit the x-ray field size at the plane of the image receptor to 125 square centimeters or less;

(iii) if provided, stepless adjustment shall at the greatest SID, provide continuous field sizes from the maximum attainable to a field size of five centimeters by five centimeters or less;

(iv) for equipment manufactured after February 25, 1978, when the angle between the image receptor and beam axis is variable, means shall be provided to indicate when the axis of the x-ray beam is perpendicular to the plane of the image receptor; and

(v) for non-circular x-ray fields used with circular image receptors, the error in alignment shall be determined along the length and width dimensions of the x-ray field which pass through the center of the visible area of the image receptor.

(3) Spot-film beam limitation. Spot-film devices shall meet the following requirements:

(a) means shall be provided between the source and the patient for adjustment of the x-ray field size in the plane of the film to the size of that portion of the film which has been selected on the spot film selector. Adjustments shall be automatically accomplished except when the x-ray field size in the plane of the film is smaller than that of the selected portion of the film. For spot film devices manufactured after June 21, 1979, if the x-ray field size is less than the size of the selected portion of the film, the means for adjustment of the field size shall be only at the operator's option;

(b) neither the length nor the width of the x-ray field in the plane of the image receptor shall differ from the corresponding dimensions of the selected portion of the image receptor by more than three percent of the SID when adjusted for full coverage of the selected portion of the image receptor. The sum, without regard to sign, of the length and width differences shall not exceed four percent of the SID;

(c) it shall be possible to adjust the x-ray field size in the plane of the film to a size smaller than the selected portion of the film. The minimum field size at the greatest SID shall be equal to, or less than, five by five centimeters;

(d) the center of the x-ray field in the plane of the film shall be aligned with the center of the selected portion of the film to within two percent of the SID; and

(e) on spot film devices manufactured after February 25, 1978, if the angle between the plane of the image receptor and beam axis is variable, means shall be provided to indicate when the axis of the x-ray beam is perpendicular to the plane of the image receptor, and compliance shall be determined with the beam axis indicated to be perpendicular to the plane of the image receptor.

(4) Override. If a means exists to override the automatic x-ray field size adjustments required in R313-28-40(2) and (3), that means:

(a) shall be designed for use only in the event of system failure;

(b) shall incorporate a signal visible at the fluoroscopist's position which will indicate whenever the automatic field size adjustment is overridden; and

(c) shall be clearly and durably labeled as follows: FOR X-RAY FIELD LIMITATION SYSTEM FAILURE.

(5) Activation of the fluoroscopic tube. X-ray production in the fluoroscopic mode shall be controlled by a dead-man switch. When recording serial fluoroscopic images, the fluoroscopist shall be able to terminate the x-ray exposure immediately, but means may be provided to permit completion of a single exposure of the series in process.

(6) Entrance EXPOSURE rate allowable limits.

(a) For fluoroscopic equipment manufactured before May 19, 1995, the following requirements apply:
(i) fluoroscopic equipment which is provided with automatic exposure rate control shall not be operable at combinations of tube potential and current which will result in an EXPOSURE rate in excess of 2.58 mC/kg (ten roentgens) per minute at the point where the center of the useful beam enters the patient, except:
(A) during recording of fluoroscopic images, or

(B) when an optional high level control is provided. When so provided, the equipment shall not be operable at combinations of tube potential and current which will result in an EXPOSURE rate in excess of 1.29 mC/kg (five roentgens) per minute at the point where the center of the useful beam enters the patient unless the high level control is activated. Special means of activation of high level controls shall be required. The high level control shall be operable only when continuous manual activation is provided by the operator. A continuous signal audible to the fluoroscopist shall indicate that the high level control is being employed.

(ii) fluoroscopic equipment which is not provided with automatic exposure rate control shall not be operable at combinations of tube potential and current which will result in a EXPOSURE rate in excess of 1.29 mC/kg (five roentgens) per minute at the point where the center of the useful beam enters the patient, except:
(A) during recording of fluoroscopic images, or

(B) when an optional high level control is activated. Special means of activation of high level controls shall be required. The high level control shall be operable only when continuous manual activation is provided by the operator. A continuous signal audible to the fluoroscopist shall indicate that the high level control is being employed.

(iii) fluoroscopic equipment which is provided with both automatic exposure rate control and a manual mode shall not be operable at combinations of tube potential and current that will result in an exposure rate of 2.58 mC/kg (ten roentgens) per minute in either mode at the point where the center of the useful beam enters the patient except:
(A) during recording of fluoroscopic images, or

(B) when an optional high level control is provided. When so provided, the equipment shall not be operable at combinations of tube potential and current which will result in an EXPOSURE rate in excess of 1.29 mC/kg (five roentgens) per minute at the point where the center of the useful beam enters the patient unless the high level control is activated. Special means of activation of high level controls shall be required. The high level control shall be operable only when continuous manual activation is provided by the operator. A continuous signal audible to the fluoroscopist shall indicate that the high level control is being employed.

(b) For fluoroscopic equipment manufactured on and after May 19, 1995, the following requirements apply:
(i) fluoroscopic equipment operable at combinations of tube potential and current which will result in an EXPOSURE rate greater than 1.29 mC/kg (five roentgens) per minute at the point where the center of the useful beam enters the patient shall be equipped with automatic exposure rate control. Provision for manual selection of technique factors may be provided.

(ii) fluoroscopic equipment shall not be operable at combinations of tube potential and current which will result in an EXPOSURE rate in excess of 2.58 mC/kg (ten roentgens) per minute at the point where the center of the useful beam enters the patient except:
(A) during recording of images from an x-ray image-intensifier tube using photographic film or a video camera when the x-ray source is operated in pulsed mode, or

(B) when an optional high level control is activated. When the high level control is activated, the equipment shall not be operable at combinations of tube potential and current which will result in an EXPOSURE rate in excess of 5.16 mC/kg (20 roentgens) per minute at the point where the center of the useful beam enters the patient. Special means of activation of high level controls shall be required. The high level control shall be operable only when continuous manual activation is provided by the operator. A continuous signal audible to the fluoroscopist shall indicate that the high level control is being employed.

(c) Compliance with the requirements of R313-28-40(6) shall be determined as follows:
(i) if the source is below the x-ray table, the EXPOSURE rate shall be measured one centimeter above the tabletop or cradle;

(ii) if the source is above the x-ray table, the EXPOSURE rate shall be measured at 30 centimeters above the tabletop with the end of the beam-limiting device or spacer positioned as closely as possible to the point of measurement;

(iii) for a C-arm type of fluoroscope, the exposure rate shall be measured 30 centimeters from the input surface of the fluoroscopic imaging assembly, with the source positioned at available SID's, provided that the end of the beam-limiting device or spacer is no closer than 30 centimeters from the input surface of the fluoroscopic imaging assembly; or

(iv) for a lateral type fluoroscope, the exposure rate shall be measured at a point 15 centimeters from the centerline of the x-ray table and in the direction of the x-ray source with the end of the beam-limiting device or spacer positioned as close as possible to the point of measurement. If the tabletop is movable, it shall be positioned as close as possible to the lateral x-ray source with the end of the beam-limiting device or spacer no closer than 15 centimeters to the x-ray table.

(d) Fluoroscopic radiation therapy simulation systems are exempt from the requirements of R313-28-40(6).

(7) Measurement of entrance EXPOSURE rates shall be performed for both maximum and typical values as follows:

(a) measurements shall be made annually or after maintenance of the system which might affect the EXPOSURE rate;

(b) results of these measurements shall be posted where the fluoroscopist may have ready access to the results while using the fluoroscope and in the record required in R313-28-31(3)(b). The measurement results shall be stated in roentgens per minute and include the machine settings used in determining results. The name of the person performing the measurements and the date the measurements were performed shall be included in the results;

(c) conditions of the annual measurement of maximum entrance EXPOSURE rate shall be performed as follows:
(i) the measurement shall be made under the conditions that satisfy the requirements of R313-28-40(6)(c);

(ii) the kVp, mA, and other selectable parameters shall be adjusted to those settings which give the maximum entrance EXPOSURE rate; and

(iii) x-ray systems that incorporate automatic exposure rate control shall have sufficient attenuative material placed in the useful beam to produce the maximum output of that system; and

(d) conditions of the annual measurement of typical entrance EXPOSURE rate are as follows:
(i) the measurement shall be made under the conditions that satisfy the requirements of R313-28-40(6)(c);

(ii) the kVp, mA, and other selectable parameters shall be those settings typical of clinical use of the x-ray system; and

(iii) the x-ray system that incorporates automatic EXPOSURE rate control shall have an appropriate phantom placed in the useful beam to produce a milliamperage and kilovoltage typical of the use of the x-ray system.

(8) Barrier transmitted radiation rate limits.

(a) The EXPOSURE rate due to transmission through the primary protective barrier with the attenuation block in the useful beam, combined with radiation from the image intensifier, if provided, shall not exceed 0.516 uC/kg (two milliroentgens) per hour at ten centimeters from accessible surfaces of the fluoroscopic imaging assembly beyond the plane of the image receptor for each mC/kg (roentgen) per minute of entrance EXPOSURE rate.

(b) Measuring compliance of barrier transmission.
(i) The EXPOSURE rate due to transmission through the primary protective barrier combined with radiation from the image intensifier shall be determined by measurements averaged over an area of 100 square centimeters with no linear dimension greater than 20 centimeters.

(ii) If the source is below the tabletop, the measurement shall be made with the input surface of the fluoroscopic imaging assembly positioned 30 centimeters above the tabletop.

(iii) If the source is above the tabletop and the SID is variable, the measurement shall be made with the end of the beam-limiting device or spacer as close to the tabletop as it can be placed, provided that it shall not be closer than 30 centimeters.

(iv) Movable grids and compression devices shall be removed from the useful beam during the measurement.

(9) Indication of potential and current. During fluoroscopy and cinefluorography, x-ray tube potential and current shall be continuously indicated.

(10) Source-skin distance. The source to skin distance shall not be less than:

(a) 38 centimeters on stationary fluoroscopic systems manufactured on or after August 1, 1974;

(b) 35.5 centimeters on stationary fluoroscopic systems manufactured prior to August 1, 1974;

(c) 30 centimeters on all mobile fluoroscopes; or

(d) 20 centimeters for all mobile fluoroscopes when used for specific surgical applications.

(11) Fluoroscopic timer.

(a) Means shall be provided to preset the cumulative on-time of the fluoroscopic x-ray tube. The maximum cumulative time of the timing device shall not exceed five minutes without resetting.

(b) A signal audible to the fluoroscopist shall indicate the completion of a preset cumulative on- time. The signal shall continue to sound while x-rays are produced until the timing device is reset.

(12) Control of scatter radiation.

(a) The tables of fluoroscopic assemblies when combined with normal operating procedures shall provide protection from scatter radiation so that unprotected parts of a staff or ancillary individual's body shall not be exposed to unattenuated scattered radiation which originates from under the table. The attenuation required shall be not less than 0.25 mm lead equivalent.

(b) Equipment configuration when combined with procedures shall not allow portions of a staff member's or ancillary person's body, except the extremities, to be exposed to unattenuated scattered radiation emanating from above the tabletop unless:
(i) the radiation has passed through not less than 0.25 mm lead equivalent material including, but not limited to, drapes, bucky-slot cover panel, or self supporting curtains, in addition to the lead equivalency provided by the protective apron referred to in R313-28-31(2)(d),

(ii) that individual is at least 120 centimeters from the center of the useful beam, or

(iii) it is not feasible to attach shielding to special procedures equipment and personnel are wearing protective aprons.

(13) Spot film exposure reproducibility. Fluoroscopic systems equipped with radiographic spot film mode shall meet the exposure reproducibility requirements of R313-28-54.

(14) Radiation therapy simulation systems. Radiation therapy simulation systems shall be exempt from all the requirements R313-28-40(1), (8), and (11) provided that:

(a) the systems are designed and used in such a manner that no individual other than the patient is in the x-ray room during periods of time when the system is producing x-rays; and

(b) the systems which do not meet the requirements of R313-28-40(11) are provided with a means of indicating the cumulative time that an individual patient has been exposed to x-rays. Procedures shall require, in these cases, that the timer be reset between examinations.

Disclaimer: These regulations may not be the most recent version. Utah may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.