Section R313-28-35 - General Requirements for Diagnostic X-Ray Systems

Universal Citation: UT Admin Code R 313-28-35

Current through Bulletin 2024-06, March 15, 2024

In addition to other requirements of R313-28, all diagnostic x-ray systems shall meet the following requirements:

(1) Warning label. The control panel containing the main power switch shall bear the warning statement, legible and accessible to view: "WARNING: This x-ray unit may be dangerous to patient and operator unless safe exposure factors and operating instructions are observed."

(2) Battery charge indicator. On battery powered generators, visual means shall be provided on the control panel to indicate whether the battery is in a state of charge adequate for proper operation.

(3) Leakage radiation from the diagnostic source assembly. The leakage radiation from the diagnostic source assembly measured at a distance of one meter in any direction from the source shall not exceed 25.8 uC/kg (100 milliroentgens) in one hour when the x-ray tube is operated at its leakage technique factors.

(4) Radiation from components other than the diagnostic source assembly. The radiation emitted by a component other than the diagnostic source assembly shall not exceed 0.516 uC/kg (two milliroentgens) in one hour at five centimeters from accessible surfaces of the component when it is operated in an assembled x-ray system under the conditions for which it was designed. Compliance shall be determined by measurements averaged over an area of 100 square centimeters with no linear dimension greater than 20 centimeters.

(5) Beam quality.

(a) The half value layer of the useful beam for a given x-ray tube potential shall not be less than the values shown in R313-28-35, Table I. If it is necessary to determine such half-value layer at an x-ray tube potential which is not listed in Table I, linear interpolation or extrapolation may be made.

TABLE I

DESIGN

OPERATING

RANGE (KILO

VOLTS PEAK

MEASURED

POTENTIAL

(KILOVOLTS PEAK)

DENTAL INTRA-ORAL

MANUFACTURED BEFORE

AUGUST 1, 1974 AND ON OR AFTER

DECEMBER 1, 1980

ALL OTHER

DIAGNOSTIC

X-RAY SYSTEMS

Below 51

30

(use prohibited)

0.3

40

(use prohibited)

0.4

50

1.5

0.5

51

1.5

1.2

60

1.5

1.3

70

1.5

1.5

Above 70

71

2.1

2.1

80

2.3

2.3

90

2.5

2.5

100

2.7

2.7

110

3.0

3.0

120

3.2

3.2

130

3.5

3.5

140

3.8

3.8

150

4.1

4.1

(b) For capacitor discharge equipment, compliance with the requirements of R313-28-35(5)(a) shall be determined with the system fully charged and a setting of 10 mAs for exposures.

(c) The required minimal half-value layer of the useful beam shall include the filtration contributed by materials which are permanently present between the focal spot of the tube and the patient.

(d) Filtration control. For x-ray systems which have variable kVp and variable filtration for the useful beam, a device shall link the kVp selector with the filters and shall prevent an exposure unless the minimum amount of filtration necessary to produce the HVL required by R313-28-35(5)(a) is in the useful beam for the given kVp which has been selected.

(6) Multiple tubes. When two or more radiographic tubes are controlled by one exposure switch, the tube or tubes which have been selected shall be clearly indicated prior to initiation of the exposure. For equipment manufactured after August 1, 1974, indications shall be both on the x-ray control panel and at or near the tube housing assembly which has been selected.

(7) Mechanical support of tube head. The tube housing assembly supports shall be adjusted so that the tube housing assembly will remain stable during an exposure unless the tube housing movement during exposure is a designed function of the x-ray system.

(8) Technique indicators.

(a) The technique factors to be used during an exposure shall be indicated before the exposure begins, except when automatic EXPOSURE controls are used, in which case the technique factors which are set prior to the exposure shall be indicated.

(b) On equipment having fixed technique factors, the requirements, in R313-28-35(8)(a) may be met by permanent markings. Indication of technique factors shall be visible from the operator's position except in the case of spot films made by the fluoroscopist.

(9) Maintaining compliance. Diagnostic x-ray systems and their associated components certified pursuant to the provisions of 21 CFR Part 1020(2006) shall be maintained in compliance with applicable requirements of that standard.

(10) Locks. All position locking, holding, and centering devices on x-ray system components and systems shall function as intended.

(11) X-ray systems which have been granted a variance by the Director, Center for Devices and Radiological Health, Food and Drug Administration (Director), from the performance standards for ionizing radiation emitting products, in accordance with 21 CFR 1010.4(2006) shall be deemed to satisfy the requirements in R313-28 that correspond to the variance granted by the Director. The registrant shall insure that labeling pursuant to 21 CFR 1010.5(f) (2006) remains legible and visible on the x-ray system.

Disclaimer: These regulations may not be the most recent version. Utah may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.