Rule R313-28 - Use of X-Rays in the Healing Arts
Section R313-28-300 - Additional Requirements Applicable to Certified Systems Only

Universal Citation: UT Admin Code R 313-28-300

Current through Bulletin 2024-06, March 15, 2024

Diagnostic x-ray systems incorporating one or more certified components shall be required to comply with the following additional requirements which relate to the certified component.

(1) Beam limitation for stationary and mobile general purpose x-ray systems.

(a) There shall be provided a means of stepless adjustment of the size of the x-ray field. The minimum field size at an SID of 100 centimeters shall be equal to or less than five centimeters by five centimeters.

(b) When a light localizer is used to define the x-ray field, it shall provide an average illumination of not less than 160 LUX (15 foot-candles) at 100 centimeters or at the maximum SID, whichever is less. The average illumination shall be based upon measurements made in the approximate center of the quadrants of the light field. Radiation therapy simulation systems are exempt from this requirement.

(2) Beam Limitation for Portable X-ray Systems. Beam limitation for portable x-ray systems shall meet the additional field limitation requirements of R313-28-51(1) or R313-28-300(1).

(3) Beam limitation and alignment on stationary general purpose x-ray systems equipped with PBL.

(a) PBL shall prevent the production of x-rays when:
(i) either the length or the width of the x-ray field in the plane of the image receptor differs, except as permitted by R313-28-300(3)(c), from the corresponding image receptor dimensions by more than three percent of the SID; or

(ii) the sum of the length and width differences as stated in R313-28-300(3)(a)(i) without regard to sign exceeds four percent of the SID.

(b) Compliance with R313-28-300(3)(a) shall be determined when the equipment indicates that the beam axis is perpendicular to the plane of the image receptor. Compliance shall be determined no sooner than five seconds after insertion of the image receptor.

(c) The PBL system shall be capable of operation, at the discretion of the operator, so that the field size at the image receptor can be adjusted to a size smaller than the image receptor through stepless adjustment of the field size. The minimum field size at a distance of 100 centimeters shall be equal to or less than five centimeters by five centimeters.

(d) The PBL system shall be designed so that if a change in image receptor does not cause an automatic return to PBL function as described in R313-28-300(3)(a), then change of the image receptor size or SID must cause the automatic return.

(4) Tube Stands for Portable X-Ray Systems. A tube stand or other mechanical support shall be used for portable x-ray systems, so that the x-ray tube housing assembly need not be hand-held during exposures.

Disclaimer: These regulations may not be the most recent version. Utah may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.