Utah Administrative Code
Topic - Environmental Quality
Title R313 - Waste Management and Radiation Control, Radiation
Rule R313-28 - Use of X-Rays in the Healing Arts
Section R313-28-160 - Computed Tomography X-ray Equipment
Universal Citation: UT Admin Code R 313-28-160
Current through Bulletin 2024-06, March 15, 2024
(1) Equipment Requirements.
(a) In the event of equipment failure
affecting data collection, means shall be provided to terminate the x-ray
exposure automatically by either de-energizing the x-ray source or intercepting
the x-ray beam with a shutter mechanism through the use of either a back-up
timer or devices which monitor equipment function.
(b) A visible signal shall indicate when the
x-ray exposure has been terminated through the means required by
R313-28-160(1)(a).
(c) The operator
shall be able to terminate the x-ray exposure at any time during a scan, or
series of scans, of greater than 0.5 second duration.
(2) Tomographic Plane Indication and Alignment.
(a) Means shall be provided to
permit visual determination of the location of a reference plane. This
reference plane can be offset from the location of the tomographic
plane.
(b) If a device using a
light source is used to satisfy R313-28-160(2)(a), the light source shall
provide illumination at levels sufficient to permit visual determination of the
location of the tomographic plane or reference plane.
(c) The total error in the indicated location
of the tomographic plane or reference plane shall not exceed 5
millimeters.
(3) Beam-On and Shutter Status Indicators.
(a) The
computed tomography (CT) x-ray control panel and CT gantry shall provide visual
indication whenever x-rays are produced and, if applicable, whether the shutter
is open or closed.
(b) Each
emergency button or switch shall be clearly labeled as to its
function.
(4) Indication of CT Conditions of Operation.
(a) The CT
x-ray system shall be designed such that technique factors, tomographic section
thickness, and scan increment shall be indicated prior to the initiation of a
scan or series of scans.
(5) Quality Assurance Procedures. Quality assurance procedures shall be conducted on the CT x-ray equipment.
(a) The quality assurance procedures shall be
in writing. Such procedures shall include, but not be limited to, the
following:
(i) Specifications of the tests
that are to be performed, including instructions to be employed in the
performance of those tests; and
(ii) Specifications of the frequency at which
tests are to be performed, the acceptable tolerance for each parameter measured
and actions to be taken if tolerances are exceeded.
(b) The parameters measured to satisfy
R313-28-160(5)(a)(ii) shall include, but not be limited to, kVp, mA and
reproducibility of dose appropriate to the type of CT procedures
performed.
(c) Records of tests
performed to satisfy the requirements of R313-28-160(5)(a) and (b) shall be
maintained for three years for inspection by the Division.
(6) Dose Calibration.
(a) Radiation measurements shall be performed
at least annually and after change or replacement of components which could
cause a change in the radiation output.
(b) The calibration of the radiation
measuring instrument shall be traceable to a national standard and shall be
calibrated at intervals not to exceed two years.
(c) Measurements shall be specified in terms
of the multiple scan average dose, using phantoms and technique factors
appropriate to the type of CT procedures performed.
Disclaimer: These regulations may not be the most recent version. Utah may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google
Privacy Policy and
Terms of Service apply.