Utah Administrative Code
Topic - Environmental Quality
Title R313 - Waste Management and Radiation Control, Radiation
Rule R313-28 - Use of X-Rays in the Healing Arts
Section R313-28-120 - Mammography X-Ray Systems - Equipment Design and Performance Standards
Universal Citation: UT Admin Code R 313-28-120
Current through Bulletin 2024-06, March 15, 2024
Only x-ray equipment meeting the following standards shall be used for mammography examinations.
(1) Equipment Design.
(a) FDA Standards. The
requirements of
21 CFR
1020.30 and
21 CFR
1020.31(2006) are adopted
and incorporated by reference.
(b)
Dedicated Equipment. The x-ray equipment shall be specifically designed for
mammography.
(c) Compression.
Devices parallel to the imaging plane shall be available to immobilize and
compress the breast during mammography procedures.
(d) Image Receptor. The x-ray equipment shall
have both an 18 cm by 24 cm and a 24 cm by 30 cm image receptor and moving
grids matched to each image receptor size.
(e) Automatic Exposure Control. X-ray
equipment used in healing arts screening shall have automatic exposure control
capabilities with a post exposure meter which indicates either
milliampere-seconds or time values.
(f) Focal Spot. The focal spot size and
source to image receptor distance configurations shall be limited to those
appropriate for mammography.
(g)
Beam Limitation. The x-ray equipment must allow for the x-ray field to extend
to or beyond the chest wall edge of the image receptor.
(h) Magnification. X-ray equipment used in a
noninvasive manner, requiring techniques beyond those utilized in standard
mammography of asymptomatic patients, shall have x-ray magnification capability
for noninvasive procedures. The equipment shall be able to provide at least one
magnification within the range of 1.4 to 2.0.
(2) Performance Standards.
(a) State Standards. The x-ray equipment
shall meet the applicable performance standards in R313- 28.
(b) Filtration. The useful beam shall have a
half-value layer between the values of the measured kilovolts peak divided by
100 and the measured kilovolts peak divided by 100 plus 0.1 mm of aluminum
equivalent. These values are to include the contribution to filtration by the
compression device.
(c) Minimum
Radiation Output. X-ray equipment installed after the effective date of this
rule shall meet the following standard: at 28 kilovolts peak on the focal spot
used in routine healing arts screening the x-ray equipment shall be capable of
sustaining a minimum output of 500 mR per second for at least three seconds.
This output shall be measured at a point 4.5 centimeters from the surface of
the patient support device when the source to image receptor distance is at its
maximum and the compression paddle is in the beam. Existing x-ray equipment
shall meet this minimum radiation output standard within one year of the
effective date of this rule.
(d)
Exposure Linearity. For kilovolts peak settings used clinically, the exposure
per mAs shall be within plus or minus ten percent of the average exposure per
mAs for those mAs stations or time stations, if applicable, that are
tested.
(e) Automatic Exposure
Control. The automatic exposure control mode shall produce consistent film
density under changing patient and examination conditions. These conditions
include breast thickness, adiposity, kilovolts peak and density settings. This
requirement will be deemed satisfied when:
(i) an automatic exposure control technique
guide is posted, and
(ii) for a
series of films obtained for attenuator thicknesses of two to seven centimeters
the resulting radiographic optical densities are within plus or minus 0.2 of
the average value when the kVp and density control setting are adjusted as
indicated on the technique guide. The attenuator used for determining
compliance shall be either acrylic or other tissue equivalent
material.
(f) Patient
Dose. The x-ray equipment must be capable of giving an average glandular dose
to an average size breast of average tissue density that does not exceed 3.0
mGy (0.3 rad) with a grid or 1.0 mGy (0.1 rad) without a grid. This will be
deemed satisfied when using an acrylic phantom of 4.5 cm thickness. In
addition, under all clinical use conditions, the average glandular dose to the
breast must be less than 5.0 mGy (0.5 rad) per film for healing arts screening
procedures.
(3) Mammography X-ray Equipment Quality Control.
(a) Initial Installation. Upon completion of
the initial installation of the x-ray equipment, and before it is commissioned
for clinical use, the equipment shall be evaluated by a mammography imaging
medical physicist who has been approved by the Board. The evaluation results
shall be submitted to the Director for review and approval.
(b) Annual Evaluation. At intervals not to
exceed 12 months or at the request of the Director, the x-ray equipment shall
be evaluated by a mammography imaging medical physicist who has been approved
by the Board.
(c) The registrant
shall develop and implement a quality control testing procedure for monitoring
the radiation performance of the x-ray equipment.
Disclaimer: These regulations may not be the most recent version. Utah may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google
Privacy Policy and
Terms of Service apply.