Utah Administrative Code
Topic - Environmental Quality
Title R313 - Waste Management and Radiation Control, Radiation
Rule R313-28 - Use of X-Rays in the Healing Arts
Section R313-28-120 - Mammography X-Ray Systems - Equipment Design and Performance Standards

Universal Citation: UT Admin Code R 313-28-120

Current through Bulletin 2024-06, March 15, 2024

Only x-ray equipment meeting the following standards shall be used for mammography examinations.

(1) Equipment Design.

(a) FDA Standards. The requirements of 21 CFR 1020.30 and 21 CFR 1020.31(2006) are adopted and incorporated by reference.

(b) Dedicated Equipment. The x-ray equipment shall be specifically designed for mammography.

(c) Compression. Devices parallel to the imaging plane shall be available to immobilize and compress the breast during mammography procedures.

(d) Image Receptor. The x-ray equipment shall have both an 18 cm by 24 cm and a 24 cm by 30 cm image receptor and moving grids matched to each image receptor size.

(e) Automatic Exposure Control. X-ray equipment used in healing arts screening shall have automatic exposure control capabilities with a post exposure meter which indicates either milliampere-seconds or time values.

(f) Focal Spot. The focal spot size and source to image receptor distance configurations shall be limited to those appropriate for mammography.

(g) Beam Limitation. The x-ray equipment must allow for the x-ray field to extend to or beyond the chest wall edge of the image receptor.

(h) Magnification. X-ray equipment used in a noninvasive manner, requiring techniques beyond those utilized in standard mammography of asymptomatic patients, shall have x-ray magnification capability for noninvasive procedures. The equipment shall be able to provide at least one magnification within the range of 1.4 to 2.0.

(2) Performance Standards.

(a) State Standards. The x-ray equipment shall meet the applicable performance standards in R313- 28.

(b) Filtration. The useful beam shall have a half-value layer between the values of the measured kilovolts peak divided by 100 and the measured kilovolts peak divided by 100 plus 0.1 mm of aluminum equivalent. These values are to include the contribution to filtration by the compression device.

(c) Minimum Radiation Output. X-ray equipment installed after the effective date of this rule shall meet the following standard: at 28 kilovolts peak on the focal spot used in routine healing arts screening the x-ray equipment shall be capable of sustaining a minimum output of 500 mR per second for at least three seconds. This output shall be measured at a point 4.5 centimeters from the surface of the patient support device when the source to image receptor distance is at its maximum and the compression paddle is in the beam. Existing x-ray equipment shall meet this minimum radiation output standard within one year of the effective date of this rule.

(d) Exposure Linearity. For kilovolts peak settings used clinically, the exposure per mAs shall be within plus or minus ten percent of the average exposure per mAs for those mAs stations or time stations, if applicable, that are tested.

(e) Automatic Exposure Control. The automatic exposure control mode shall produce consistent film density under changing patient and examination conditions. These conditions include breast thickness, adiposity, kilovolts peak and density settings. This requirement will be deemed satisfied when:
(i) an automatic exposure control technique guide is posted, and

(ii) for a series of films obtained for attenuator thicknesses of two to seven centimeters the resulting radiographic optical densities are within plus or minus 0.2 of the average value when the kVp and density control setting are adjusted as indicated on the technique guide. The attenuator used for determining compliance shall be either acrylic or other tissue equivalent material.

(f) Patient Dose. The x-ray equipment must be capable of giving an average glandular dose to an average size breast of average tissue density that does not exceed 3.0 mGy (0.3 rad) with a grid or 1.0 mGy (0.1 rad) without a grid. This will be deemed satisfied when using an acrylic phantom of 4.5 cm thickness. In addition, under all clinical use conditions, the average glandular dose to the breast must be less than 5.0 mGy (0.5 rad) per film for healing arts screening procedures.

(3) Mammography X-ray Equipment Quality Control.

(a) Initial Installation. Upon completion of the initial installation of the x-ray equipment, and before it is commissioned for clinical use, the equipment shall be evaluated by a mammography imaging medical physicist who has been approved by the Board. The evaluation results shall be submitted to the Director for review and approval.

(b) Annual Evaluation. At intervals not to exceed 12 months or at the request of the Director, the x-ray equipment shall be evaluated by a mammography imaging medical physicist who has been approved by the Board.

(c) The registrant shall develop and implement a quality control testing procedure for monitoring the radiation performance of the x-ray equipment.

Disclaimer: These regulations may not be the most recent version. Utah may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.