Utah Administrative Code
Topic - Environmental Quality
Title R309 - Drinking Water
Rule R309-211 - Monitoring and Water Quality: Distribution System - Total Coliform Requirements
Section R309-211-7 - Repeat Monitoring and E coli Requirements

Universal Citation: UT Admin Code R 309-211-7

Current through Bulletin 2024-18, September 15, 2024

(1) Repeat monitoring.

(a) If a sample taken under R309-211-5 though R309-211-6 is total coliform-positive, the system must collect a set of repeat samples within 24 hours of being notified of the positive result. The system must collect no fewer than three repeat samples for each total coliform-positive sample found. The Director may extend the 24-hour limit on a case-by-case basis if the system has a logistical problem in collecting the repeat samples within 24 hours that is beyond its control. Alternatively, the Director may implement criteria for the system to use in lieu of case-by-case extensions. In the case of an extension, the Director must specify how much time the system has to collect the repeat samples. The Director cannot waive the requirement for a system to collect repeat samples in paragraphs (1)(a) through (1)(c) of this section.

(b) The system must collect all repeat samples on the same day, except that the Director may allow a system with a single service connection to collect the required set of repeat samples over a three-day period or to collect a larger volume repeat sample(s) in one or more sample containers of any size, as long as the total volume collected is at least 300 ml.

(c) The system must collect an additional set of repeat samples in the manner specified in paragraphs (1)(a) through (1)(c) of this section if one or more repeat samples in the current set of repeat samples is total coliform-positive. The system must collect the additional set of repeat samples within 24 hours of being notified of the positive result, unless the Director extends the limit as provided in paragraph (1)(a) of this section. The system must continue to collect additional sets of repeat samples until either total coliforms are not detected in one complete set of repeat samples or the system determines that a coliform treatment technique trigger specified in R309-211-8(1) has been exceeded as a result of a repeat sample being total coliform-positive and notifies the Director. If a trigger identified in R309-211-8 is exceeded as a result of a routine sample being total coliform-positive, systems are required to conduct only one round of repeat monitoring for each total coliform-positive routine sample.

(d) After a system collects a routine sample and before it learns the results of the analysis of that sample, if it collects another routine sample(s) from within five adjacent service connections of the initial sample, and the initial sample, after analysis, is found to contain total coliforms, then the system may count the subsequent sample(s) as a repeat sample instead of as a routine sample.

(e) Results of all routine and repeat samples taken under R309-211-5 through R309-211-7 not invalidated by the Director must be used to determine whether a coliform treatment technique trigger specified in R309-211-8 has been exceeded.

(2) Escherichia coli (E. coli) testing.

(a) If any routine or repeat sample is total coliform-positive, the system must analyze that total coliform-positive culture medium to determine if E. coli are present. If E. coli are present, the system must notify the Director by the end of the day when the system is notified of the test result, unless the system is notified of the result after the Director office is closed and the Director does not have either an after-hours phone line or an alternative notification procedure, in which case the system must notify the Director before the end of the next business day.

(b) The Director has the discretion to allow a system, on a case-by-case basis, to forgo E. coli testing on a total coliform-positive sample if that system assumes that the total coliform-positive sample is E. coli-positive. Accordingly, the system must notify the Director as specified in paragraph (2)(a) of this section and the provisions of R309-200-5(6)(b) apply.

Disclaimer: These regulations may not be the most recent version. Utah may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.