The requirement in Subsection 58-46a-303(1)(b)(ii) for calibration of each appropriate technical instruments used in practice is defined as follows:
(1) each audiometer used in the fitting of hearing aids shall be calibrated when necessary, but not less than annually;
(2) the calibration shall include to ANSI standards calibration of frequency accuracy, acoustic output, attenuator linearity, and harmonic distortion; and
(3) calibration shall be accomplished by the manufacturer, or a properly trained person, or an institution of higher learning equipped with proper instruments for calibration of an audiometer.
Disclaimer: These regulations may not be the most recent version. Utah may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.