Current through Reg. 49, No. 38; September 20, 2024
(e)
Definitions. The following words and terms when used in this section shall have
the following meaning unless the context clearly indicates otherwise.
(1) Absorbed dose (D)--The mean energy
imparted by ionizing radiation to matter. Absorbed dose is determined as the
quotient of dE by dM, where dE is the mean energy imparted by ionizing
radiation to matter of mass dM. The SI unit of absorbed dose is joule per
kilogram and the special name of the unit of absorbed dose is the gray (Gy).
The previously used special unit of absorbed dose (rad) is being replaced by
the gray.
(2) Absorbed dose
rate--Absorbed dose per unit time, for machines with timers, or dose monitor
unit per unit time for linear accelerators.
(3) Air kerma--The kinetic energy released in
air by ionizing radiation. Kerma is the quotient of dE by dM, where dE is the
sum of the initial kinetic energies of all the charged ionizing particles
liberated by uncharged ionizing particles in air of mass dM. The SI unit of air
kerma is joule per kilogram and the special name for the unit of kerma is the
gray (Gy).
(4) Barrier--(See
definition for protective barrier).
(5) Beam axis--The axis of rotation of the
beam limiting device.
(6)
Beam-flattening filter--(See field-flattening filter).
(7) Beam-limiting device--A field defining
collimator, integral to the therapeutic radiation machine, which provides a
means to restrict the dimensions of the useful beam.
(8) Beam monitoring system--A system designed
and installed in the radiation head to detect and measure the radiation present
in the useful beam.
(9) Beam
quality--A term that describes the penetrating power of the x-ray beam. This is
identified numerically by half-value layer and is influenced by kilovolt peak
(kVp) and filtration.
(10) Beam
quality (accelerator)--A term that describes the type and penetrating power of
the ionizing radiation produced for certain machine settings.
(11) Beam scattering foil--A thin piece of
material (usually metallic) placed in the beam to scatter a beam of electrons
in order to provide a more uniform electron distribution in the useful
beam.
(12) Central axis of the
beam--An imaginary line passing through the center of the useful beam and the
center of the plane figure formed by the edge of the first beam-limiting
device.
(13) Certified physician--A
physician licensed by the Texas Medical Board and certified in radiation
oncology or therapeutic radiology.
(14) Coefficient of variation or C--The ratio
of the standard deviation to the mean value of a population of observations. It
is estimated using the following equation:
Attached
Graphic
(15)
Collimator--A device or mechanism by which the x-ray beam is restricted in
size.
(16) Computed tomography
(CT)--The production of a tomogram by the acquisition and computer processing
of x-ray transmission data.
(17)
Continuous pressure type switch--A switch so constructed that a circuit closing
contact can be maintained only by continuous pressure on the switch by the
operator.
(18) Control panel--The
part of the radiation machine where the switches, knobs, push buttons, and
other hardware necessary for manually setting the technique factors are
located. For purposes of this section console is an equivalent term.
(19) CT conditions of operation--All
selectable parameters governing the operation of a CT x-ray system including,
but not limited to, nominal tomographic section thickness, filtration, and the
technique factors as defined in this subsection.
(20) Detector--(See definition for radiation
detector).
(21) Diaphragm--A device
or mechanism by which the x-ray beam is restricted in size.
(22) Dose monitor unit (DMU)--A unit response
from the beam monitoring system from which the absorbed dose can be
calculated.
(23) Dosimetry
system--A system of devices used for the detection, measurement, and display of
qualitative and quantitative radiation exposures.
(24) Electronic brachytherapy--A method of
radiation therapy using electrically generated x-rays to deliver a radiation
dose at a distance of up to a few centimeters by intracavitary, intraluminal or
interstitial application, or by applications with the source in contact with
the body surface or very close to the body surface.
(25) Electronic brachytherapy device--The
system used to produce and deliver therapeutic radiation including the x-ray
tube, the control mechanism, the cooling system, and the power
source.
(26) Electronic
brachytherapy source--The x-ray tube component used in an electronic
brachytherapy device.
(27) External
beam radiation therapy--Therapeutic irradiation in which the source of
radiation is at a distance from the body.
(28) Field-flattening filter--A filter used
to homogenize the absorbed dose rate over the radiation field.
(29) Field size--The dimensions along the
major axes of an area in a plane perpendicular to the central axis of the beam
at the normal treatment or examination source to image distance and defined by
the intersection of the major axes and the 50% isodose line.
(30) Filter--Material placed in the useful
beam to change beam quality in therapeutic radiation machines subject to
subsection (h) of this section.
(31) Focal spot--The area projected on the
anode of the x-ray tube that is bombarded by the electrons accelerated from the
cathode and from which the useful beam originates.
(32) Gantry--That part of the radiation
therapy system supporting and allowing possible movements of the radiation head
about the center of rotation.
(33)
Gray (Gy)--For purposes of this section, the SI unit of absorbed dose, kerma,
and specific energy imparted equal to 1 joule per kilogram. For purposes of
this section the previous unit of absorbed dose (rad) is being replaced by the
gray (1 Gy = 100 rad).
(34)
Half-value layer (HVL)--The thickness of a specified material which attenuates
x-radiation or gamma radiation to an extent such that the exposure rate (air
kerma rate), or absorbed dose rate is reduced to one-half of the value measured
without the material at the same point.
(35) Healing arts--Any treatment, operation,
diagnosis, prescription, or practice for the ascertainment, cure, relief,
palliation, adjustment, or correction of any human disease, ailment, deformity,
injury, or unhealthy or abnormal physical or mental condition.
(36) Image receptor--Any device, such as a
fluorescent screen or radiographic film, that transforms incident x-ray photons
either into a visible image or into another form that can be made into a
visible image by further transformations.
(37) Institutional Review Board (IRB)--Any
board, committee, or other group formally designated by an institution to
review, approve the initiation of, and conduct periodic review of biomedical
research involving human subjects.
(38) Interlock--A device preventing the start
or continued operation of equipment unless certain predetermined conditions
prevail.
(39) Interruption of
irradiation--The stopping of irradiation with the possibility of continuing
irradiation without resetting of operating conditions at the control
panel.
(40) Irradiation--The
exposure of a living being or matter to ionizing radiation.
(41) Isocenter--The center of the sphere
through which the useful beam axis passes while the gantry moves through its
full range of motions.
(42)
Kilovolt (kV) (kilo electron volt (keV))--The energy equal to that acquired by
a particle with one electron charge in passing through a potential difference
of one thousand volts in a vacuum. (Note: current convention is to use kV for
photons and keV for electrons.)
(43) Kilovolt peak--kVp (See definition for
peak tube potential).
(44) Lead
equivalent--The thickness of lead affording the same attenuation, under
specified conditions, as the material in question.
(45) Leakage radiation--Radiation emanating
from the source(s) assembly except for the useful beam and radiation produced
when the exposure switch or timer is not activated.
(46) Leakage technique factors--The technique
factors associated with the source assembly that is used in measuring leakage
radiation.
(47) Licensed medical
physicist--An individual holding a current Texas license under the Medical
Physics Practice Act, Texas Occupations Code, Chapter 602, with a specialty in
therapeutic radiological physics.
(48) Light field--The area illuminated by
light, simulating the radiation field.
(49) mA--Milliampere.
(50) Medical event--An event that meets the
criteria specified in subsection (i) of this section.
(51) Megavolt (MV) (megaelectron volt
(MeV))--The energy equal to that acquired by a particle with one electron
charge in passing through a potential difference of one million volts in a
vacuum.
(52) Mobile electronic
brachytherapy device--An electronic brachytherapy device that is transported
from one address to be used at another address.
(53) Moving beam radiation therapy--Radiation
therapy with any planned displacement of radiation field or patient relative to
each other, or with any planned change of absorbed dose distribution. It
includes arc, skip, conformal, intensity modulation and rotational
therapy.
(54) Nominal treatment
distance--The following nominal treatment distances shall apply.
(A) For electron irradiation, the distance
from the scattering foil, virtual source, or exit window of the electron beam
to the entrance surface of the irradiated object along the central axis of the
useful beam, as specified by the manufacturer.
(B) For x-ray irradiation, the virtual source
or target to isocenter distance along the central axis of the useful beam to
the isocenter. For non-isocentric equipment, this distance shall be that
specified by the manufacturer.
(55) Output--The exposure rate (air kerma
rate), dose rate, or a quantity related to these rates from a therapeutic
radiation machine.
(56) Peak tube
potential--The maximum value of the potential difference in kilovolts across
the x-ray tube during an exposure.
(57) Phantom--An object behaving in
essentially the same manner as tissue, with respect to absorption or scattering
of the ionizing radiation in question.
(58) Physician--An individual licensed by the
Texas Medical Board.
(59) Port
film--An x-ray exposure made with a radiation therapy system to visualize a
patient's treatment area using radiographic film.
(60) Portable shielding--Moveable shielding
that can be placed in the primary or secondary beam to reduce the radiation
exposure to the patient, occupational worker or a member of the public. The
shielding can be easily moved to position with use of mobility devices or by
hand.
(61) Prescribed dose--The
total dose and dose per fraction as documented in the written directive. The
prescribed dose is an estimation from measured data from a specified
therapeutic machine using assumptions that are clinically acceptable for the
treatment technique and historically consistent with the clinical calculations
previously used for patients treated with the same clinical
technique.
(62) Primary dose
monitoring system--A system that will monitor the useful beam during
irradiation and that will terminate irradiation when a preselected number of
dose monitor units have been delivered.
(63) Primary protective barrier--(See
definition for protective barrier).
(64) Protective apron--An apron made of
radiation absorbing materials used to reduce radiation exposure.
(65) Protective barrier--A barrier of
radiation absorbing materials used to reduce radiation exposure. The types of
protective barriers are as follows:
(A)
primary protective barrier--A barrier sufficient to attenuate the useful beam
to the required degree.
(B)
secondary protective barrier--A barrier sufficient to attenuate the stray
radiation to the required degree.
(66) Protective glove--A glove made of
radiation absorbing materials used to reduce radiation exposure.
(67) Radiation detector--A device which, in
the presence of radiation provides, by either direct or indirect means, a
signal or other indication suitable for use in measuring 1 or more quantities
of incident radiation.
(68)
Radiation field--(See definition for useful beam).
(69) Radiation head--The structure from which
the useful beam emerges.
(70)
Radiation oncologist--A physician with a specialty in radiation
therapy.
(71) Radiation therapy
simulation system (simulator)--An x-ray system intended for localizing and
confirming the volume to be irradiated during radiation treatment and
confirming the position and size of the therapeutic irradiation
field.
(72) Radiation therapy
system--An x-ray system that utilizes prescribed doses of ionizing radiation
for treatment.
(73) Scan--The
complete process of collecting x-ray transmission data for the production of a
tomogram. Data can be collected simultaneously during a single scan for the
production of one or more tomograms.
(74) Scan increment--The amount of relative
displacement of the patient with respect to the CT x-ray system between
successive scans measured along the direction of such displacement.
(75) Scan sequence--A preselected set of 2 or
more scans performed consecutively under preselected CT conditions of
operation.
(76) Scan time--The
period of time between the beginning and end of x-ray transmission data
accumulation for a single scan.
(77) Scattered radiation--Radiation that has
been deviated in direction during passage through matter.
(78) Secondary dose monitoring system--A
system which will terminate irradiation in the event of failure of the primary
dose monitoring system.
(79)
Secondary protective barrier (See definition for protective barrier).
(80) Shutter--A device attached to the tube
housing assembly which can totally intercept the useful beam and which has a
lead equivalency not less than that of the tube housing assembly.
(81) Source-to-skin distance (SSD)--The
distance from the source to the skin of the patient.
(82) Spot check--Those tests and analyses
performed at specified intervals for the purpose of verifying the consistent
output of radiation equipment.
(83)
Stationary beam therapy--Radiation therapy without displacement of one or more
mechanical axes relative to the patient during irradiation.
(84) Supervision--The delegating of the task
of applying radiation in accordance with this section to persons not licensed
in the healing arts or veterinary medicine, who provide services under the
physician's control. The physician or veterinarian assumes full responsibility
for these tasks and shall assure that the tasks will be administered
correctly.
(85) Target--That part
of an x-ray tube or accelerator onto which a beam of accelerated particles is
directed to produce ionizing radiation or other particles.
(86) Termination of irradiation--The stopping
of irradiation in a fashion which will not permit continuance of irradiation
without the resetting of operating conditions at the control panel.
(87) Therapeutic radiation machine--X ray or
electron producing equipment designed and used for external beam radiation
therapy.
(88) Traceable to a
national standard--This indicates that a quantity or a measurement has been
compared to a national standard, for example, National Institute of Standards
and Technology, directly or indirectly through one or more intermediate steps
and that all comparisons have been documented.
(89) Tube housing assembly--The tube housing
with tube installed. It includes high-voltage and/or filament transformers and
other appropriate elements when such are contained within the tube
housing.
(90) Useful
beam--Radiation that passes through the window, aperture, cone, or other
collimating device of the source housing. Also referred to as the primary
beam.
(91) Veterinarian--An
individual licensed by the Texas Board of Veterinary Medical
Examiners.
(92) Virtual source--A
point from which radiation appears to originate.
(93) Wedge filter--An added filter effecting
continuous progressive attenuation on all or part of the useful beam.
(94) Written directive--An order in writing
for the administration of radiation to a specific patient as specified in
subsection (h)(1)(F)(ii) of this section.