Current through Reg. 49, No. 38; September 20, 2024
(e) Definitions. The following words and
terms, when used in this section, shall have the following meanings unless the
context clearly indicates otherwise.
(1)
Accessible surface--The external surface of the enclosure or housing provided
by the manufacturer.
(2) Air
kerma--The kinetic energy released in air by ionizing radiation. Kerma is the
quotient of dE by dM, where dE is the sum of the initial kinetic energies of
all the charged ionizing particles liberated by uncharged ionizing particles in
air of mass dM. The SI unit of air kerma is joule per kilogram and the special
name for the unit of kerma is the gray (Gy). For purposes of this section, when
exposure in air measured in roentgen (R) is to be converted to dose in air
measured in gray (Gy), a nationally recognized standard air conversion factor
shall be used.
(3) Aluminum
equivalent--The thickness of type 1100 aluminum alloy affording the same
attenuation, under specified conditions, as the material in question. The
nominal chemical composition of type 1100 aluminum alloy is 99% minimum
aluminum, 0.12% copper.
(4)
Attenuate--To reduce the exposure rate upon passage of radiation through
matter.
(5) Attenuation block--A
block or stack, having dimensions 20 centimeters (cm) by 20 cm by 3.8 cm, of
type 1100 aluminum alloy or other materials having equivalent attenuation. The
nominal chemical composition of type 1100 aluminum alloy is 99% minimum
aluminum, 0.12% copper.
(6)
Automatic exposure control (AEC)--A device that automatically controls one or
more technique factors in order to obtain a required quantity of radiation at
preselected locations (See definition for phototimer).
(7) Automatic exposure rate control (AERC)--A
device that automatically controls one or more technique factors in order to
obtain a required quantity of radiation per unit time at preselected
locations.
(8) Barrier (See
definition for protective barrier).
(9) Beam axis--A line from the source through
the centers of the x-ray fields.
(10) Beam-limiting device--A device that
provides a means to restrict the dimensions of the x-ray field.
(11) Beam quality (diagnostic x-ray)--A term
that describes the penetrating power of the x-ray beam. This is identified
numerically by half-value layer and is influenced by kilovolt peak (kVp) and
filtration.
(12) Bone
densitometer--A device intended for medical purposes to measure bone density
and mineral content by x-ray transmission measurements through the bone and
adjacent tissues.
(13) Calibration
of instruments--The comparative response or reading of an instrument relative
to a series of known radiation values over the range of the
instrument.
(14) Category 1
continuing medical education units (CMEU)--Educational activities designated as
Category I and approved by the Accreditation Council for Continuing Medical
Education, the American Osteopathic Association, a state medical society, or an
equivalent organization.
(15)
Central axis of the beam--A line passing through the virtual source and the
center of the plane figure formed by the edge of the first beam-limiting
device.
(16) Certified x-ray
systems--X-ray systems that have been certified in accordance with Title 21,
Code of Federal Regulations (CFR).
(17) Certified radiologist--A physician
certified by the American Board of Radiology, the American Osteopathic Board of
Radiology, the Royal College of Physicians and Surgeons of Canada, or Le
College des Medecins du Quebec.
(18) Coefficient of variation or C--The ratio
of the standard deviation to the mean value of a population of observations. It
is estimated using the following equation:
Attached
Graphic
(19)
Collimator--A device or mechanism by which the x-ray beam is restricted in
size.
(20) Computed tomography
(CT)--The production of a tomogram by the acquisition and computer processing
of x-ray transmission data.
(21)
Computed tomography dose index (CTDI)--CTDI represents the average absorbed
dose along the z-axis from a series of contiguous irradiations. CTDI
approximates the average central dose values associated with a spatially
complex dose distribution in a reference acrylic phantom for one particular set
of exam techniques. CTDI is defined exclusively for axial scanning.
(22) Control panel--The part of the radiation
machine control upon which are mounted the switches, knobs, push buttons, and
other hardware necessary for manually setting the technique factors.
(23) CT conditions of operation--All
selectable parameters governing the operation of a CT system including, but not
limited to, nominal tomographic section thickness, filtration, and the
technique factors as defined in this subsection.
(24) CT gantry--The tube housing assemblies,
beam-limiting devices, detectors, and the supporting structures and frames that
hold these components.
(25)
Cumulative air kerma (CAK)--The air kerma accumulated at a specific point in
space relative to the fluoroscopic gantry during a procedure. CAK does not
include tissue backscatter and is measured in Gy. Some manufacturers identify
CAK as cumulative dose (CD).
(26)
Diagnostic source assembly--The tube housing assembly with a beam-limiting
device attached.
(27) Diagnostic
x-ray system--An x-ray system designed for irradiation of any part of the human
body for the purpose of diagnosis or visualization.
(28) Dose-area product (DAP)--Dose-area
product is the integral of air kerma (absorbed dose to air) across the entire
x-ray beam emitted from the x-ray tube. DAP is a surrogate measurement for the
entire amount of energy delivered to the patient by the beam. DAP is measured
in R x cm2 (cGy x
cm2).
(29) Dose-length product (DLP)--DLP is the
CTDI multiplied by the scan length (slice thickness x number of slices) in
centimeters. It reflects the total energy absorbed attributable to the complete
scan acquisition but is independent of what is actually scanned.
(30) Entrance exposure (Entrance air
kerma)--The entrance exposure in air expressed in roentgens (R) or the entrance
dose in air (air kerma) expressed in gray (Gy), measured at the point where the
center of the useful beam enters the patient.
(31) Entrance exposure rate (air kerma
rate)--The exposure (air kerma) per unit time at the point where the center of
the useful beam enters the patient.
(32) Field emission equipment--Equipment that
uses an x-ray tube in which electron emission from the cathode is due solely to
the action of an electric field.
(33) Field size--The dimensions along the
major axes of an area in a plane perpendicular to the central axis of the beam
at the normal treatment or examination source-to-image distance and defined by
the intersection of the major axes and the 50% isodose line.
(34) Filter--Material placed in the useful
beam to preferentially absorb selected radiations.
(35) Fluoroscopic system--A system in which
x-ray photons produce a fluoroscopic image. It includes the image receptors
such as the image intensifier and spot-film device, electrical interlocks, if
any, and structural material providing linkage between the image receptor and
diagnostic source assembly.
(36)
Fluoroscopically-Guided Interventional (FGI) Procedures--An interventional
diagnostic or therapeutic procedure performed via percutaneous or other access
routes, usually with local anesthesia or intravenous sedation, which uses
external ionizing radiation in the form of fluoroscopy to localize or
characterize a lesion, diagnostic site, or treatment site, to monitor the
procedure, and to control and document therapy. FGI procedures may include but
not be limited to:
(A) TIPS creation
(transjugular intrahepatic portosystemic shunt);
(B) Embolization (any location, any
lesion);
(C) Stroke
therapy;
(D) Biliary
drainage;
(E) Angioplasty with or
without stent placement;
(F)
Stent-graft placement;
(G)
Chemoembolization;
(H) Angiography
and intervention for gastrointestinal hemorrhage;
(I) Carotid stent placement;
(J) RF (radiofrequency) cardiac
ablation;
(K) Complex placement of
cardiac EP (electrophysiology) devices; and
(L) PCI (percutaneous coronary intervention)
(single or multiple vessel).
(37) Focal spot--The area projected on the
anode of the x-ray tube bombarded by the electrons accelerated from the cathode
and from which the useful beam originates.
(38) General purpose x-ray system--Any x-ray
system that is not limited by design to radiographic examinations of specific
anatomical regions.
(39) Gonadal
shield--A protective barrier for the testes or ovaries.
(40) Half-value layer (HVL)--The thickness of
a specified material that attenuates the beam of radiation to an extent such
that the exposure rate (air kerma rate) is reduced to one-half of its original
value.
(41) Healing arts--Any
system, treatment, operation, diagnosis, prescription, or practice for the
ascertainment, cure, relief, palliation, adjustment, or correction of any human
disease, ailment, deformity, injury, or unhealthy or abnormal physical or
mental condition.
(42) Healing arts
screening--The testing of asymptomatic human beings using radiation machines
for the detection or evaluation of health indications when such tests are not
specifically and individually ordered by a licensed practitioner of the healing
arts legally authorized to prescribe such x-ray tests for the purpose of
diagnosis or treatment.
(43) High
level control for fluoroscopy--Any selected mode having an entrance exposure
rate (air kerma rate) above 10 roentgens per minute (R/min) or (100 mGy/min).
This mode shall meet the high level requirements in subsection
(m)(3)(A)(i)(II), (ii)(II), or (iii)(II) of this section.
(44) Image intensifier--A device, installed
in its housing that instantaneously converts an x-ray pattern into a
corresponding light or digital image.
(45) Image receptor--Any device, such as a
fluorescent screen or radiographic film that transforms incident x-ray photons
either into a visible image or into another form that can be made into a
visible image by further transformations.
(46) Irradiation--The exposure of matter to
ionizing radiation.
(47)
kV--Kilovolt.
(48) kVp--Kilovolt
peak (See definition for peak tube potential).
(49) kWs--Kilowatt-second. It is equivalent
to 10 E 3 watt-second, where 1 watt-second = 1 kV x 1 milliampere (mA) x 1
second.
(50) Lead equivalent--The
thickness of lead affording the same attenuation, under specified conditions,
as the material in question.
(51)
Leakage radiation--Radiation emanating from the diagnostic source assembly
except for the useful beam and radiation produced when the exposure switch or
timer is not activated.
(52)
Leakage technique factors--The technique factors associated with the diagnostic
source assembly that is used in measuring leakage radiation. They are defined
as follows:
(A) for diagnostic source
assemblies intended for capacitor energy storage equipment, the maximum-rated
peak tube potential and the maximum-rated number of exposures in an hour for
operation at the maximum-rated peak tube potential with the quantity of charge
per exposure being 10 millicoulombs (10 milliampere-second (mAs)) or the
minimum obtainable from the unit, whichever is larger;
(B) for diagnostic source assemblies intended
for field emission equipment rated for pulsed operation, the maximum-rated peak
tube potential and the maximum-rated number of x-ray pulses in an hour for
operation at the maximum-rated peak tube potential; or
(C) for all other diagnostic source
assemblies, the maximum-rated peak tube potential and the continuous tube
current for the maximum-rated peak tube potential.
(53) Licensed medical physicist--An
individual holding a current Texas license under the Medical Physics Practice
Act, Texas Occupations Code, Chapter 602, with a specialty in diagnostic
medical physics.
(54)
mA--Milliampere.
(55)
mAs--Milliampere-second.
(56)
Medical research--The investigation of various health risks and
diseases.
(57) Mobile service
operation--The provision of x-ray systems and personnel at temporary sites for
limited time periods. The x-ray systems may be fixed inside a motorized vehicle
or may be a radiation machine that may be removed from the vehicle and taken
into a facility for use.
(58)
Multiple slice tomogram system--A computed tomography x-ray system that obtain
x-ray transmission data simultaneously during a single scan to produce more
than one tomogram.
(59) Nominal
tomographic section thickness--The full-width at half-maximum of the
sensitivity profile taken at the center of the cross sectional volume over
which x-ray transmission data are collected.
(60) Non-certified x-ray systems--X-ray
systems manufactured and assembled prior to certification requirements of Title
21, CFR, effective as specified in Title 21, CFR, Part 1020.30(a).
(61) Patient--An individual subjected to
healing arts examination, diagnosis, or treatment.
(62) Peak tube potential--The maximum value
of the potential difference in kilovolts across the x-ray tube during an
exposure.
(63) Phantom--A volume of
material behaving in a manner that can be related to tissue with respect to the
attenuation and scattering of radiation.
(64) Phototimer--A method for controlling
exposures to image receptors by the amount of radiation that reaches a
radiation detection device. The radiation detection device is part of an
electronic circuit that controls the duration of time the tube is activated
(See definition for automatic exposure control).
(65) Physician--An individual licensed by the
Texas Medical Board.
(66) Portable
x-ray systems--X-ray systems that are mounted on a permanent base with wheels
and/or casters for moving while completely assembled. Portable x-ray systems
may also include x-ray systems designed to be hand-carried.
(67) Practitioner of the healing arts
(practitioner)--A person licensed to practice healing arts by either the Texas
Medical Board as a physician, the Texas Board of Chiropractic Examiners, or the
Texas State Board of Podiatric Medical Examiners.
(68) Primary protective barrier--(See
definition for protective barrier).
(69) Protective apron--An apron made of
radiation attenuating materials used to reduce radiation exposure.
(70) Protective barrier--A barrier of
radiation absorbing materials used to reduce radiation exposure. The types of
protective barriers are as follows:
(A)
primary protective barrier--A barrier sufficient to attenuate the useful beam
to the required degree.
(B)
secondary protective barrier--A barrier sufficient to attenuate the stray
radiation to the required degree.
(71) Protective glove--A glove made of
radiation attenuating materials used to reduce radiation exposure.
(72) Radiation machine--Any device capable of
producing ionizing radiation except those devices with radioactive material as
the only source of radiation.
(73)
Radiation oncologist--A physician with a specialty in radiation
therapy.
(74) Radiograph--An image
receptor on which the image is created directly or indirectly by an x-ray
exposure and results in a permanent record.
(75) Radiologist--A physician with a
specialty in using ionizing and non-ionizing radiation for medical imaging and
interpretation for diagnostic and treatment purposes.
(76) Reference level--The suggested levels at
which a facility should review its methods and determine if acceptable image
quality can be achieved at a lower radiation output level as determined by
measurements taken from a selected phantom. The specific purpose of the
reference level is to provide a benchmark for comparison of imaging equipment
performance under prescribed conditions and is not intended to define a maximum
or minimum exposure limit for any patient.
(77) Reference plane--A plane that is
displaced from and parallel to the tomographic plane.
(78) Roentgen (R)--The special unit of
exposure. One roentgen (R) equals 2.58 x 10-4 C/kg
of air.
(79) Scan--The complete
process of collecting x-ray transmission data for the production of a tomogram.
Data can be collected simultaneously during a single scan for the production of
one or more tomograms.
(80) Scan
increment--The amount of relative displacement of the patient with respect to
the CT system between successive scans measured along the direction of such
displacement.
(81) Scan sequence--A
preselected set of 2 or more scans performed consecutively under preselected CT
conditions of operation.
(82) Scan
time--The period of time between the beginning and end of x-ray transmission
data accumulation for a single scan.
(83) Scattered radiation--Radiation that has
been deviated in direction during passage through matter.
(84) Secondary protective barrier (See
definition for protective barrier).
(85) Shutter--A device attached to the tube
housing assembly that can totally intercept the useful beam and that has a lead
equivalency not less than that of the tube housing assembly.
(86) Single tomogram system--CT system that
obtains x-ray transmission data during a scan to produce a single
tomogram.
(87) Source--The focal
spot of the x-ray tube.
(88)
Source-to-image receptor distance--The distance from the source to the center
of the input surface of the image receptor.
(89) Source-to-skin distance (SSD)--The
distance from the source to the skin of the patient.
(90) Special purpose x-ray system--Any x-ray
system that is limited by design to radiographic examinations of specific
anatomical regions. Special purpose x-ray systems include, but are not limited
to, dedicated chest units, cystography units, and head and skull
units.
(91) Special procedures--The
application of special x-ray systems and specialized techniques to obtain
required diagnostic information. Special procedures include, but are not
limited to, angiography, cardiac catheterization, myelography, and
surgery.
(92) Spot film--A
radiograph that is made during a fluoroscopic examination to permanently record
conditions that exist during that fluoroscopic procedure.
(93) Spot film device--A device intended to
transport and/or position a radiographic image receptor between the x-ray
source and fluoroscopic image receptor. It includes a device intended to hold a
cassette over the input end of an image intensifier for the purpose of making a
radiograph.
(94) Stationary x-ray
system--A stationary x-ray system that is installed in a fixed
location.
(95) Stray radiation--The
sum of leakage and scattered radiation.
(96) Supervision--The delegating, by the
practitioner, of the task of applying radiation to persons who perform tasks
under the practitioner's control and who are certified under the Medical
Radiologic Technologist Act, Texas Occupations Code, Chapter 601. The
practitioner assumes full responsibility for these tasks and shall assure that
the tasks will be administered correctly.
(97) Target--The part of a radiation machine
head that by design intercepts a beam of accelerated particles with subsequent
emission of other radiation.
(98)
Technique chart--A chart that provides technical factors, anatomical
examination, patient thickness for examination being performed, and
source-to-image distance needed to make clinical radiographs when the
radiographic system is in manual mode.
(99) Technique factors--The conditions of
operation that are specified as follows:
(A)
for capacitor energy storage equipment, peak tube potential in kV and quantity
of charge in mAs.
(B) for field
emission equipment rated for pulsed operation, peak tube potential in kV and
number of x-ray pulses;
(C) for CT
systems designed for pulsed operations, peak tube potential in kV, scan time in
seconds, and either tube current in mA, x-ray pulse width in seconds, and the
number of x-ray pulses per scan or the product of tube current, x-ray pulse
width, and the number of x-ray pulses in mAs;
(D) for CT systems not designed for pulsed
operation, peak tube potential in kV, and either tube current in mA and scan
time in seconds or the product of tube current and exposure time in mAs when
the scan time and exposure time are equivalent; and
(E) for all other x-ray systems, peak tube
potential in kV and either tube current in mA and exposure time in seconds or
the product of tube current and exposure time in mAs.
(100) Tomogram--The depiction of the x-ray
attenuation properties of a section through the body.
(101) Tomographic plane--The geometric plane
that is identified as corresponding to the output.
(102) Tomographic section--The volume of an
object whose x-ray attenuation properties are imaged in a tomogram.
(103) Traceable to a national standard--This
indicates that a quantity or a measurement has been compared to a national
standard, for example, the National Institute of Standards and Technology,
directly or indirectly through one or more intermediate steps and that all
comparisons have been documented.
(104) Tube--An x-ray tube, unless otherwise
specified.
(105) Tube housing
assembly--The tube housing with tube installed. It includes high-voltage and/or
filament transformers and other appropriate elements when such are contained
within the tube housing.
(106)
Useful beam--Radiation that passes through the window, aperture, cone, or other
collimating device of the source housing. Also referred to as the primary
beam.
(107) X-ray control panel--A
device that controls input power to the x-ray high-voltage generator and/or the
x-ray tube. It includes equipment such as timers, phototimers, automatic
brightness stabilizers, and similar devices that control the technique factors
of an x-ray exposure.
(108) X-ray
field--That area of the intersection of the useful beam and any one of the set
of planes parallel to and including the plane of the image receptor, whose
perimeter is the locus of points at which the exposure rate (air kerma rate) is
1/4th of the maximum in the intersection.
(109) X-ray high-voltage generator--A device
that transforms electrical energy from the potential supplied by the x-ray
control to the tube operating potential. The device may also include means for
transforming alternating current to direct current, filament transformers for
the x-ray tubes, high-voltage switches, electrical protective devices, and
other appropriate elements.
(110)
X-ray system--An assemblage of components for the controlled production of
x-rays and include a radiation machine. It includes minimally an x-ray
high-voltage generator, an x-ray control, a tube housing assembly, a
beam-limiting device, and the necessary supporting structures. Additional
components that function with the system are considered integral parts of the
system.
(111) X-ray tube--Any
electron tube that is designed to be used primarily for the production of x
rays.
(i) General
operating requirements.
(1) Technique chart.
A technique chart relevant to the particular radiation machine shall be
provided or electronically displayed in the vicinity of the control panel and
used by all operators.
(2)
Operating and safety procedures.
(A) Each
registrant shall have and implement written operating and safety procedures.
The procedures shall include, but are not limited to, the following as
applicable:
(i) posting notices to workers in
accordance with §
289.203(b)
of this title;
(ii) instructions to
workers in accordance with §
289.203(c)
of this title;
(iii) notifications
and reports to individuals in accordance with §
289.203(d)
of this title;
(iv) ordering x-ray
exams in accordance with §
289.231(b)(1)
of this title;
(v) occupational
dose requirements in accordance with §
289.231(m)
of this title;
(vi) personnel
monitoring requirements in accordance with §289.231(n), (q), and (s) of
this title;
(vii) posting of a
radiation area in accordance with §
289.231(x)
of this title;
(viii) use of a
technique chart in accordance with paragraph (1) of this subsection;
(ix) use of protective devices in accordance
with paragraph (4) of this subsection;
(x) credentialing requirements for
individuals operating radiation machines in accordance with paragraph (5) of
this subsection;
(xi) exposure of
individuals other than the patient in accordance with paragraph (7) of this
subsection;
(xii) holding of
patients or image receptors in accordance with the following;
(I) a list of circumstances in which
mechanical holding devices cannot be routinely utilized; and
(II) a procedure used for selecting an
individual to hold or support the patient or image receptor.
(xiii) gonadal shielding in
accordance with paragraph (13) of this subsection;
(xiv) control of scattered radiation in
accordance with subsection (m)(8) of this section; and
(xv) film processing program or digital image
processing protocols in accordance with subsections (p), (q), and (r) of this
section.
(B) The
registrant shall maintain the operating and safety procedures in accordance
with subsection (s)(1) of this section for inspection by the agency.
(C) The procedures shall be made available to
each individual operating a radiation machine, including any restrictions of
the operating technique required for the safe operation of the particular x-ray
system.
(D) The registrant shall
document that each individual operating a radiation machine has read the
operating and safety procedures at least annually and shall maintain this
documentation for inspection by the agency. The documentation shall include the
following:
(i) name and signature of
individual;
(ii) date individual
read the operating and safety procedures; and
(iii) initials of the RSO.
(3) Occupational dose
limits and personnel monitoring. Except as otherwise exempted, all individuals
who are associated with the operation of a radiation machine are subject to the
occupational dose limits of §
289.231(m)
of this title regarding dose limits to individuals, and the personnel
monitoring requirements of §
289.231(n)
of this title.
(4) Protective
devices. Protective devices shall be utilized when required, as in paragraphs
(8)(B) and (C), (10), and (13) of this subsection, and subsection (m)(8) of
this section.
(A) Protective devices shall be
of no less than 0.25 millimeter (mm) lead equivalent material except as
specified in paragraph (13) of this subsection and subsection (m)(8)(B)(i) of
this section.
(B) Protective
devices, including aprons, gloves, and shields shall be checked annually for
defects such as holes, cracks, and tears. These checks may be performed by the
registrant by visual or tactile means, or x-ray imaging. If a defect is found,
protective devices shall be replaced or removed from service until repaired. A
record of this test shall be made and maintained by the registrant in
accordance with subsection (s)(1) of this section for inspection by the
agency.
(5) Operator
credentialing. Individuals who operate radiation machines for human use shall
meet the appropriate credentialing requirements of rules issued in accordance
with the Medical Radiologic Technologist Certification Act, Texas Occupations
Code, Chapter 601. Copies of the credentialing document shall be maintained at
the location(s) where the individual is working.
(6) Practice of medical physics. Surveys,
tests, or evaluations required by this section may constitute the practice of
medical physics and, therefore, require a license from the Texas Board of
Licensure for Professional Medical Physicists in accordance with the Medical
Physics Practice Act, Texas Occupations Code, Chapter 602.
(7) Exposure of individuals other than the
patient. No individual other than a patient, operator, and ancillary personnel
shall be in the x-ray room or area while exposures are being made unless such
individual's assistance is required.
(8) Holding of patient or image receptor.
(A) When a patient or image receptor must be
held in position during radiography, mechanical supporting or restraining
devices shall be used when the exam permits.
(B) If a patient or image receptor must be
held by an individual during an exposure, that individual shall be protected
with appropriate shielding devices described in paragraph (4) of this
subsection.
(C) In those cases
where the patient must hold the image receptor, any portion of the body other
than the area of clinical interest struck by the useful beam shall be protected
by not less than 0.25 mm lead equivalent material.
(9) Viewing system and contact with patient.
(A) Windows, mirrors, closed circuit
television, or another method shall be provided to permit the operator to
continuously observe the patient during irradiation.
(B) The operator shall be able to maintain
verbal, visual, and aural contact with the patient.
(10) Operator position. The operator position
during the exposure shall be such that the operator's exposure is as low as
reasonably achievable (ALARA) and the operator is a minimum of 6 feet from the
source of radiation or protected by an apron, gloves, or other shielding having
a minimum of 0.25 mm lead equivalent material.
(11) Holding of tube. In no case shall an
individual hold the tube or tube housing assembly supports during any
radiographic exposure.
(12) Patient
protection. Notwithstanding the provisions of paragraph (7) of this subsection,
other patients who are in line with the primary beam and who cannot be removed
from the room shall be protected by whole body protective barriers of a minimum
of 0.25 mm lead equivalent material or so positioned that the nearest portion
of their body is at least 6 feet from both the tube head and the nearest edge
of the image receptor.
(13) Gonadal
shielding. Gonadal shielding shall be used on patients when the gonads are in
or within 5 cm of the useful beam. This requirement does not apply if the
shielding will interfere with the diagnostic procedure. Gonadal shielding shall
be of at least 0.5 mm lead equivalent material.
(14) Measurements of the radiation output for
x-ray, fluoroscopic, and CT systems. Measurements of the radiation output of
x-ray, fluoroscopic, and CT systems shall be performed with a calibrated
dosimetry system in accordance with the following.
(A) The dosimetry system calibration shall be
traceable to a national standard.
(B) Dosimetry systems shall be calibrated
within 24 months from the date of the prior calibration.
(C) Measurements of radiation output shall be
performed with a dosimetry system that has been calibrated within the 24 months
preceding the date of the measurement.
(D) Record of the dosimetry system
calibration shall include:
(i) manufacturer's
name, model and serial number of each calibrated instrument;
(ii) date of the calibration; and
(iii) name of the individual recording the
information.
(E) The
registrant shall record the dosimetry system calibration information specified
in subparagraph (D)(i) - (iii) of this paragraph and maintain that record in
accordance with subsection (s)(1) of this section for inspection by the
agency.