Texas Administrative Code
Title 13 - CULTURAL RESOURCES
Part 1 - TEXAS STATE LIBRARY AND ARCHIVES COMMISSION
Chapter 6 - STATE RECORDS
Subchapter B - MICROFILMING STANDARDS FOR STATE AGENCIES
Section 6.25 - Microfilm Production Tests and Inspections
Universal Citation: 13 TX Admin Code ยง 6.25
Current through Reg. 49, No. 38; September 20, 2024
(a) The following tests must be performed on microfilm as indicated.
(1) Methylene blue
test.
(A) A methylene blue test for
conventionally processed silver-gelatin film must be performed on all original
microfilm in compliance with standard ANSI/NAPM IT9.17.
(B) A test must be performed on a strip of
processed, clear, and unexposed film approximately six inches long from the
original microfilm or on a process control strip at least once each week during
which processing is done.
(C) A
test must also be performed whenever a batch of fixer or developer is changed,
or when changes in processing such as replacement or addition of filter, water
softener, or replenishing system are made, or when water is changed.
(D) If processing is performed by a service
bureau, a test performed for one agency may satisfy this requirement for more
than one agency.
(E) If film fails
to meet the standards established by these sections, it must be rewashed and
retested within 14 days of initial processing.
(F) Once a year, two test strips must be
taken from the same microfilm; one must be sent to an independent testing
laboratory and the other to the laboratory which performs the weekly test, for
verification of test results.
(G)
If either test result is greater than 0.014 grams per square meter, the records
administrator or records custodian must identify and remedy the cause of the
deviation from the required standard.
(2) Density test.
(A) A density test must be performed on all
original microfilm in compliance with standard ANSI/AIIM MS23 and the
background density must be in accordance with that standard.
(B) Densitometer used must be calibrated
daily, when in use.
(3)
Resolution test.
(A) A resolution test must
be performed on all original microfilm in compliance with standard ANSI/AIIM
MS23.
(B) The resolution target
must meet the following standards:
(i)
ANSI/AIIM MS17; and/or
(ii)
ANSI/AIIM MS19; and/or
(iii)
ANSI/AIIM MS51.
(C)
Photocopies of the resolution target must not be used for testing.
(D) Resolution target images must be
evaluated under a microscope to determine if resolution loss from duplicating
is acceptable. A duplicate of a first-generation film must have a resolution
loss of no more than one test pattern of the test objects described in
ANSI/AIIM MS51.
(b) Original processed microfilm must be visually inspected according to the following procedures.
(1) Within two weeks of completion of the
methylene blue test, a visual inspection of microfilm must be completed to
verify legibility.
(2) Film of
essential records or records having a retention period of 10 years or more must
be inspected image by image.
(3)
Film of non-essential records having a retention period of less than 10 years
must be inspected at least every three meters (10 feet) of each roll or every
third fiche.
(4) Images of
documents must be uniformly placed on the film and must be free of any defects
in the filming area that would interfere with the documents being
read.
(5) If a defect is found on
any microfilm, the microfilm immediately preceding and following the sample of
microfilm on which the defect was found must be inspected. If a defect is found
on those microfilms, the uninspected microfilm preceding and/or following those
microfilms must be inspected image by image until all defective film has been
identified.
(c) Testing and Inspection of Equipment.
(1) Cameras and
ancillary equipment must be calibrated, tested, or otherwise inspected and
adjusted at least twice annually or more often if required to comply with
manufacturer's specifications or recommended operating and maintenance
procedures.
(2) Processors must be
monitored on a scheduled basis with process control strips (sensitometric
strips) at a minimum at the start of processing each day and whenever a batch
of film, fixer, or developer is changed, or when changes in processing such as
replacement or addition of filter, water softener, or replenishing system are
made, or when water is changed.
Disclaimer: These regulations may not be the most recent version. Texas may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google
Privacy Policy and
Terms of Service apply.