South Dakota Administrative Rules
Title 44 - Department of Health
Article 44:12 - Tattooing and body piercing
Chapter 44:12:02 - Minimum sanitation standards for body piercing
Section 44:12:02:17 - Sterilization methods

Universal Citation: SD Admin Rules 44:12:02:17
Current through Register Vol. 51, page 43, September 23, 2024

Equipment requiring sterilization shall be sterilized by using the method of steam pressure sterilization with at least 15 pounds of pressure per square inch at a temperature of 250 degrees Fahrenheit in an autoclave for the length of time instructed in the manufacturer's operational manual. This shall be certified by a chemical indicator which is attached to the autoclave bag and turns color when the required temperature has been reached. The owner or operator of the body piercing establishment shall provide lab result slips that state that each autoclave has been bacteriologically tested monthly with commercially prepared packets of Bacillus stearothermophilus or another Bacillus species and has passed the test. These lab result slips shall be kept for at least one year. Dry heat is not an acceptable method of sterilization for the purpose of this article.

General Authority: SDCL 34-1-17(8).

Law Implemented: SDCL 9-34-17, 34-1-17(8).

Disclaimer: These regulations may not be the most recent version. South Dakota may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.