South Dakota Administrative Rules
Title 44 - Department of Health
Article 44:03 - Radiation safety
Chapter 44:03:01 - Radiation control
Section 44:03:01:08.07 - Equipment standards for mobile X ray equipment
Current through Register Vol. 51, page 43, September 23, 2024
In addition to § 44:03:01:08.02 all mobile X ray equipment must:
(1) Meet the specifications in 21 C.F.R. § 1020.30, in effect on March 31, 2000, 21 C.F.R. § 1020.31, in effect on July 2, 1999, and 21 C.F.R. § 1020.32, in effect on May 19, 1994; and
(2) Have the exposure switch located so the operator can stand at least six feet from the useful beam. unless sufficient shielding is provided to protect the operator from stray radiation, in which case the exposure switch may be located closer.
General Authority: SDCL 34-21-4.1, 34-21-15.
Law Implemented: SDCL 34-21-4.1, 34-21-15.