Current through Register Vol. 48, No. 9, September 27, 2024
6.4.1 Equipment
requirements.
6.4.1.1 Leakage radiation. When
the tube is operated at its leakage technique factors, the leakage radiation
shall not exceed the values specified at the distance stated for the
classification of that x-ray system shown in Table 1.
TABLE 1. LEAKAGE LIMITS FOR THERAPEUTIC X-RAY SYSTEMS OF LESS
THAN 1 MeV.
System
Contact Therapy
0-150 kVp (manufactured or
installed prior to January 1, 1994)
|
Leakage Limit 100 mR/hr
1 R in 1 hr.
|
Measurement Location
5 cm from surface of tube housing
1 m from source
|
0-150 kVp (manufactured on or after January 1,
1994)
|
100 mR in 1 hr
|
1 m from source
|
151-500 kVp 500-999 kVp
|
1 R in 1 hr
0.1 percent of 1 R in 1 hr.
|
1 m from source
1 m from source useful beam
|
6.4.1.2 Permanent Beam-Limiting Devices.
Permanent fixed diaphragms or cones used for limiting the useful beam shall
provide the same or a higher degree of protection as required for the tube
housing assembly.
6.4.1.3 Removable
and Adjustable Beam-Limiting Device.
6.4.1.3.1
Removable beam-limiting devices shall, for the portion of the useful beam to be
blocked by these devices, transmit not more than one percent (1%) of the useful
beam at the maximum kV and maximum treatment filter. This requirement does not
apply to auxiliary blocks or materials placed in the useful beam to shape the
useful beam to the individual patient.
6.4.1.3.2 Adjustable beam-limiting devices
shall, for the portion of the x-ray beam to be blocked by these devices,
transmit not more than five percent (5%) of the useful beam at the maximum kV
and maximum treatment filter.
6.4.1.3.3 Adjustable beam-limiting devices
installed after May 25, 2001, shall meet the requirements of
RHB 6.4.1.3.
6.4.1.4 The filter system shall be
so designed that:
6.4.1.4.1 The filters cannot
be accidentally displaced at any possible tube orientation;
6.4.1.4.2 For equipment installed after
January 1, 1994, an interlock system prevents irradiation if the proper filter
is not in place;
6.4.1.4.3 The
radiation at five centimeters (5 cm) from the filter insertion slot opening
does not exceed thirty Roentgens (30 R)(7.74 mC/kg) per hour under any
operating conditions; and
6.4.1.4.4
Each filter is marked as to its material of construction and its thickness. For
wedge filters, the wedge angle shall appear on the wedge or wedge
tray.
6.4.1.5 Tube
Immobilization. The tube housing assembly shall be capable of being immobilized
for stationary treatments.
6.4.1.6
Focal Spot Marking. The tube housing assembly shall be so marked that it is
possible to determine the location of the focal spot to within five millimeters
(5 mm), and such markings shall be readily accessible for use during
calibration procedures.
6.4.1.7
Beam Block. Contact therapy tube housing assemblies shall have a removable
shield of at least 0.5 millimeter lead equivalency at one hundred kilovoltage
peak (100 kVp) that can be positioned over the entire useful beam exit port
during periods when the beam is not in use.
6.4.1.8 Beam Monitoring System. Systems of
greater than one hundred fifty (150 kVp) manufactured after January 1, 1994,
shall be provided with a beam monitoring system which:
6.4.1.8.1 Shall have the detector of the
monitor system interlocked to prevent incorrect positioning;
6.4.1.8.2 Shall not allow irradiation until a
preselected value of exposure has been made at the treatment control
panel;
6.4.1.8.3 Shall
independently terminate irradiation when the preselected exposure has been
reached;
6.4.1.8.4 Shall be so
designed that, in the event of a system malfunction or electrical power
failure, the dose administered to a patient prior to the system malfunction or
power failure can be accurately determined;
6.4.1.8.5 Shall have a display at the control
panel from which the dose at a reference point in soft tissue can be
calculated;
6.4.1.8.6 Shall have a
control panel display which maintains the administered dose reading until
intentionally reset to zero (0); and
6.4.1.8.7 Shall have a control panel display
which does not have scale multiplying factors and utilizes a design such that
increasing dose is displayed by increasing numbers.
6.4.1.9 Timer.
6.4.1.9.1 A timer which has a display shall
be provided at the treatment control panel. The timer shall have a preset time
selector.
6.4.1.9.2 The timer shall
activate with the production of radiation and retain its reading after
irradiation is interrupted. After irradiation is terminated and before
irradiation can be reinitiated, it shall be necessary to reset the elapsed time
indicator to zero (0).
6.4.1.9.3
The timer shall terminate irradiation when a preselected time has elapsed, if
any dose monitoring system present has not previously terminated
irradiation.
6.4.1.9.4 The timer
shall permit accurate presetting and determination of exposure times as short
as one (1) second.
6.4.1.9.5 The
timer shall not permit an exposure if set at zero (0).
6.4.1.9.6 The timer shall not activate until
the shutter is opened when irradiation is controlled by a shutter mechanism
unless calibration includes a timer factor to compensate for mechanical
lag.
6.4.1.9.7 Timers shall be
accurate to within one percent (1%) of the selected value or one (1) second,
whichever is greater.
6.4.1.10 Control Panel Functions. The control
panel, in addition to the displays required in other provisions of this Part,
shall have:
6.4.1.10.1 An indication of
whether electrical power is available at the control panel and if activation of
the x-ray tube is possible;
6.4.1.10.2 An indication of whether x-rays
are being produced;
6.4.1.10.3
Means for indicating x-ray tube potential and current;
6.4.1.10.4 Means for terminating an exposure
at any time;
6.4.1.10.5 A locking
device which will prevent unauthorized use of the x-ray system; and
6.4.1.10.6 For x-ray systems manufactured
after May 25, 2001, a positive display of specific filters in the
beam.
6.4.1.11 Multiple
Tubes. When a control panel may energize more than one (1) x-ray tube:
6.4.1.11.1 It shall be possible to activate
only one (1) x-ray tube at any time;
6.4.1.11.2 There shall be an indication at
the control panel identifying which x-ray tube is activated; and
6.4.1.11.3 There shall be an indication at
the tube housing assembly when that tube is energized.
6.4.1.12 Source-to-Skin Distance (SSD). There
shall be means of determining initially the SSD to within one centimeter (1 cm)
and of producing this measurement to within two millimeters (2 mm)
thereafter.
6.4.1.13 Shutters.
Unless it is possible to bring the x-ray output to the prescribed exposure
parameters within five (5) seconds, the beam shall be attenuated by a shutter
having a lead equivalency not less than that of the tube housing assembly.
6.4.1.13.1 After the unit is at operating
parameters, the shutter shall be controlled electrically by the operator from
the control panel.
6.4.1.13.2 An
indication of shutter position shall appear on the control panel.
6.4.2 Facility Design
Requirements for Therapy X-ray Systems Capable of Operating Above 50 kVp.
6.4.2.1 Aural Communication. Provision shall
be made for two-way aural communication between the patient and the operator at
the control panel. However, where excessive noise levels or treatment
requirements make aural communication impractical, other methods of
communication shall be used.
6.4.2.2 Viewing Systems.
6.4.2.2.1 Windows, mirrors, closed circuit
television, or an equivalent system shall be provided to permit continuous
observation of the patient during irradiation and shall be so located that the
operator can observe the patient from the control panel.
6.4.2.2.2 When the primary viewing system is
by electronic means, an alternate viewing system, which may be electronic,
shall be available for use in the event of failure of the primary viewing
system.
6.4.2.2.3 Should both
systems described in
RHB 6.4.2.2.2 above fail
or be inoperative, treatment shall not be performed with the unit until one of
the systems is restored.
6.4.2.3 Barriers. With equipment operating at
voltages above fifty kilovoltage peak (50 kVp), the required barriers shall be
an integral part of the building.
6.4.2.4 Multiple Access. Treatment rooms to
which access is possible through more than one entrance shall be provided with
flashing warning lights in a readily observable position near the outside of
all access doors, which will indicate when the useful beam is "on." Interlocks
shall be provided such that all entrance doors must be closed, including doors
to any interior booths, before treatment can be initiated or continued. If the
radiation beam is interrupted by any door opening, it shall not be possible to
restore the machine to operation without closing the door and reinitiating
irradiation by manual action at the control panel.
6.4.3 Additional Requirements for X-ray
Systems Capable of Operating Above 150 kVp.
6.4.3.1 All protective barriers shall be
fixed except for entrance doors or beam interceptors.
6.4.3.2 The control panel shall be within a
protective booth equipped with an interlocked door or located outside the
treatment room or in a totally enclosed booth, which has a ceiling, inside the
room.
6.4.3.3 Interlocks shall be
provided such that all entrance doors must be closed, including doors to any
interior booth, before treatment can be initiated or continued. If the
radiation beam is interrupted by any door opening, it shall not be possible to
restore the machine to operation without closing the door and reinitiating
irradiation by manual action at the control panel.
6.4.3.4 When any door referred to in
RHB 6.4.3.3 is opened while
the x-ray tube is activated, the exposure at a distance of one meter (1 m) from
the source shall be reduced to less than one hundred milliroentgen (100 mR) per
hour.
6.4.3.5 A scram button or
other emergency power cut-off switch shall be located and easily identifiable
in all accessible high radiation areas.
6.4.3.6 All safety and warning devices,
including interlocks, shall be tested and appropriately serviced after each
five hundred (500) hours of operation or at intervals not to exceed six (6)
months, whichever comes first. Documentation shall be kept and available for
review of all testing and servicing.
6.4.4 Surveys, Calibrations, and Spot Checks.
6.4.4.1 Surveys.
6.4.4.1.1 All new facilities, and existing
facilities not previously surveyed shall have a survey made by or under the
direction of a qualified expert who is authorized by the Department to perform
such surveys. Such surveys shall be done after any change in the facility or
equipment which might cause a significant increase in radiation hazard. A
record shall be made of the therapeutic operating conditions and radiation
levels measured at specific control points. One (1) of these control points
must be at the normal work station of the operator.
6.4.4.1.2 The registrant shall obtain a
written report of the survey from the qualified expert. A copy of the initial
report shall be transmitted by the registrant to the Department within thirty
(30) calendar days of the first patient treatment following the survey. The
registrant shall maintain all subsequent reports for inspection by the
Department.
6.4.4.1.3 The survey
and report shall indicate all instances where the installation, in the opinion
of the qualified expert, is in violation of applicable rules or
regulations.
6.4.4.1.4 The
registrant shall maintain sufficient calibrated and operable radiation survey
instruments to make physical radiation surveys as required by this regulation.
Each radiation survey instrument shall meet the requirements of
RHB 1.4.4.
6.4.4.2 Calibrations. Calibrations
of x-ray systems subject to the requirements of this Part shall meet the
following requirements:
6.4.4.2.1 The
calibration of an x-ray system shall be performed at intervals not to exceed
one (1) year and after any change or replacement of components which could
cause a change in the radiation output on output.
6.4.4.2.2 The calibration of the radiation
output of the x-ray system shall be performed by or under the direction of a
radiological physicist who is physically present at the facility during such
calibration.
6.4.4.2.3 Calibration
of the radiation output of an x-ray system shall be performed with a calibrated
dosimetry system. The calibration of such system shall meet the requirements of
RHB 1.4.4.
6.4.4.2.4 The calibration shall be such that
the dose at a reference point in a water or tissue equivalent phantom can be
calculated to within an uncertainty of five percent (5%). For superficial
units, free-in-air calibrations are acceptable.
6.4.4.2.5 The calibration of the x-ray system
shall include, but not be limited to, the following determinations:
6.4.4.2.5.1 Verification that the x-ray
system is operating in compliance with the design specifications;
6.4.4.2.5.2 Half-value layer for each kV
setting and filter combination used;
6.4.4.2.5.3 The exposure rates as a function
of field size, technique factors, filter, and treatment distance used;
and
6.4.4.2.5.4 The degree of
congruence between the radiation field and the field indicated by the
localizing device if such device is present, which shall be within five
millimeters (5 mm) for any field edge.
6.4.4.2.6 Records of calibrations shall be
maintained by the registrant for five (5) years after completion of the
calibration. The records shall be available for review.
6.4.4.2.7 A copy of the most recent x-ray
system calibration shall be available at or in the general area of the control
panel.
6.4.4.2.8 A copy of the most
recent x-ray system calibration shall be submitted to the Department upon
request.
6.4.4.3 Spot
Checks. Spot checks shall be performed on x-ray systems capable of operation at
greater than one hundred fifty kilovoltage peak (150 kVp). Such spot checks
shall meet the following requirements:
6.4.4.3.1 The spot check procedures shall be
in writing and shall have been developed by a radiological physicist. A copy of
the procedures shall be submitted to the Department upon request.
6.4.4.3.2 If the radiological physicist does
not perform the spot check measurement, the results of the spot check
measurements shall be reviewed by the radiological physicist within seven (7)
treatment days and a record made of the review.
6.4.4.3.3 The spot check procedures shall
specify the frequency at which tests or measurements are to be performed. The
spot check procedures shall specify that the spot check shall be performed
during the calibration specified in
RHB 6.4.4.2. The acceptable
tolerance for each parameter measured in the spot check when compared to the
value for that parameter determined in the calibration specified in
RHB 6.4.4.2 shall be
stated.
6.4.4.3.4 The written spot
check procedures shall include special operating instructions which shall be
carried out whenever a parameter in
RHB 6.4.4.2 exceeds an
acceptable tolerance.
6.4.4.3.5
Whenever a spot check indicates a significant change in the operating
characteristics of a system, as specified in the spot check procedures, the
system shall be recalibrated, as required in
RHB 6.4.4.2.
6.4.4.3.6 Records of spot check measurements
and any necessary corrective actions shall be maintained by the registrant for
two (2) years after completion of the spot check measurements. A copy of the
most recent spot check shall be available at or in the area of the control
panel.
6.4.4.3.7 Where a spot check
involves a radiation measurement, such measurement shall be obtained using a
system satisfying the requirements of
RHB 6.4.4.2.3 or which
has been intercompared with a system meeting those requirements within the
previous year.
6.4.4.4
Prohibited use. The x-ray system shall not be used in the administration of
radiation therapy unless the requirements of
RHB 6.4.4.2 and 6.4.4.3
have been met.