Current through Register Vol. 48, No. 9, September 27, 2024
5.13.1 Daily
quality control tests. Film processors used to develop mammograms shall be
adjusted and maintained to meet the technical development specifications for
the mammography film in use. A processor performance test shall be performed on
each day that examinations are performed before any clinical films are
processed that day. The test shall include an assessment of base plus fog
density, mid- density, and density difference, using the mammography film used
clinically at the facility.
5.13.1.1 The base
plus fog density shall be within plus 0.03 of the established operating
level.
5.13.1.2 The mid-density
shall be within plus or minus 0.15 of the established operating
level.
5.13.1.3 The density
difference shall be within plus or minus 0.15 of the established operating
level.
5.13.2 Weekly
quality control tests. Facilities with screen-film systems shall perform a
phantom image quality evaluation test, using an FDA-approved phantom, at least
weekly.
5.13.2.1 The optical density of the
film at the center of an image of the phantom shall be at least 1.20 when
exposed under a typical clinical condition.
5.13.2.2 The optical density of the film at
the center of the phantom image shall not change by more than plus or minus
0.20 from the established operating level.
5.13.2.3 The phantom image shall achieve at
least the minimum score established by the accreditation body.
5.13.2.4 The density difference between the
background of the phantom and an added test object, used to assess image
contrast, shall be measured and shall not vary by more than plus or minus 0.05
from the established operating level.
5.13.3 Quarterly quality control tests.
Facilities with screen-film systems shall perform the following quality control
tests at least quarterly:
5.13.3.1 Fixer
retention in film. The residual fixer shall be no more than five micrograms per
square centimeter (5 µg/cm2).
5.13.3.2 Repeat analysis. If the total repeat
or reject rate changes from the previously determined rate by more than two
percent (2%) of the total films included in the analysis, the reason(s) for the
change shall be determined. Any corrective actions shall be recorded and the
results of these corrective actions shall be assessed.
5.13.4 Semiannual quality control tests.
Facilities with screen-film systems shall perform the following quality control
tests at least semiannually:
5.13.4.1 Darkroom
fog. The optical density attributable to darkroom fog shall not exceed 0.05
when a mammography film of the type used in the facility, which has a
mid-density of no less than 1.20, is exposed to typical darkroom conditions for
two (2) minutes while such film is placed on the counter top emulsion side up.
If the darkroom has a safelight used for mammography film, it shall be on
during this test.
5.13.4.2
Screen-film contact. Testing for screen-film contact shall be conducted using
40 mesh copper screen. All cassettes used in the facility for mammography shall
be tested.
5.13.4.3 Compression
device performance. The maximum compression force for the initial power drive
shall be between one hundred eleven newtons (111 N)(25 lbs) and two hundred
nine newtons (209 N)(45 lbs).
5.13.5 Annual quality control tests.
Facilities with screen-film systems shall perform the following quality control
tests at least annually:
5.13.5.1 Automatic
exposure control (AEC) performance.
5.13.5.1.1 The AEC shall be capable of
maintaining film optical density within plus or minus 0.30 of the mean optical
density when the thickness of a homogeneous material is varied over a range of
two to six centimeters (2 to 6 cm) and the kVp is varied appropriately for such
thicknesses over the kVp range used clinically in the facility. If this
requirement cannot be met, a technique chart shall be developed showing
appropriate techniques (kVp and density control settings) for different breast
thicknesses and compositions that shall be used so that optical densities
within plus or minus 0.30 of the average under phototimed conditions can be
produced.
5.13.5.1.2 After October
28, 2002, the AEC shall be capable of maintaining film optical density within
plus or minus 0.15 of the mean optical density when thickness of a homogeneous
material is varied over a range of two to six centimeters (2 to 6 cm) and the
kVp is varied appropriately for such thicknesses over the kVp range used
clinically in the facility.
5.13.5.1.3 The optical density of the film in
the center of the phantom image shall not be less than 1.20.
5.13.5.2 Kilovoltage peak accuracy
and reproducibility. The kVp shall be accurate within plus or minus five
percent (5%) of the indicated or selected kVp at:
5.13.5.2.1 The lowest clinical kVp that can
be measured by a kVp test device;
5.13.5.2.2 The most commonly used clinical
kVp;
5.13.5.2.3 The highest
available clinical kVp; and
5.13.5.2.4 At the most commonly used clinical
setting of kVp, the coefficient of variation of reproducibility of the kVp
shall be equal to or less than 0.02. The kVp shall be checked annually or upon
new x-ray tube installation.
5.13.5.3 Focal spot condition. Until October
28, 2002, focal spot condition shall be evaluated by measuring focal spot
dimensions or by determining system resolution. After October 28, 2002,
facilities shall evaluate focal spot condition only by determining the system
resolution. For focal spot dimensions, the measured values of the focal spot
length (dimension parallel to the anode cathode axis) and width (dimension
perpendicular to the anode cathode axis) shall be within the following
tolerance limits:
Focal Spot Tolerance Limit
Nominal Focal Spot Size (mm)
|
Maximum Width (mm)
|
Measured Dimensions Length (mm)
|
0.10
|
0.15
|
0.15
|
0.15
|
0.23
|
0.23
|
0.20
|
0.30
|
0.30
|
0.30
|
0.45
|
0.65
|
0.40
|
0.60
|
0.85
|
0.60
|
0.90
|
1.30
|
5.13.5.3.1
System Resolution.
5.13.5.3.1.1 Each x-ray
system used for mammography, in combination with the mammography screen-film
combination used in the facility, shall provide a minimum resolution of eleven
(11) cycles per millimeter (mm)(line-pairs/mm) when a high contrast resolution
bar test pattern is oriented with the bars perpendicular to the anode cathode
axis, and a minimum resolution of thirteen (13) line-pairs/mm when the bars are
parallel to that axis.
5.13.5.3.1.2
The bar pattern shall be placed four and one-half centimeters (4.5 cm) above
the breast support surface, centered with respect to the chest wall edge of the
image receptor, and with the edge of the pattern within one centimeter (1 cm)
of the chest wall edge of the image receptor.
5.13.5.3.1.3 When more than one (1) target
material is provided, the measurement shall be made using the appropriate focal
spot for each target material.
5.13.5.3.1.4 When more than one (1)
source-image receptor distance is provided, the test shall be performed at the
SID most commonly used clinically.
5.13.5.3.1.5 Test kVp shall be set at the
value used clinically by the facility for a standard breast and shall be
performed in the AEC mode, if available. If necessary, a suitable absorber may
be placed in the beam to increase exposure times. The screen-film cassette
combination used by the facility shall be used to test for this requirement and
shall be placed in the normal location used for clinical procedures.
5.13.5.3.2 Focal spot dimensions.
Measured values of the focal spot length (dimension parallel to the anode
cathode axis) and width (dimension perpendicular to the anode cathode axis)
shall be within the tolerance limits specified in this Part. The focal spot
shall be checked annually or upon new x-ray tube installation.
5.13.5.4 Exposure Reproducibility.
The coefficient of variation of exposure shall not exceed 0.05 for any specific
combination of selected technique factors. This requirement shall be deemed to
have been met if, when four (4) exposures are made at identical technique
factors, the value of the average exposure ( E) is greater than or equal to
five (5) times the maximum exposure (Emax) minus the minimum exposure (Emin): 5
(Emax - Emin). This requirement shall be checked annually or upon a new
mammography x-ray unit or a new tube installation.
5.13.5.5 Timer Reproducibility. The
coefficient of variation of the timer shall not exceed 0.05. This requirement
shall be deemed to have been met if, with a selected timer setting, the average
exposure period (T) shall be greater than or equal to five (5) times the
maximum exposure period (Tmax) minus the minimum exposure period (Tmin) when
four (4) timer tests are performed: T >= 5 (Tmax- Tmin). This requirement
shall be checked annually or upon a new mammography x-ray unit or a new tube
installation.
5.13.5.6 Timer
Accuracy. Deviation of the selected time setting from indicated time values
shall not exceed the limits specified for that system by its manufacturer. In
the absence of manufacturer's specifications, the deviation shall not exceed
ten percent (10%) of the indicated time value. This requirement shall be
checked annually or upon a new mammography x-ray unit or a new tube
installation.
5.13.5.7 Linearity.
The following requirements apply when the equipment is operated on a power
supply as specified by the manufacturer for any fixed x-ray tube potential
within the range of forty percent to one hundred percent (40% to 100%) of the
maximum rated:
5.13.5.7.1 Equipment having
independent selection of x-ray tube current (mA). The average ratios of
exposure to the indicated milli Ampere-seconds product (C/kg/mAs (or mR/mAs))
obtained at any tube current settings shall not differ by more than 0.10 times
their sum. This is: [X1-X2] < 0.10 (X1+X2); where X1 and X2 are the average
C/kg/mAs (or mR/mAs) values obtained at any two (2) tube current
settings.
5.13.5.7.2 Equipment
having a combined x-ray tube current-exposure time product (mAs) selector, but
not a separate tube current (mA) selector. The average ratios of exposure to
the indicated milliAmpere-seconds product (C/kg/mAs (or mR/mAs)) obtained at
any two mAs selector settings shall not differ by more than 0.10 times their
sum. This is [X1-X2] <0.10 (X1+X2); where X1 and X2 are the average C/kg/mAs
(or mR/mAs) values obtained at any two mAs selector settings.
5.13.5.7.3 Measuring Compliance.
Determination of compliance shall be based on four (4) exposures, at each of
the two (2) settings. The two (2) settings may include any two (2) focal spot
sizes provided that neither focal spot size is equal to or less than .45
millimeter, in which case the two (2) settings shall be restricted to the same
focal spot size. For purposes of this requirement, focal spot size is the
nominal focal spot size specified by the tube manufacturer. Linearity shall
also be checked annually or upon new x-ray tube installation.
5.13.5.8 Beam quality and
half-value layer (HVL). For mammography systems operating at x-ray tube
potentials of less than fifty kilovoltage peak (50 kVp), the HVL in millimeters
of aluminum of the useful beam shall be equal to or greater than the product of
the measured tube potential in kilovolts multiplied by 0.01. The HVL shall be
measured with the compression device in the beam and shall be measured at the
same tube potential used in Appendix A of this Part, Mammography Dose
Measurement Protocol, and Appendix B of this Part, Mammography Phantom Image
Evaluation. The HVL shall be checked annually and after repairs to the system
have been made that could affect the filtration or upon new x-ray tube
installation.
5.13.5.9 Breast
entrance air kerma and AEC reproducibility. The coefficient of variation for
both air kerma and mAs shall not exceed 0.05.
5.13.5.10 Dosimetry. The average glandular
dose delivered during a single craniocaudal view of a phantom simulating a
standard breast shall not exceed 3.0 milligray (mGy) (0.3 rad) per exposure.
The dose shall be determined with technique factors and conditions used
clinically for a standard breast. The average glandular dose shall be checked
annually or upon new tube installation.
5.13.5.11 X-ray field/light field/image
receptor/compression paddle alignment.
5.13.5.11.1 All systems shall have
beam-limiting devices that allow the entire chest wall edge of the x-ray field
to extend to the chest wall edge of the image receptor and provide means to
assure that the x-ray field does not extend beyond any edge of the image
receptor by more than two percent (2%) of the SID. This requirement is for both
large and small cassettes sizes.
5.13.5.11.2 If a light field that passes
through the x-ray beam limitation device is provided, it shall be aligned with
the x-ray field so that the total of any misalignment of the edges of the light
field and the x-ray field along either the length or the width of the visually
defined field at the plane of the breast support surface shall not exceed two
percent (2%) of the SID.
5.13.5.11.3 The chest wall edge of the
compression paddle shall not extend beyond the chest wall edge of the image
receptor by more than one percent (1%) of the SID when tested with the
compression paddle placed above the breast support surface at a distance
equivalent to standard breast thickness. The shadow of the vertical edge of the
compression paddle shall not be visible on the image.
5.13.5.12 Uniformity of screen speed.
Uniformity of screen speed of all the cassettes in the facility shall be tested
and the difference between the maximum and minimum optical densities shall not
exceed 0.30. Screen artifacts shall also be evaluated during this
test.
5.13.5.13 System artifacts.
System artifacts shall be evaluated with a high-grade, defect-free sheet of
homogeneous material large enough to cover the mammography cassette and shall
be performed for all cassette sizes used in the facility using a grid
appropriate for the cassette size being tested. System artifacts shall also be
evaluated for all available focal spot sizes and target filter combinations
used clinically.
5.13.5.14
Radiation output.
5.13.5.14.1 The system
shall be capable of producing a minimum output of 4.5 Gy air kerma per second
(513 mR per second) when operating at twenty-eight kilovoltage peak (28 kVp) in
the standard mammography (moly/moly) mode at any SID where the system is
designed to operate and when measured by a detector with its center located
four and one-half centimeters (4.5 cm) above the breast support surface with
the compression paddle in place between the source and the detector. After
October 28, 2002, the system, under the same measuring conditions, shall be
capable of producing a minimum output of 7.0 Gy air kerma per second (800 mR
per second) when operating at twenty-eight kilovoltage peak (28 kVp) in the
standard (moly/moly) mammography mode at any SID where the system is designed
to operate.
5.13.5.14.2 The system
shall be capable of maintaining the required minimum radiation output averaged
over a 3.0-second period.
5.13.5.15 Decompression. If the system is
equipped with a provision for automatic decompression after completion of an
exposure or interruption of power to the system, the system shall be tested to
confirm that it provides:
5.13.5.15.1 An
override capability to allow maintenance of compression;
5.13.5.15.2 A continuous display of the
override status; and
5.13.5.15.3 A
manual emergency compression release that can be activated in the event of
power or automatic release failure.
5.13.6 The quality assurance requirements of
RHB 4.2.16 and film
processing requirements of
RHB 4.2.17.2 shall be met
except where otherwise mentioned.
5.13.7 Quality control tests - other
modalities. For systems with image receptor modalities other than screen-film,
the quality assurance program shall be substantially the same as the quality
assurance program recommended by the image receptor manufacturer, except that
the average glandular dose must meet the requirements of
RHB 5.13.5.10.
5.13.8 Mobile Units. The facility shall
verify that mammography units used to produce mammograms at more than one (1)
location meet the requirements in
RHB 5.13.1 through
5.13.7. In addition, at each
examination location, before any examinations are conducted, the mobile
mammography system shall be tested using the mammography phantom image
evaluation to establish the adequacy of the image quality produced by the
unit.
5.13.9 Use of test results.
5.13.9.1 After completion of the tests
specified in
RHB 5.13.1 through
5.13.8, the facility shall
compare the test results to the corresponding specified action limits; or for
non-screen film modalities, to the manufacturer's recommended action limits; or
for post-move, pre-examination testing of mobile units, to the limits
established in the test method used by the facility.
5.13.9.2 If the test results fall outside the
action limits, the source of the problem shall be identified and corrective
actions shall be taken and documented:
5.13.9.2.1 Before any further examinations
are performed or any films are processed using the component of the mammography
system that failed any of the tests described in
RHB 5.13.1, 5.13.2, 5.13.4.1,
5.13.4.2, 5.13.4.3, 5.13.5.10, 5.13.6, 5.13.7, or 5.13.8.
5.13.9.2.2 Within thirty (30) calendar days
of the test date for all other tests described in
RHB 5.13.