South Carolina Code of Regulations
Chapter 61 - DEPARTMENT OF HEALTH AND ENVIRONMENTAL CONTROL
Subchapter 61-64 - X-Rays (Title B)
Part IX - PERSONNEL SECURITY SCREENING SYSTEMS USING X-RAY EQUIPMENT
Section 61-64.IX.RHB 9.4 - Shielding
Current through Register Vol. 48, No. 9, September 27, 2024
Prior to installation or replacement, the registrant shall submit a floor plan and equipment arrangement which has been prepared by a registered Class II vendor and submitted to the Department for review and acceptance.
9.4.1 The floor plan must include, at a minimum:
9.4.2 An inspection zone shall be established around the personnel security screening system where bystanders are prohibited during the operation of the device. A means shall be provided for any operator responsible for initiating a scan to maintain full visual surveillance of the inspection zone. The registrant shall ensure only the scanned individual is within two meters (2 m) of the scanner when in operation.
9.4.3 The Department may require a shielding plan, as described in RHB 4.4.