South Carolina Code of Regulations
Chapter 61 - DEPARTMENT OF HEALTH AND ENVIRONMENTAL CONTROL
Subchapter 61-64 - X-Rays (Title B)
Part IV - USE OF X-RAYS IN THE HEALTH PROFESSIONS
Section 61-64.IV.RHB 4.11 - Computed Tomography (CT) X-ray Systems
Universal Citation: SC Code Regs 61-64.IV.RHB 4.11
Current through Register Vol. 48, No. 9, September 27, 2024
In addition to the applicable provisions of this regulation, the requirements of this Part apply to all stationary, transportable, and mobile CT X-ray systems.
4.11.1 Equipment Requirements.
4.11.1.1 Tomographic
Plane Indication and Alignment.
4.11.1.1.1 For
any single tomogram system, means shall be provided to permit visual
determination of the tomographic plane or a reference plane offset from the
tomographic plane.
4.11.1.1.2 For
any multiple tomogram system, means shall be provided to permit visual
determination of the location of a reference plane. The reference plane can be
offset from the location of the tomographic planes.
4.11.1.1.3 If a device using a light source
is used to satisfy the requirements of
RHB 4.11.1.1.1 or
4.11.1.1.2, the light source shall provide illumination levels sufficient to
permit visual determination of the location of the tomographic plane or
reference plane under ambient light conditions.
4.11.1.2 Indication of CT Conditions of
Operation. The CT x-ray system shall be designed such that the CT conditions of
operation to be used during a scan or a scan sequence shall be indicated prior
to the initiation of a scan or a scan sequence. On equipment having all or some
of these conditions of operation at fixed values, this requirement may be met
by permanent markings. Indication of CT conditions of operation shall be
visible from any position from which scan initiation is possible.
4.11.1.3 Beam-On and Shutter Status
Indicators and Control Switches
4.11.1.3.1 The
x-ray control and gantry shall provide visual indication whenever x-rays are
produced and, if applicable, whether the shutter is open or closed.
4.11.1.3.2 All emergency buttons or switches
shall be clearly labeled as to their functions.
4.11.1.4 Termination of Exposure.
4.11.1.4.1 Means shall be provided to
terminate the x-ray exposure automatically by either de-energizing the x-ray
source or shuttering the x-ray beam in the event of equipment failure affecting
data collection. Such termination shall occur within an interval that limits
the total scan time to no more than one hundred ten percent (110%) of its
preset value through the use of either a backup timer or devices which monitor
equipment function.
4.11.1.4.2 A
visible signal shall indicate when the x-ray exposure has been terminated
through the means required by
RHB
4.11.1.4.1.
4.11.1.4.3 The operator shall be able to
terminate the x-ray exposure at any time during a scan, or series of scans
under x-ray system control, of greater than one-half (0.5) second
duration.
4.11.1.5
Additional Requirements Applicable to CT X-ray Systems Containing a Gantry
Manufactured After September 3, 1985.
4.11.1.5.1 The total error in the indicated
location of the tomographic plane or reference plane shall not exceed five
millimeters (5 mm).
4.11.1.5.2 If
the x-ray production period is less than one-half (0.5) second, the indication
of x-ray production shall be actuated for at least one-half (0.5) second.
Indicators at or near the gantry shall be discernible from any point external
to the patient opening where insertion of any part of the human body into the
primary beam is possible.
4.11.1.5.3 The deviation of indicated scan
increment versus actual increment shall not exceed to within one millimeter (1
mm) with any mass from zero to one hundred kilograms (0 to 100 kg) resting on
the support device. The patient support device shall be incremented from a
typical starting position to the maximum incremented distance or thirty
centimeters (30 cm), whichever is less, and then returned to the starting
position. Measurement of actual versus indicated scan increment can be taken
anywhere along this travel.
4.11.1.5.4 Premature termination of the x-ray
exposure by the operator shall necessitate resetting of the CT conditions of
operation prior to the initiation of another scan.
4.11.2 Facility Design Requirements.
4.11.2.1 The control panel and
x-ray control must be mounted in a permanently protected area outside the
computed tomography room. The operator is required to remain in that protected
area during the entire exposure, except when performing procedures requiring
the use of exposure switches located on or near the CT gantry and designed to
provide a delay before initiating x-rays and provided all requirements of
RHB 4.2.9 are met.
4.11.2.2 Aural Communication. Provision shall
be made for two-way aural communication between the patient and the operator at
the control panel.
4.11.2.3
Facilities designed with an open area in the control room that leads to the
gantry shall mark this open area conspicuously and indicate not to stand or sit
in this area during x-ray exposures.
4.11.2.4 Viewing Systems.
4.11.2.4.1 Windows, mirrors, closed-circuit
television, or an equivalent shall be provided to permit continuous observation
of the patient during irradiation and shall be so located that the operator can
observe the patient from the control panel.
4.11.2.4.2 When the primary viewing system is
by electronic means, an alternate viewing system (which may be electronic)
shall be available for use in the event of failure of the primary viewing
system.
4.11.3 Equipment Performance Tests and Routine Quality Control
4.11.3.1 Equipment
Performance Tests
4.11.3.1.1 Equipment
performance tests shall be performed by a Class IX vendor.
4.11.3.1.2 Evaluation standards and
tolerances shall be established by the Class IX vendor and maintained by the
facility. These standards and tolerances shall meet nationally recognized
standards and tolerances for the CT x-ray system and shall include the required
minimum criteria for performance tests provided by Appendix F.
4.11.3.1.3 The measurements of the radiation
output of the CT x-ray system shall be performed with a calibrated dosimetry
system. The calibration of such system shall be traceable to a national
standard. The dosimetry system shall have been calibrated within the preceding
two (2) years.
4.11.3.1.4 Records
of equipment performance tests performed shall be maintained for inspection by
the Department.
4.11.3.2
Routine Quality Control (QC)
4.11.3.2.1 A
routine QC program shall be developed by or have written approval by a Class IX
vendor and include:
4.11.3.2.1.1 Instructions
on performing routine QC;
4.11.3.2.1.2 Frequency and conditions of QC
testing;
4.11.3.2.1.3 Acceptable
tolerances for items evaluated; and
4.11.3.2.1.4 Daily use of a water equivalent
phantom to evaluate CT number, noise, and artifacts.
4.11.3.2.2 The CT operator shall have access
to the QC program and the results of the most recent routine QC completed on
the system.
4.11.3.2.3 Routine QC
records shall be documented and maintained for inspection by the Department.
Records shall be maintained for two (2) years or the next Department
inspection, whichever is later.
4.11.4 Ancillary personnel who are not necessary for the safety of the patient shall not be present in the area of the CT unit while exposures are being made.
4.11.5 Cone Beam Computed Tomography (CBCT) Systems
4.11.5.1 The registrant shall follow
QC recommendations provided by the CBCT manufacturer. In the absence of
manufacturer-provided QC recommendations, the registrant shall implement and
document QC guidelines established by a Class IX vendor in accordance to
nationally recognized guidelines or those recognized by the
Department.
4.11.5.2 As applicable,
all provisions of
RHB 4.4 and 4.11 apply, except
4.11.2.4 and 4.11.3.2.1 through 4.11.3.2.2.
4.11.5.3 The minimum source-skin distance
shall not be less than thirty centimeters (30 cm), except veterinary
equipment.
4.11.5.4 Beam alignment.
The x-ray field in the plane of the image receptor shall not exceed beyond the
edge of the image receptor by more than two percent (2%) of the SID when the
axis of the x-ray beam is perpendicular to the plane of the image receptor. In
addition, the center of the x-ray field shall be aligned with the center of the
image receptor to within two percent (2%) of the SID.
4.11.5.5 The registrant shall implement and
document a policy addressing deviations from established protocols.
4.11.5.6 The following information shall be
readily available to the CBCT operator:
4.11.5.6.1 Instructions on performing routine
QC, including the use of the CBCT phantom(s);
4.11.5.6.2 A schedule of routine QC
appropriate for the system;
4.11.5.6.3 Allowable variations set by the
Class IX vendor, if required, for the indicated parameters; and
4.11.5.6.4 The results of at least the most
recent routine QC completed on the system.
Disclaimer: These regulations may not be the most recent version. South Carolina may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google
Privacy Policy and
Terms of Service apply.