Section 61-63.VIII.RHA 8.9 - Radiation Survey Instruments
Current through Register Vol. 48, No. 9, September 27, 2024
8.9.1 The licensee shall keep a calibrated and operable radiation survey instrument capable of detecting beta and gamma radiation at each field station and temporary jobsite to make the radiation surveys required by this Part and by Part III of these regulations. To satisfy this requirement, the radiation survey instrument must be capable of measuring 0.001 mSv (0.1 mrem) per hour through at least 0.5 mSv (50 mrem) per hour
8.9.2 The licensee shall have available additional calibrated and operable radiation detection instruments sensitive enough to detect the low radiation and contamination levels that could be encountered if a sealed source ruptured. The licensee may own the instruments or may have a procedure to obtain them quickly from a second party.
8.9.3 The licensee shall have each radiation survey instrument required under RHA 8.9.1 calibrated as follows:
8.9.4 The licensee shall retain calibration records for a period of 3 years after the date of calibration for inspection by the Department.