Section 61-63.V.RHA 5.8 - Radiation Survey Instruments
Universal Citation: SC Code Regs 61-63.V.RHA 5.8
Current through Register Vol. 48, No. 9, September 27, 2024
5.8.1 The licensee shall keep sufficient calibrated and operable radiation survey instruments at each location where radioactive material is present to make the radiation surveys required by this part and by Part III. Instrumentation required by this section must be capable of measuring a range from 2 millirems (0.02 millisieverts) per hour through 1 rem (0.01 sievert) per hour.
5.8.2 The licensee shall have each radiation survey instrument required under RHA 5.8.1 calibrated:
5.8.2.1 At intervals not to exceed 6 months and
after instrument servicing, except for battery changes;
5.8.2.2 For linear scale instruments, at two
points located approximately one-third and two-thirds of full-scale on each scale;
for logarithmic scale instruments, at mid-range of each decade, and at two points of
at least one decade; and for digital instruments, at 3 points between 2 and 1000
millirems (0.02 and 10 millisieverts) per hour; and
5.8.2.3 So that an accuracy within plus or minus
20 percent of the calibration source can be demonstrated at each point
checked.
5.8.3 Each licensee shall maintain records of the calibrations of its radiation survey instruments and retain each record for 3 years after it is made.
Disclaimer: These regulations may not be the most recent version. South Carolina may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.