South Carolina Code of Regulations
Chapter 61 - DEPARTMENT OF HEALTH AND ENVIRONMENTAL CONTROL
Subchapter 61-63 - Radioactive Materials (Title A)
Part IV - USE OF RADIONUCLIDES IN THE HEALTH PROFESSIONS
Subpart L - RECORDS
Section 61-63.IV.L.RHA 4.109 - Records of Dosimetry Equipment Used with Remote Afterloader Units, Teletherapy Units, and Gamma Stereotactic Radiosurgery Units

Current through Register Vol. 48, No. 9, September 27, 2024

4.109.1 A licensee shall retain a record of the calibration, intercomparison, and comparisons of its dosimetry equipment done in accordance with RHA 4.63 for the duration of the license.

4.109.2 For each calibration, intercomparison, or comparison, the record must include--

4.109.2.1 The date;

4.109.2.2 The manufacturer's name, model numbers and serial numbers of the instruments that were calibrated, intercompared, or compared as required by RHA 4.63.1 and 4.63.2;

4.109.2.3 The correction factor that was determined from the calibration or comparison or the apparent correction factor that was determined from an intercomparison; and

4.109.2.4 The names of the individuals who performed the calibration, intercomparison, or comparison.

Disclaimer: These regulations may not be the most recent version. South Carolina may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.