South Carolina Code of Regulations
Chapter 61 - DEPARTMENT OF HEALTH AND ENVIRONMENTAL CONTROL
Subchapter 61-63 - Radioactive Materials (Title A)
Part IV - USE OF RADIONUCLIDES IN THE HEALTH PROFESSIONS
Subpart H - PHOTON EMITTING REMOTE AFTERLOADER UNITS, TELETHERAPY UNITS, AND GAMMA STEREOTACTIC RADIOSURGERY UNITS
Section 61-63.IV.H.RHA 4.70 - Additional Technical Requirements for Mobile Remote Afterloader Units
Current through Register Vol. 48, No. 9, September 27, 2024
4.70.1 A licensee providing mobile remote afterloader service shall--
4.70.2 In addition to the periodic spot-checks required by RHA 4.68, a licensee authorized to use mobile afterloaders for medical use shall perform checks on each remote afterloader unit before use at each address of use. At a minimum, checks must be made to verify the operation of--
4.70.3 In addition to the requirements for checks in RHA 4.70.2, a licensee shall ensure overall proper operation of the remote afterloader unit by conducting a simulated cycle of treatment before use at each address of use.
4.70.4 If the results of the checks required in RHA 4.70.2 indicate the malfunction of any system, a licensee shall lock the control console in the off position and not use the unit except as may be necessary to repair, replace, or check the malfunctioning system.
4.70.5 A licensee shall retain a record of each check required by RHA 4.70.2 in accordance with RHA 4.114.