South Carolina Code of Regulations
Chapter 61 - DEPARTMENT OF HEALTH AND ENVIRONMENTAL CONTROL
Subchapter 61-63 - Radioactive Materials (Title A)
Part IV - USE OF RADIONUCLIDES IN THE HEALTH PROFESSIONS
Subpart H - PHOTON EMITTING REMOTE AFTERLOADER UNITS, TELETHERAPY UNITS, AND GAMMA STEREOTACTIC RADIOSURGERY UNITS
Section 61-63.IV.H.RHA 4.67 - Periodic Spot-Checks for Teletherapy Units
Current through Register Vol. 48, No. 9, September 27, 2024
4.67.1 A licensee authorized to use teletherapy units for medical use shall perform output spot-checks on each teletherapy unit once in each calendar month that include determination of--
4.67.2 A licensee shall perform measurements required by RHA 4.67.1 in accordance with written procedures established by the authorized medical physicist. That individual need not actually perform the spot-check measurements.
4.67.3 A licensee shall have the authorized medical physicist review the results of each spot-check within 15 days. The authorized medical physicist shall notify the licensee as soon as possible in writing of the results of each spot-check.
4.67.4 A licensee authorized to use a teletherapy unit for medical use shall perform safety spot-checks of each teletherapy facility once in each calendar month and after each source installation to assure proper operation of--
4.67.5 If the results of the checks required in RHA 4.67.4 indicate the malfunction of any system, a licensee shall lock the control console in the off position and not use the unit except as may be necessary to repair, replace, or check the malfunctioning system.
4.67.6 A licensee shall retain a record of each spot-check required by RHA 4.67.1 and 4.67.4 and a copy of the procedures required by RHA 4.67.2, in accordance with RHA 4.111.