South Carolina Code of Regulations
Chapter 61 - DEPARTMENT OF HEALTH AND ENVIRONMENTAL CONTROL
Subchapter 61-63 - Radioactive Materials (Title A)
Part IV - USE OF RADIONUCLIDES IN THE HEALTH PROFESSIONS
Subpart H - PHOTON EMITTING REMOTE AFTERLOADER UNITS, TELETHERAPY UNITS, AND GAMMA STEREOTACTIC RADIOSURGERY UNITS
Section 61-63.IV.H.RHA 4.66 - Full Calibration Measurements on Gamma Stereotactic Radiosurgery Units
Current through Register Vol. 48, No. 9, September 27, 2024
4.66.1 A licensee authorized to use a gamma stereotactic radiosurgery unit for medical use shall perform full calibration measurements on each unit--
4.66.2 To satisfy the requirement of RHA 4.66.1, full calibration measurements must include determination of--
4.66.3 A licensee shall use the dosimetry system described in RHA 4.63.1 to measure the output for one set of exposure conditions. The remaining radiation measurements required in RHA 4.66.2.1 may be made using a dosimetry system that indicates relative dose rates.
4.66.4 A licensee shall make full calibration measurements required by RHA 4.66.1 in accordance with published protocols accepted by nationally recognized bodies.
4.66.5 A licensee shall mathematically correct the outputs determined in RHA 4.66.2.1 at intervals not exceeding 1 month for cobalt-60 and at intervals consistent with 1 percent physical decay for all other radionuclides.
4.66.6 Full calibration measurements required by RHA 4.66.1 and physical decay corrections required by RHA 4.66.5 must be performed by the authorized medical physicist.
4.66.7 A license shall retain a record of each calibration in accordance with RHA 4.110.