South Carolina Code of Regulations
Chapter 61 - DEPARTMENT OF HEALTH AND ENVIRONMENTAL CONTROL
Subchapter 61-63 - Radioactive Materials (Title A)
Part IV - USE OF RADIONUCLIDES IN THE HEALTH PROFESSIONS
Subpart H - PHOTON EMITTING REMOTE AFTERLOADER UNITS, TELETHERAPY UNITS, AND GAMMA STEREOTACTIC RADIOSURGERY UNITS
Section 61-63.IV.H.RHA 4.65 - Full Calibration Measurements on Remote Afterloader Units

Current through Register Vol. 48, No. 9, September 27, 2024

4.65.1 A licensee authorized to use a remote afterloader unit for medical use shall perform full calibration measurements on each unit--

4.65.1.1 Before the first medical use of the unit;

4.65.1.2 Before medical use under the following conditions:
4.65.1.2.1 Following replacement of the source or following reinstallation of the unit in a new location outside the facility; and

4.65.1.2.2 Following any repair of the unit that includes removal of the source or major repair of the components associated with the source exposure assembly; and

4.65.1.3 At intervals not exceeding 1 quarter for high dose-rate, medium dose-rate, and pulsed dose-rate remote afterloader units with sources whose half-life exceeds 75 days; and

4.65.1.4 At intervals not exceeding 1 year for low dose-rate remote afterloader units.

4.65.2 To satisfy the requirement of RHA 4.65.1, full calibration measurements must include, as applicable, determination of:

4.65.2.1 The output within 5 percent;

4.65.2.2 Source positioning accuracy to within 1 millimeter;

4.65.2.3 Source retraction with backup battery upon power failure;

4.65.2.4 Length of the source transfer tubes;

4.65.2.5 Timer accuracy and linearity over the typical range of use;

4.65.2.6 Length of the applicators; and

4.65.2.7 Function of the source transfer tubes, applicators, and transfer tube-applicator interfaces.

4.65.3 A licensee shall use the dosimetry system described in RHA 4.63.1 to measure the output.

4.65.4 A licensee shall make full calibration measurements required by RHA 4.65.1 in accordance with published protocols accepted by nationally recognized bodies.

4.65.5 In addition to the requirements for full calibrations for low dose-rate remote afterloader units in RHA 4.65.2, a licensee shall perform an autoradiograph of the source(s) to verify inventory and source(s) arrangement at intervals not exceeding 1 quarter.

4.65.6 For low dose-rate remote afterloader units, a licensee may use measurements provided by the source manufacturer that are made in accordance with RHA 4.65.1 through 4.65.5.

4.65.7 A licensee shall mathematically correct the outputs determined in RHA 4.65.2.1 for physical decay at intervals consistent with 1 percent physical decay.

4.65.8 Full calibration measurements required by RHA 4.65.1 and physical decay corrections required by RHA 4.65.7 must be performed by the authorized medical physicist.

4.65.9 A licensee shall retain a record of each calibration in accordance with RHA 4.110.

Disclaimer: These regulations may not be the most recent version. South Carolina may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.